Volumetric-modulated arc therapy using multicriteria optimization for body and extremity sarcoma.

Michael R Young1, David L Craft, Caroline M Colbert, Kyla Remillard, Liam Vanbenthuysen, Yi Wang.   

Abstract

This study evaluates the implementation of volumetric-modulated arc therapy (VMAT) using multicriteria optimization (MCO) in the RayStation treatment planning system (TPS) for complex sites, namely extremity and body sarcoma. The VMAT-MCO algorithm implemented in RayStation is newly developed and requires an integrated, comprehensive analysis of plan generation, delivery, and treatment efficiency. Ten patients previously treated by intensity-modulated radiation therapy (IMRT) with MCO were randomly selected and replanned using VMAT-MCO. The plan quality was compared using homogeneity index (HI) and conformity index (CI) of the planning target volume (PTV) and dose sparing of organs at risk (OARs). Given the diversity of the tumor location, the 10 plans did not have a common OAR except for skin. The skin D50 and Dmean was directly compared between VMAT-MCO and IMRT-MCO. Additional OAR dose points were compared on a plan-by-plan basis. The treatment efficiency was compared using plan monitor units (MU) and net beam-on time. Plan quality assurance was performed using the Sun Nuclear ArcCHECK phantom and a gamma criteria of 3%/3 mm. No statistically significant differences were found between VMAT- and IMRT-MCO for HI and CI of the PTV or D50 and Dmean to the skin. The VMAT-MCO plans showed general improvements in sparing to OARs. The VMAT-MCO plan set showed statistically significant improvements over the IMRT-MCO set in treatment efficiency per plan MU (p < 0.05) and net beam-on time (p < 0.01). The VMAT-MCO plan deliverability was validated. Similar gamma passing rates were observed for the two modalities. This study verifies the suitability of VMAT-MCO for sarcoma cancer and highlighted the comparability in plan quality and improve-ment in treatment efficiency offered by VMAT-MCO as compared to IMRT-MCO.
© 2016 The Authors.

Entities:  

Mesh:

Year:  2016        PMID: 27929501      PMCID: PMC5690529          DOI: 10.1120/jacmp.v17i6.6547

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


I. INTRODUCTION

Volumetric‐modulated arc therapy (VMAT) is the dynamic evolution of intensity‐modulated radiation therapy (IMRT). VMAT uses arcs to produce highly conformal dose distributions in complex patient geometry via concerted action of gantry rotation, multileaf collimator (MLC) motion, and dose rate modulation. Unlike the fixed gantry angles characteristic of IMRT delivery, VMAT involves dynamic gantry movement, which requires accurate synchronization of all moving components. The decreased treatment time, potential increased OAR dose sparing, and optimization of monitor units (MUs) afforded by VMAT over IMRT planning makes it ideal for clinical implementation. VMAT planning is currently available in several commercial treatment planning systems (TPS), including Pinnacle SmartArc (Philips, Inc., Andover, MA), Monaco (Elekta, Inc., Stockholm, Sweden), Eclipse (Varian Medical Systems, Inc., Palo Alto, CA), and RayStation (RaySearch Laboratories, Stockholm, Sweden). Each system has its advantages and disadvantages for VMAT planning, and requires an expert planner in order to achieve high‐quality plans. In this study we focused on VMAT planning in RayStation, a TPS popularized by its development and implementation of multicriteria optimization (MCO). The MCO algorithm increases IMRT plan quality by allowing planners to better approach an optimal plan by interactively balancing several treatment objectives and constraints. Planners and physicians navigate high‐dimensional Pareto surfaces and weigh trade‐offs between objectives in real time to improve planning and plan quality. The mathematics and utility of MCO are elucidated in greater detail by Craft and Bortfeld and recent studies suggest dosimetric and efficiency advantages of MCO in generating optimal IMRT treatment plans. Algorithmic improvements in RayStation v4.7.2 have fully extended MCO to VMAT planning, but to date VMAT‐MCO has only been assessed and validated for prostate. We aim to implement fully‐functional VMAT‐MCO planning for more advanced treatment sites, and begin with sarcoma. Our integrative analysis includes an evaluation of sarcoma cancer VMAT‐MCO plan generation, delivery efficiency, and dosimetric comparison with IMRT‐MCO. The purposes of this study are twofold: to generate and deliver the first sarcoma cancer treatments using RayStation VMAT‐MCO, and subsequently to compare the generated plans’ dosimetric and delivery characteristics to paired IMRT‐MCO plans.

