Karen M Tobias1, Leslie Serrano2, Xiaocun Sun3, Bente Flatland4.
Abstract
BACKGROUND: Preanalytic factors such as time and temperature can have significant effects on laboratory test results. For example, ammonium concentration will increase 31% in blood samples stored at room temperature for 30 min before centrifugation. To reduce preanalytic error, blood samples may be placed in precooled tubes and chilled on ice or in ice water baths; however, the effectiveness of these modalities in cooling blood samples has not been formally evaluated. The purpose of this study was to evaluate the effectiveness of various cooling modalities on reducing temperature of EDTA whole blood samples.Entities:
Keywords: Blood; Cooling; Ice bath; Ice water bath; Laboratory error; Preanalytic; Sample storage; Temperature
Year: 2016 PMID: 27917319 PMCID: PMC5131627 DOI: 10.7717/peerj.2732
Source DB: PubMed Journal: PeerJ ISSN: 2167-8359 Impact factor: 2.984
Summary of treatment comparisons (treatment 1 vs. treatment 2) showing effectiveness of blood sample cooling conditions on the mean temperature drop over 15 min.
| Ice 20% PCV | Ice 40% PCV | P/C 20% PCV | P/C 40% PCV | Water/Ice 20% PCV | Water/Ice 40% PCV | |
|---|---|---|---|---|---|---|
| Ice 20% PCV | – | p = 0.6405 | p < 0.0001 | – | p < 0.0001 | – |
| Ice 40% PCV | p = 0.6405 | – | – | p = 0.7974 | – | p < 0.0001 |
| P/C 20% PCV | p < 0.0001 | – | – | p = 0.1697 | p < 0.0001 | – |
| P/C 40% PCV | – | p = 0.7974 | p = 0.1697 | – | – | p < 0.0001 |
| Water/Ice 20% PCV | p < 0.0001 | – | p < 0.0001 | – | – | p = 0.9983 |
| Water/Ice 40% PCV | – | p < 0.0001 | – | p < 0.0001 | p = 0.9983 | – |
Notes:
Ice, ice bath; P/C, precooled tube in ice bath; water/ice, ice water bath; PCV, packed cell volume.
p value = Tukey’s adjusted probability value, with significance set at p < 0.05.
Mean temperature (°ᵒC) per minute of normal and simulated anemic blood samples for three cooling conditions.
| Time (MIN) | Ice 20% PCV | Ice 40% PCV | Precooled 20% PCV | Precooled 40% PCV | Water/Ice 20% PCV | Water/Ice 40% PCV |
|---|---|---|---|---|---|---|
| 0 | 37.7 | 37.7 | 37.7 | 37.7 | 37.7 | 37.7 |
| 1 | 27.74 | 27.6 | 22.57 | 23.91 | 17.53 | 19.07 |
| 2 | 21.94 | 21.88 | 17.9 | 19.74 | 9.26 | 10.09 |
| 3 | 18.35 | 17.93 | 14.68 | 16.41 | 5.26 | 5.53 |
| 4 | 15.5 | 15.07 | 12.23 | 13.76 | 3.07 | 3.09 |
| 5 | 13.22 | 12.75 | 10.36 | 11.68 | 1.84 | 1.78 |
| 6 | 11.4 | 10.85 | 8.86 | 9.99 | 1.12 | 1.06 |
| 7 | 9.88 | 9.27 | 7.65 | 8.64 | 0.71 | 0.66 |
| 8 | 8.62 | 7.99 | 6.63 | 7.51 | 0.47 | 0.44 |
| 9 | 7.54 | 6.94 | 5.78 | 6.56 | 0.33 | 0.31 |
| 10 | 6.65 | 6.05 | 5.06 | 5.74 | 0.24 | 0.24 |
| 11 | 5.88 | 5.3 | 4.46 | 5.05 | 0.20 | 0.2 |
| 12 | 5.22 | 4.67 | 3.98 | 4.45 | 0.17 | 0.18 |
| 13 | 4.65 | 4.12 | 3.5 | 3.94 | 0.16 | 0.16 |
| 14 | 4.14 | 3.66 | 3.12 | 3.51 | 0.15 | 0.16 |
| 15 | 3.69 | 3.27 | 2.8 | 3.14 | 0.14 | 0.15 |
Note:
PCV, Packed cell volume; MIN, minutes.
Figure 1Mean speed of temperature drop over 15 min (+SEM) under conditions of ice 20%, ice 40%, P/C 20%, P/C 40%, water/ice 20% and water/ice 40% (N = 30 for each treatment group).
Ice, ice bath; P/C, precooled tube in ice bath; Water/Ice, ice water bath.