Influence Of Inter Electrode Atrial Lead Distance On Acapconfirm™ Viability.

Irigoyen J A Lapuerta1, Andre I Valverde1, Escalera P Vigil1, Gonzalez M Vasserot1.   

Abstract

Introduction: The AcapConfirm™ feature available with the Zephyr pacemaker family (St. Jude Medical) is designed for monitoring patient's atrial capture threshold periodically, and automatically adjusting the atrial pulse amplitude. Previous studies showed a relative low proportion of patients at three months follow-up with recommended automatic atrial capture after the AcapConfirm™ viability test. The purpose of the present study is to evaluate the effect of inter electrode distance on the viability of the AcapConfirm™ algorithm. Methods and
Results: 132 patients (66 woman and 86 men; 71, 08 ± 8, 04 years old) were enrolled into this prospective evaluation. Sixty six bipolar leads (models 1882 (54p) and LPA1200M (12p)) with an inter electrode distance of 10 mm (Group A) were compared with sixty six bipolar leads (model 1999) with an inter electrode distance of 1,1mm (Group B). Set-up test AcapConfirm viability and manual step-down atrial threshold test as well as automatic threshold testing by AcapConfirm™ were performed at 3 months after implantation. A positive viability of the AcapConfirm™ algorithm was much lower in Group B (37, 9%; 95% confidence interval, 10, 3% - 65, 4%) versus thirty two patients (48, 5%; 95% confidence interval, 20, 9% - 76%) in Group A. However, the difference was not statistically significant (χ2=1, 51; p=0, 33). The most frequent reason to reject the AcapConfirm activation was a too small evoked response to polarization ratio (N9).At 3 months, threshold results from the AcapConfirm™ positive test were: 0, 53 ± 0, 13 V in Group B versus 0, 67 ± 0, 18 V in Group A (p< 0, 01). The differences between automatic and manual measurements were ≤0.25V in all patients.
Conclusion: We observed that a short inter electrode distance (1,1mm) is more likely correlated with a lower frequency of AcapConfirm™ viability and threshold that a standard inter electrode distance (10mm). A small evoked response to polarization ratio was the most common cause of a negative test of AcapConfirm™ viability.

Entities:  

Keywords:  Acapconfirm; Automatic Threshold; Interelectrode Distance

Year:  2016        PMID: 27909528      PMCID: PMC5129681          DOI: 10.4022/jafib.1386

Source DB:  PubMed          Journal:  J Atr Fibrillation        ISSN: 1941-6911


  9 in total

1.  Benefits in projected pacemaker longevity and in pacing related costs conferred by automatic threshold tracking.

Authors:  G Boriani; M Biffi; A Branzi; A Mininno; R Sigliano
Journal:  Pacing Clin Electrophysiol       Date:  2000-11       Impact factor: 1.976

2.  Atrial evoked response integral for automatic capture verification in atrial pacing.

Authors:  Giuseppe Boriani; Mauro Biffi; Douglas Cameron; Sergio Datteri; Jeff Snell; Nils Holmström; Euljoon Park; Gene A Bornzin
Journal:  Pacing Clin Electrophysiol       Date:  2003-01       Impact factor: 1.976

3.  Automatic adjustment of pacing output in the clinical setting.

Authors:  Antonio Luiz P Ribeiro; Leonor G Rincón; Bruna G Oliveira; Caio R Vinha; Douglas Melatto; Ana A Torres; Vladimir C V Barros; Paul A Levine
Journal:  Am Heart J       Date:  2004-01       Impact factor: 4.749

4.  Muscle noise effects on atrial evoked response sensing: implications on atrial auto-threshold and auto-capture determination.

Authors:  Hung-Fat Tse; Sunipa Saha; Ankur Garg; Derek Bohn; Yim Ling Lee; Chu-Pak Lau
Journal:  Pacing Clin Electrophysiol       Date:  2011-01-05       Impact factor: 1.976

5.  The measurement of the paced depolarization integral using the braided endocardial lead.

Authors:  R M Lu; A K Dawson
Journal:  Pacing Clin Electrophysiol       Date:  1992-11       Impact factor: 1.976

6.  Autocapture enhancements: unipolar and bipolar lead compatibility and bipolar pacing capability on bipolar leads.

Authors:  Ludwig Binner; John Messenger; Johannes Sperzel; Hansjürgen Bondke; Yves Vandekerckhove; John Poore; Jöerg Scheiner; Marc Berkhof; Euljoon Park; Nils Holmström; Gene A Bornzin
Journal:  Pacing Clin Electrophysiol       Date:  2003-01       Impact factor: 1.976

7.  The impact of automatic threshold tracking on pulse generator longevity in patients with different chronic stimulation thresholds.

Authors:  L Simeon; F Duru; M Fluri; H R Jenzer; M Rahn; R Candinas
Journal:  Pacing Clin Electrophysiol       Date:  2000-11       Impact factor: 1.976

8.  Evaluation of the atrial evoked response for capture detection with high-polarization leads.

Authors:  Johannes Sperzel; Ludwig Binner; Giuseppe Boriani; Mauro Biffi; Jeff Snell; Jörg Scheiner; Euljoon Park; Gene Bornzin
Journal:  Pacing Clin Electrophysiol       Date:  2005-01       Impact factor: 1.976

9.  Reduction of pacing output coupling capacitance for sensing the evoked response.

Authors:  J Sperzel; J Neuzner; T Schwarz; Q Zhu; A König; G N Kay
Journal:  Pacing Clin Electrophysiol       Date:  2001-09       Impact factor: 1.976

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.