II. MATERIALS AND METHODS

A. Sarcoma cancer cases

Ten randomly‐selected sarcoma cases were planned with VMAT‐MCO by dosimetrists using patient CT scans and the target volumes for plans previously implemented clinically using IMRT‐MCO. Of the 10 plans used, 5 were for extremity sarcoma and 5 for body sarcoma patients. Sarcomas grow in connective tissue in various regions of the body, and thus planning target volumes (PTV), clinical target volumes (CTV), and their respective prescriptions varied by case. To account for the diversity of sarcoma locations, we compared treatments of the right arm, forearm, leg, gluteus, calf, left torso, and pelvis. Table 1 shows the target volumes and prescription for each plan. Sarcoma planning standards at our clinic require that 100% of the PTV receives 95% of the prescribed dose and that the maximum dose to the PTV is less than 110% of the prescribed dose. Each plan optimization used the same objectives and constraints as the corresponding IMRT‐MCO plan.
Table 1

Target volume and prescription dose for each selected sarcoma treatment

# of Fields # of Control Points
Plan # Type Target Rx Dose (cGy) IMRT VMAT IMRT VMAT
1ExtremityRt Calf440074138189
2ExtremityRt Thigh440054100126
3ExtremityRt Arm500052150222
4ExtremityRt Forearm660054116139
5ExtremityRt Leg500064164169
6BodyLt Pelvis690062100196
7BodyPelvis540472166282
8BodyLt Torso500082116216
9BodyRt Gluteus500072182178
10BodyRt Gluteus160062194178
Target volume and prescription dose for each selected sarcoma treatment

B. VMAT‐MCO

One goal of MCO is to allow for planners of any level to achieve high‐quality treatment plans in less time and with easier physician interaction than rival approaches. It allows for real‐time navigation of the optimization solution set, affording planners and physicians a valuable tool for finding and selecting Pareto‐optimal plans for treatment. The utility of MCO, and particularly IMRT‐MCO, compared to non‐MCO algorithms has been studied (e.g., McGarry et al. and Hong et al. ), but this comparison has not yet been fully extended to VMAT. RayStation MCO uses a fluence‐based approach in its MCO module, meaning that the precomputed plans are optimized via fluence maps rather than the deliverable segment‐based dose engine. Computing a deliverable plan after navigation can lead to plan degradation. This has not proven to be a large problem clinically for MCO‐IMRT, but initial experience with MCO‐VMAT in RayStation v.4.0 showed that the discrepancy between navigated dose and final deliverable plan dose was often unacceptably large. This is because VMAT is a more challenging optimization problem due to its inherent nonconvexity. Clinical VMAT algorithms such as that implemented in RayStation also put a high focus on delivery efficiency, which can compromise plan quality, particularly for highly modulated fields. If they instead focused on fluence map reproduction fidelity, they could use sliding window‐like VMAT plans to reproduce optimal fluence maps. This would lead to longer plan delivery times, but would make MCO‐VMAT more similar to MCO‐IMRT, which achieves high fidelity in dose recreation by allowing for each fluence map to be recreated with many MLC segments.

C. Plan generation, delivery, and measurement

C.1 Plan generation

Plans in this study were generated with the RayStation v.4.7.2 VMAT‐MCO module. In order to help minimize study uncertainty, each VMAT‐MCO plan was generated by the dosimetrist who generated its corresponding IMRT‐MCO plan. IMRT‐MCO plans used between five and nine fields, depending on treatment site, and the number of fields and dose control points for each plan can be found in Table 1. VMAT‐MCO plans used two or four arcs, depending on treatment site, and the number of fields and dose control points for each plan can also be found in Table 1. All VMAT‐MCO plans were set at 2° gantry spacing at variable gantry speeds and dose rates. Each planner iteratively set constraints and fine‐tuned objectives for MCO optimization to approach an optimal treatment plan. Objectives and constraints varied for each plan, but generally sought maximal dose sparing to relevant organs at risk (OAR) and maximal target coverage to the PTV. Plans then underwent dose reconstruction using machine parameters, at which point each VMAT‐MCO plan was confirmed to be clinically viable by a physician.

