Abstract
AIM: To evaluate the agreement of multichannel intraluminal impedance-pH monitoring (MII-pHM) and gastroesophageal reflux scintigraphy (GES) for the diagnosis of gastroesophageal reflux disease.Entities:
Keywords: Children; Esophageal pH monitoring; Gastroesophageal reflux disease; Multichannel intraluminal impedance; Scintigraphy
Mesh:
Substances:
Year: 2016 PMID: 27920480 PMCID: PMC5116603 DOI: 10.3748/wjg.v22.i43.9595
Source DB: PubMed Journal: World J Gastroenterol ISSN: 1007-9327 Impact factor: 5.742
Comparison of pH monitoring and gastroesophageal reflux scintigraphy
| pH monitoring positive | 25 | 9 | 34 |
| pH monitoring negative | 22 | 4 | 26 |
| Total | 47 | 13 | 60 |
Kappa value= -0.118 [P = 0.302, 95%CI: -0.232-(-0.004)]; Overall percent agreement = 48.3%; When pH monitoring is considered as reference test, positive percent agreement = 73.5% and negative percent agreement = 15.4%.
Comparison of multichannel intraluminal impedance monitoring and gastroesophageal reflux scintigraphy
| MII monitoring positive | 34 | 10 | 44 |
| MII monitoring negative | 13 | 3 | 16 |
| Total | 47 | 13 | 60 |
Kappa value = -0.042 (P = 0.741, 95%CI: -0.17-0.086); Overall percent agreement = 61.7%; When MII monitoring is considered as a reference test, positive percent agreement = 77.3% and negative percent agreement = 18.8%. MII: Multichannel intraluminal impedance.
Comparison of multichannel intraluminal impedance/pH monitoring and gastroesophageal reflux scintigraphy
| MII-pHM positive | 41 | 10 | 51 |
| MII-pHM negative | 6 | 3 | 9 |
| Total | 47 | 13 | 60 |
Kappa value 0.116 (P = 0.357, 95%CI: -0.01-0.24); Overall percent agreement = 73.3%; When MII-pHM is considered as reference test, positive percent agreement = 80.4% and negative percent agreement = 33.3%. MII-pHM: Multichannel intraluminal impedance-pH monitoring.
Comparison of multichannel intraluminal impedance and pH monitoring
| MII monitoring positive | 27 | 17 | 44 |
| MII monitoring negative | 7 | 9 | 16 |
| Total | 34 | 26 | 60 |
Kappa = 0.147 (P = 0.223, 95%CI: 0.026-0.268). Overall percent agreement = 60%; MII: Multichannel intraluminal impedance.
Sensitivity, specificity, positive and negative predictive values of diagnostic tests according to assumed gold standard test
| MII-pHM | - | - | - | - | 87.2 | 23.1 | 80.4 | 33.3 |
| (73.6-94.7) | (6.2-54.0) | (66.5-89.7) | (9.0-69.1) | |||||
| pHM alone | 66.7 | 100 | 100 | 34.6 | 53.2 | 30.8 | 73.5 | 15.4 |
| (52.0-78.9) | (62.9-100.0) | (87.4-100.0) | (17.9-55.6) | (38.2-67.6) | (10.4-61.1) | (55.3-86.5) | (5.0-35.7) | |
| MII alone | 73.3 | 100 | 100 | 56.3 | 72.3 | 23.1 | 77.3 | 18.6 |
| (73.1-93.8) | (62.9-100.0) | (90.0-100.0) | (30.6-79.2) | (57.1-83.9) | (6.0-54.0) | (61.8-88.0) | (5.0-46.3) | |
| GES | 80.4 | 33.3 | 87.2 | 23.1 | - | - | - | - |
| (66.5-89.7) | (9.0-69.1) | (73.6-94.7) | (6.1-54.0) | |||||
GES: Gastroesophageal scintigraphy; MII-pHM: Multichannel intraluminal impedance-pH monitoring; pHM: Esophageal pH monitoring; MII: Multichannel intraluminal impedance monitoring; PPV: Positive predictive value; NPV: Negative predictive value.