The prediction of radiofrequency ablation zone volume using vascular indices of 3-dimensional volumetric colour Doppler ultrasound in an in vitro blood-perfused bovine liver model.

Samuel Chang1, Anthony C Lanctot1, Martin D McCarter2, Katherine M Roberts3, Deborah H Glueck3, Gerald D Dodd1.   

Abstract

OBJECTIVE: To determine the most reliable predictor of radiofrequency (RF) ablation zone volume among three-dimensional (3D) volumetric colour Doppler vascular indices in an in vitro blood-perfused bovine liver model.
METHODS: 3D colour Doppler volume data of the local hepatic parenchyma were acquired from 37 areas of 13 bovine livers connected to an in vitro oxygenated blood perfusion system. Doppler vascular indices of vascularization index (VI), flow index (FI) and vascularization flow index (VFI) were obtained from the volume data using 3D volume analysis software. 37 RF ablations were performed at the same locations where the ultrasound data were obtained from. The relationship of these vascular indices and the ablation zone volumes measured from gross specimens were analyzed using a general linear mixed model fit with random effect for liver and backward stepwise regression analysis.
RESULTS: FI was significantly associated with ablation zone volumes measured on gross specimens (p = 0.0047), but explained little of the variance (Rβ2 = 0.21). Ablation zone volume decreased by 0.23 cm3 (95% confidence interval: -0.38, -0.08) for every 1 increase in FI. Neither VI nor VFI was significantly associated with ablation zone volumes (p > 0.05).
CONCLUSION: Although FI was associated with ablation zone volumes, it could not sufficiently explain their variability, limiting its clinical applicability. VI, FI and VFI are not clinically useful in the prediction of RF ablation zone volume in the liver. Advances in knowledge: Despite a significant association of FI with ablation zone volumes, VI, FI and VFI cannot be used for their prediction. Different Doppler vascular indices need to be investigated for clinical use.

Entities:  

Mesh:

Year:  2016        PMID: 27925468      PMCID: PMC5685127          DOI: 10.1259/bjr.20160661

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  14 in total

1.  Assessment of normal fetal liver blood flow using quantitative three-dimensional power Doppler ultrasound.

Authors:  Chiung-Hsin Chang; Chen-Hsiang Yu; Huei-Chen Ko; Fong-Ming Chang; Hsi-Yao Chen
Journal:  Ultrasound Med Biol       Date:  2003-07       Impact factor: 2.998

2.  Quantitative tissue blood flow measurement of the liver parenchyma: comparison between xenon CT and perfusion CT.

Authors:  Kazuhiko Hashimoto; Takamichi Murakami; Keizo Dono; Masatoshi Hori; Tonsok Kim; Masayuki Kudo; Shigeru Marubashi; Atsushi Miyamoto; Yutaka Takeda; Hiroaki Nagano; Koji Umeshita; Hironobu Nakamura; Morito Monden
Journal:  Dig Dis Sci       Date:  2007-02-23       Impact factor: 3.199

3.  Radiofrequency ablation of the liver: effect of variation of portal venous blood flow on lesion size in an in-vitro perfused bovine liver.

Authors:  Liping Huang; Gerald D Dodd; Anthony C Lanctot
Journal:  Acad Radiol       Date:  2012-05-15       Impact factor: 3.173

4.  Radiofrequency ablation of porcine liver in vivo: effects of blood flow and treatment time on lesion size.

Authors:  E J Patterson; C H Scudamore; D A Owen; A G Nagy; A K Buczkowski
Journal:  Ann Surg       Date:  1998-04       Impact factor: 12.969

5.  In vitro blood-perfused bovine liver model: a physiologic model for evaluation of the performance of radiofrequency ablation devices.

Authors:  Michael D Orsi; Gerald D Dodd; Rex A Cardan; Hayden W Head; Stephanie K Burns; Dana Blakemore; Jeffrey Blume; Tyler J Green
Journal:  J Vasc Interv Radiol       Date:  2011-07-28       Impact factor: 3.464

6.  Effect of vascular occlusion on radiofrequency ablation of the liver: results in a porcine model.

Authors:  S B Chinn; F T Lee; G D Kennedy; C Chinn; C D Johnson; T C Winter; T F Warner; D M Mahvi
Journal:  AJR Am J Roentgenol       Date:  2001-03       Impact factor: 3.959

7.  Three-dimensional Doppler evaluation of single spherical samples from the placenta: intra- and interobserver reliability.

Authors:  W P Martins; J C Lima; A W Welsh; E Araujo Júnior; E Araujo; A H Miyague; F M Filho; N J Raine-Fenning
Journal:  Ultrasound Obstet Gynecol       Date:  2012-08       Impact factor: 7.299

8.  Influence of large peritumoral vessels on outcome of radiofrequency ablation of liver tumors.

Authors:  David S K Lu; Steven S Raman; Piyaporn Limanond; Donya Aziz; James Economou; Ronald Busuttil; James Sayre
Journal:  J Vasc Interv Radiol       Date:  2003-10       Impact factor: 3.464

9.  Radiofrequency ablation of hepatic tumors: variability of lesion size using a single ablation device.

Authors:  Richard S Montgomery; Andres Rahal; Gerald D Dodd; John R Leyendecker; Linda G Hubbard
Journal:  AJR Am J Roentgenol       Date:  2004-03       Impact factor: 3.959

10.  Radiofrequency ablation of the porcine liver with complete hepatic vascular occlusion.

Authors:  C K Chang; Mary Pat Hendy; J Michael Smith; Matthew H Recht; Richard E Welling
Journal:  Ann Surg Oncol       Date:  2002-07       Impact factor: 5.344

View more
  1 in total

1.  Contrast-enhanced ultrasound to ultrasound fusion during microwave ablation: feasibility study in a perfused porcine liver model.

Authors:  Wayne Monsky; Christina Keravnou; Michalakis Averkiou
Journal:  J Ultrasound       Date:  2019-02-27
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.