C.2 Plan delivery

All 20 plans (10 VMAT and 10 IMRT) were delivered on an Elekta Agility linear accelerator (Elekta, Stockholm, Sweden) at an energy of 6 M V. The Elekta Agility was previously commissioned for both IMRT and VMAT treatments. Machine parameters include a maximum field size of , a leaf pitch of 0.5 cm, and 160 MLC leaves that move with a maximum speed of 3.5 cm/s.

C.3 Plan measurement

All 20 plans were measured using a Sun Nuclear ArcCHECK, a cylindrical dose measurement tool which consists of 1,386 0.019 mm2 diode detectors each spaced 1.0 cm apart. Prior studies have validated the ArcCHECK as an IMRT and VMAT quality assurance (QA) measuring tool (e.g., Nelms et al. ). For dosimetric evaluation, we use a gamma passing rate of 3%/3 mm. The ArcCHECK is a three‐dimensional measuring device which allows for entry and exit dose measurements to provide valuable information for VMAT QA potentially missed by two‐dimensional QA devices.

D. Plan comparison metrics

VMAT‐MCO and IMRT‐MCO plans were compared by plan quality and treatment efficiency. Plan quality was evaluated by comparing OAR dose sparing, conformity index (CI), and homogeneity index (HI) averages for the 10 patient sets. Due to the wide spread of targets among the selected sarcoma plans, only dose‐sparing comparisons of D50 and to an external body contour of the skin were evaluated between modalities. To compensate for the wide range in prescription doses, skin doses for each plan were scaled to a prescription dose of 5000 cGy by a factor of 5000 cGy/RxDose. The variation in target by plan prevents further direct comparisons between the VMAT‐MCO and IMRT‐MCO sets. To compensate, dosimetrists also selected plan‐specific OAR DVH points to generally compare dose sparing between modalities. Points of interest (POIs) specific to each extremity sarcoma plan include V40 (%) of the tibia, humerus, and right femur. POIs specific to each body sarcoma plan include of the colon, of the right femur, V20 of the left lung, of the cauda equina, and of the rectum. The target coverage parameters HI and CI are defined as: where, for is the dose to 5% of the PTV and is the dose to 95% of the PTV. For is the PTV volume receiving the prescribed dose and is the total patient volume receiving the prescribed dose. Both HI and CI are perfect at unity, and depreciate as they approach zero. Treatment efficiency was compared by treatment time (net beam‐on time, in seconds) and monitor unit number for all plans. Plan quality assurance was evaluated with a 3%/3 mm gamma criteria using a Sun Nuclear ArcCHECK device. Current clinical standards require the global gamma passing rate percentage greater than 90% for each plan. All parameters were assessed for statistical significance between treatment modalities (VMAT‐MCO vs. IMRT‐MCO) by the two‐tailed Wilcoxon signed‐rank test.

III. RESULTS

All 10 VMAT‐MCO generated treatments met physician‐imposed PTV coverage (100% of the PTV receiving more than 95% of prescription dose) and plan‐specific OAR dose‐sparing constraints. PTV homogeneity ( for VMAT‐MCO vs. for IMRT‐MCO) and conformity ( for VMAT‐MCO vs. for IMRT‐MCO) showed no significant differences between planning modalities. Skin dose, scaled to a PTV prescription dose of 5000 cGy, slightly decreased on average for VMAT‐MCO plans at D50 ( vs. ). A summary of target coverage and skin dose‐sparing comparisons is shown in Table 2.
Table 2

Average plan results for IMRT‐MCO and VMAT‐MCO plans

Index IMRT‐MCO VMAT‐MCO Statistically significant?
PTV HI 0.92±0.04 0.93±0.03 No
PTV CI 0.89±0.08 0.91±0.08 No
Skin Dmean(cGy)a 990±590 1020±570 No
Skin D50(cGy)a 350±500 340±570 No
Delivery time 519±173 166±32 Yes (p<0.01)
Output 759±370 488±136 Yes (p<0.05)

Scaled to Rx Dose of 5000 cGy.

