The evidence based practice for optimal sample quality for ammonia measurement.

Nora Nikolac1, Jelena Omazic2, Ana-Maria Simundic2.   

Abstract

OBJECTIVES: Pre-analytical factors are a major source of variability in laboratory results. Failure to identify these factors can lead to falsely increased or decreased results and to erroneous clinical decisions. We aimed to investigate several pre-analytical factors influencing ammonia measurement, and to quantify their effect on the apparent increase in ammonia concentration. DESIGN AND METHODS: Blood samples were taken from 20 healthy volunteers and submitted to five different sets of conditions: placing sample on ice vs. room temperature immediately after phlebotomy; centrifugation at room temperature vs. 0°C; measurement at 60 min vs. 30 min after sampling; storing sample at room temperature vs. 4°C; and use of stopper (i.e. open vs. closed tube). Ammonia was measured in all samples. Additionally, alanine aminotransferase (ALT), gamma glutamyltransferase (GGT), free hemoglobin, and complete blood count were determined.
RESULTS: Samples placed on ice immediately after centrifugation, samples spun at 0°C, and samples stored at 4°C all had lower changes in ammonia concentration than samples with less favorable treatments (P=0.008; P=0.033, and P=0.001, respectively). The observed biases exceeded clinically relevant acceptance criteria. Most of the tested parameters were significantly associated with increased ammonia. Multiple linear regression analysis identified only three variables that contributed significantly to the prediction of the dependent variable (i.e. increased NH3): ALT, GGT, and storage.
CONCLUSION: Pre-analytical factors cause significant errors in ammonia measurement. An increase in ammonia concentration is most strongly associated with ALT and GGT activity and with storage temperature.
Copyright © 2014 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Ammonia; Centrifugation; Phlebotomy; Preanalytical error; Storage conditions

Mesh:

Substances:

Year:  2014        PMID: 24915632     DOI: 10.1016/j.clinbiochem.2014.05.068

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  5 in total

1.  Clinical utility of breath ammonia for evaluation of ammonia physiology in healthy and cirrhotic adults.

Authors:  Lisa A Spacek; Matthew Mudalel; Frank Tittel; Terence H Risby; Steven F Solga
Journal:  J Breath Res       Date:  2015-12-14       Impact factor: 3.262

2.  The effectiveness of cooling conditions on temperature of canine EDTA whole blood samples.

Authors:  Karen M Tobias; Leslie Serrano; Xiaocun Sun; Bente Flatland
Journal:  PeerJ       Date:  2016-11-29       Impact factor: 2.984

3.  Psychiatric adult-onset of urea cycle disorders: A case-series.

Authors:  Adrien Bigot; Paul Brunault; Christian Lavigne; François Feillet; Sylvie Odent; Elsa Kaphan; Christel Thauvin; Vanessa Leguy; Pierre Broué; Michel C Tchan; François Maillot
Journal:  Mol Genet Metab Rep       Date:  2017-07-06

4.  Ammonium quantification in human plasma by proton nuclear magnetic resonance for staging of liver fibrosis in alcohol-related liver disease and nonalcoholic fatty liver disease.

Authors:  Marc Azagra; Elisa Pose; Francesco De Chiara; Martina Perez; Emma Avitabile; Joan-Marc Servitja; Laura Brugnara; Javier Ramon-Azcón; Irene Marco-Rius
Journal:  NMR Biomed       Date:  2022-05-09       Impact factor: 4.478

5.  Repeated Measures of Blood and Breath Ammonia in Response to Control, Moderate and High Protein Dose in Healthy Men.

Authors:  Lisa A Spacek; Arthur Strzepka; Saurabh Saha; Jonathan Kotula; Jeffrey Gelb; Sarah Guilmain; Terence Risby; Steven F Solga
Journal:  Sci Rep       Date:  2018-02-07       Impact factor: 4.379

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.