Shirin Sakhdari1, Ahmad Reza Talaeipour2, Maziar Talaeipour3, Maryam Pazhutan4, Sanaz Heidarkhan Tehrani4, Mohammad Javad Kharazifard5.
Abstract
OBJECTIVES: This study sought to assess the diagnostic accuracy of cone beam computed tomography (CBCT) with different voxel sizes and intraoral digital radiography with photostimulable phosphor (PSP) plate for detection of periapical (PA) bone lesions.Entities:
Keywords: Cone-Beam Computed Tomography; Diagnosis; Periapical Periodontitis; Radiography, Dental, Digital
Year: 2016 PMID: 27928235 PMCID: PMC5139934
Source DB: PubMed Journal: J Dent (Tehran) ISSN: 1735-2150
Fig. 1:(A) Radiograph of bone blocks using parallel digital sensor radiography; (B) Specimens in a Plexiglas cylinder in CBCT system
Fig. 2:(A) Axial section; (B) cross section and (C) sagittal section of blocks containing teeth. Images of all three sections of teeth were observed and evaluated simultaneously by the observers.
Calculation of kappa and weighted kappa statistics for assessment of inter-observer and intra-observer reliability
| PSP | 0.361 | 0.57 | 0.87 | 0.90 |
| 200μCBCT | 0.42 | 0.58 | 0.86 | 0.89 |
| 250μCBCT | 0.45 | 0.59 | 0.79 | 0.87 |
| 300μCBCT | 0.35 | 0.49 | 0.84 | 0.88 |
PSP: Photostimulable phosphor plate, CBCT: Cone beam computed tomography
The mean complete and absolute sensitivity and specificity for detection of periapical bone defects on the four types of images
| PSP | 44.4±6.2 | 15±17.5 | 74.9±2.1 | 51.6±6.3 |
| 200μ CBCT | 55.58±23 | 30±13.8 | 87.8±3.8 | 73.33±12.1 |
| 250μ CBCT | 36.1±16.6 | 19.9±9.4 | 83.88±5.2 | 63.3±15.8 |
| 300μ CBCT | 31.1±21.6 | 16.65±6.6 | 72.8±4.5 | 46.63±23.08 |
PSP: Photostimulable phosphor plate, CBCT: Cone beam computed tomography
The mean complete and absolute NPV and PPV for detection of periapical bone defects on the four types of images
| PSP | 44.4±6.2 | 15±17.5 | 74.9±2.1 | 51.6±6.3 |
| 200μ CBCT | 55.58±23 | 30±13.8 | 87.8±3.8 | 73.33±12.1 |
| 250μ CBCT | 36.1±16.6 | 19.9±9.4 | 83.88±5.2 | 63.3±15.8 |
| 300μ CBCT | 31.1±21.6 | 16.65±6.6 | 72.8±4.5 | 46.63±23.08 |
PSP: Photostimulable phosphor plate, CBCT: Cone beam computed tomography