The wide spread of target location (e.g., right gluteus compared to right calf) within the patient cohort prevented additional baseline OAR dose‐sparing comparisons between modalities. To overcome this, we compared a relevant OAR DVH point for each plan pair, and we observed a general improvement for VMAT‐MCO over IMRT‐MCO plans, as seen in Table 3. VMAT‐MCO provided superior dose sparing to 9 of the 10 points of interest, with the most significant improvements found in plan 9 (cauda equina reduced from 4,048 cGy in IMRT‐MCO to 2,410 cGy in VMAT‐MCO) and plan 6 (colon reduced from 1,779 cGy in IMRT‐MCO to 1,028 cGy in VMAT‐MCO). Figure 1 shows a dose distribution comparison between VMAT‐MCO and IMRT‐MCO for plan 9 (body sarcoma), which indicates the improvements in target coverage for the PTV (right gluteus) and OAR sparing for the cauda equina, rectum, right femur, bladder, and small bowel. This illustrates the general improvements made by VMAT‐MCO over IMRT‐MCO in terms of dose sparing. Figure 2 shows a dose distribution comparison between VMAT‐MCO and IMRT‐MCO for plan 3 (extremity sarcoma), which indicates the improvement in target coverage for the PTV (right arm) and OAR sparing for the humerus.
Table 3

Selected DVH points of interest (POIs) for each plan

Plan # OAR POI IMRT‐MCO VMAT‐MCO
1TibiaV404.9%4.2%
2HumerusV4077.0%67.0%
3HumerusV4045.0%47.0%
4TibiaV4053.0%50.0%
5Rt femurV4026.0%25.5%
6Colon Dmean 1779 cGy1028 cGy
7Rt femur Dmax 5149 cGy5033 cGy
8Lt lungV203.1%0.5%
9Cauda equina Dmax 4048 cGy2410 cGy
10Rectum Dmax 1729 cGy1636 cGy
Figure 1

Dose distribution comparison for plan 9 (PTV right gluteus to 5000 cGy).

Figure 2

Dose distribution comparison for plan 3 (PTV right arm to 5000 cGy).

Average plan results for IMRT‐MCO and VMAT‐MCO plans Scaled to Rx Dose of 5000 cGy. Selected DVH points of interest (POIs) for each plan Plan delivery time showed statistically significant improvements of 213% on average for VMAT‐MCO plans (average delivery time of s) as compared to IMRT‐MCO plans (average delivery time of seconds), as illustrated in Fig. 3(a). Monitor unit number also showed statistically significant improvements for VMAT‐MCO plans (average of monitor units) as compared to IMRT‐MCO (average of monitor units), as seen in Fig. 3(b). VMAT‐MCO plan deliverability was confirmed and dosimetric measurements showed a slight improvement in gamma passing rate average for VMAT‐MCO () as compared to IMRT‐MCO (). Individual plan results for plan quality and treatment efficiency can be found in Tables 4 and 5, respectively.
Figure 3

Delivery time (a) and plan MU (b) comparisons between VMAT‐MCO and IMRT‐MCO plans.

Table 4

Overview of homogeneity, conformity, and dose‐sparing comparisons between VMAT‐MCO and IMRT‐MCO plans

IMRT‐MCO VMAT‐MCO
Plan # HI CI Dmean (cGy) D50 (cGy) HI CI Dmean (cGy) D50 (cGy)
10.940.97193915120.940.9819891656
20.930.96710240.930.9471028
30.960.89372320.960.921964
40.840.82848230.910.8977625
50.890.9625620.910.96107252
60.920.9614293950.930.991282190
70.950.8414074440.960.811419438
80.940.92621430.940.97655133
90.860.7115996350.860.741494475
100.930.863991090.940.8728573
Mean 0.92±0.04 0.89±0.08 0.93±0.03 0.91±0.08
Table 5

Comparison of delivery time and output for IMRT‐MCO and VMAT‐MCO

IMRT‐MCO VMAT‐MCO
Plan Time (s) Output (MU) Time (s) Output (MU)
1497595160384
2339576121317
3421455160409
4441487132398
5493547130406
6402612198709
7545513220418
83961075184586
97971206172506
108541527186704
Mean 519±173 759±370 166±32 488±136
Dose distribution comparison for plan 9 (PTV right gluteus to 5000 cGy). Dose distribution comparison for plan 3 (PTV right arm to 5000 cGy). Delivery time (a) and plan MU (b) comparisons between VMAT‐MCO and IMRT‐MCO plans. Overview of homogeneity, conformity, and dose‐sparing comparisons between VMAT‐MCO and IMRT‐MCO plans Comparison of delivery time and output for IMRT‐MCO and VMAT‐MCO

IV. DISCUSSION

We successfully generated a set of 10 clinical VMAT‐MCO sarcoma plans, and subsequently evaluated and compared the cohort to paired IMRT‐MCO plans in terms of plan quality and treatment efficiency. The optimization and individualization afforded by MCO provides high‐quality IMRT sarcoma plans, and it is vital to confirm that plan quality is maintained using VMAT‐MCO before transitioning to a VMAT‐MCO standard in the clinic. This study evaluated the VMAT‐MCO algorithm for the complex sarcoma site; to our knowledge VMAT‐MCO has only been studied dosimetrically for prostate. Non‐MCO VMAT has been tested and commissioned on complex sites for other TPS (e.g., Varian RapidArc), but inherent algorithmic and delivery differences between systems require a VMAT‐MCO‐specific validation in order to proceed with necessary clinical confidence in treatment. This same degree of confidence was required for MCO‐IMRT studies. In theory, the increase in gantry angle freedom for VMAT compared to IMRT allows for a more uniform spread of skin dose, and improved dose conformity and homogeneity in some cases. In this study we confirm VMAT‐MCO has comparable plan quality and OAR sparing to IMRT‐MCO for sarcoma plans. We found dose sparing to the skin to be improved in VMAT‐MCO plans, and target coverage was maintained and at times improved. Dose sparing to other OARs could not be directly intercompared between modalities for statistical significance due to the variation of relevant structures by plan. VMAT‐MCO body sarcoma treatments showed more pronounced improvements in dose sparing over IMRT‐MCO treatments in the body sarcoma subset, which is due to the higher degree of OAR overlap in more centrally located targets. For example, the more uniform distribution of dose permitted by VMAT's increased gantry degree freedom allowed for significantly improved sparing in V20 for the left lung in plan 8; the corresponding IMRT plan solution space was confined to only eight gantry angles which required depositing a higher dose to a larger volume of the proximal lung in order to achieve prescription dose in the PTV. While the actual patient benefit of homogeneity and conformity remains largely unproven, it was an objective in both IMRT‐MCO and VMAT‐MCO optimization, and thus represents a relevant comparison metric. The most significant improvements made using VMAT‐MCO are the decreases in monitor units () and treatment delivery time () which are made possible by the reduced “beam‐on” time of VMAT as compared to IMRT. Furthermore, the dynamic nature of VMAT delivery helps overcome the trade‐off between treatment time and treatment plan quality which can impede IMRT treatment efficiency. The decrease in monitor units per patient plan for VMAT extends linac life span, conserves energy, and has been shown to reduce the likelihood of secondary cancers. The decrease in delivery time minimizes patient discomfort, reduces the chances of patient movement during treatment, and allows for more patients to be treated. We tested the gamma passing rate in order to confirm quality assurance for each plan. While VMAT shows slightly improved results over IMRT, the uncertainty of the metrics’ utility disqualifies it from being used conclusively in intercomparison. RayStation has resolved issues with full‐fledged implementation of MCO for VMAT planning, which permits the extension of VMAT‐MCO planning to complex treatment sites. We first examined VMAT‐MCO for sarcomas and plan to undergo a similarly exhaustive commissioning process on other sites. The exciting developments made in VMAT‐MCO require detailed, site‐specific commissioning, and we successfully validated the clinical feasibility and utility of VMAT‐MCO in lieu of IMRT‐MCO. The results of this study confirm the improvements in treatment efficiency and dosimetric quality afforded by VMAT‐MCO planning for complex treatment sites.

V. CONCLUSION

We compared dosimetric quality and treatment efficiency between the VMAT‐MCO and IMRT‐MCO planning modules in RayStation for 10 sarcoma cancer patients receiving standard fractionation treatment. This study highlights the advantages in treatment efficiency and uncompromised plan quality made possible by VMAT‐MCO as compared to IMRT‐MCO.

COPYRIGHT

This work is licensed under a Creative Commons Attribution 3.0 Unported License.
  18 in total

1.  The tradeoff between treatment plan quality and required number of monitor units in intensity-modulated radiotherapy.

Authors:  David Craft; Philipp Süss; Thomas Bortfeld
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-04-01       Impact factor: 7.038

2.  Commissioning of volumetric modulated arc therapy (VMAT).

Authors:  James L Bedford; Alan P Warrington
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-02-01       Impact factor: 7.038

3.  Evaluating IMRT and VMAT dose accuracy: practical examples of failure to detect systematic errors when applying a commonly used metric and action levels.

Authors:  Benjamin E Nelms; Maria F Chan; Geneviève Jarry; Matthieu Lemire; John Lowden; Carnell Hampton; Vladimir Feygelman
Journal:  Med Phys       Date:  2013-11       Impact factor: 4.071

4.  A comparison of the gamma index analysis in various commercial IMRT/VMAT QA systems.

Authors:  Mohammad Hussein; Pejman Rowshanfarzad; Martin A Ebert; Andrew Nisbet; Catharine H Clark
Journal:  Radiother Oncol       Date:  2013-10-04       Impact factor: 6.280

5.  How many plans are needed in an IMRT multi-objective plan database?

Authors:  David Craft; Thomas Bortfeld
Journal:  Phys Med Biol       Date:  2008-05-01       Impact factor: 3.609

6.  Plan averaging for multicriteria navigation of sliding window IMRT and VMAT.

Authors:  David Craft; Dávid Papp; Jan Unkelbach
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

7.  Comparison of Elekta VMAT with helical tomotherapy and fixed field IMRT: plan quality, delivery efficiency and accuracy.

Authors:  Min Rao; Wensha Yang; Fan Chen; Ke Sheng; Jinsong Ye; Vivek Mehta; David Shepard; Daliang Cao
Journal:  Med Phys       Date:  2010-03       Impact factor: 4.071

8.  Multicriteria optimization in intensity-modulated radiation therapy treatment planning for locally advanced cancer of the pancreatic head.

Authors:  Theodore S Hong; David L Craft; Fredrik Carlsson; Thomas R Bortfeld
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-11-15       Impact factor: 7.038

9.  Evaluation of a new VMAT QA device, or the "X" and "O" array geometries.

Authors:  Vladimir Feygelman; Geoffrey Zhang; Craig Stevens; Benjamin E Nelms
Journal:  J Appl Clin Med Phys       Date:  2011-01-31       Impact factor: 2.102

10.  Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator.

Authors:  James L Bedford; Michael D R Thomas; Gregory Smyth
Journal:  J Appl Clin Med Phys       Date:  2013-05-06       Impact factor: 2.102

View more
  1 in total

1.  Planning comparison of five automated treatment planning solutions for locally advanced head and neck cancer.

Authors:  J Krayenbuehl; M Zamburlini; S Ghandour; M Pachoud; S Tanadini-Lang; J Tol; M Guckenberger; W F A R Verbakel
Journal:  Radiat Oncol       Date:  2018-09-10       Impact factor: 3.481

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.