| Literature DB >> 27904747 |
Hsi Liu1, Thomas H Taylor2, Kevin Pettus1, Steve Johnson1, John R Papp1, David Trees1.
Abstract
BACKGROUND: We assessed the validity of testing for antimicrobial susceptibility of clinical and mutant Neisseria gonorrhoeae (GC) isolates by disk diffusion in comparison to agar dilution, and Etest® (bioMerieux, France), respectively, for three third generation extended spectrum cephalosporins (ESC): ceftriaxone (CRO), cefixime (CFX), and cefpodoxime (CPD).Entities:
Keywords: Antibiotic; Cefixime; Ceftriaxone; Gonorrhoeae; Neisseria; Susceptibility testing
Year: 2016 PMID: 27904747 PMCID: PMC5122013 DOI: 10.1186/s13756-016-0148-x
Source DB: PubMed Journal: Antimicrob Resist Infect Control ISSN: 2047-2994 Impact factor: 4.887
Categorical Comparisons of the Susceptibility Testing Results of Disk Diffusion with Agar Dilution
| Disk diffusion, zone diameter | Agar dilution, MIC | |||
|---|---|---|---|---|
| Susceptible | Non-Susceptible | |||
| Ceftriaxone (CRO) | <=0.25 ug/ml | >0.25 ug/ml | Totals | |
| Susceptible | > = 35 mm | 97 (84.4%) | 0 (0.00%) | 97 (84.4%) |
| Non- Susceptible | <35 mm | 8 (7.0%) | 10 (8.7%) | 18 (15.6%) |
| Totals | 105 (91.3%) | 10 (8.7%) | 115 (100.0%) | |
| Cefixime (CFX) | <=0.25 ug/ml | >0.25 ug/ml | Totals | |
| Susceptible | > = 31 mm | 94 (81.7%) | 0 (0.0%) | 94 (81.7%) |
| Non-Susceptible | <31 mm | 9 (7.8%) | 12 (10.4%) | 21 (18.3%) |
| Totals | 103 (89.6%) | 12 (10.4%) | 115 (100.0%) | |
| Cefpodoxime (CPD) | <=0.5 μg/ml | >0.5 μg/ml | Totals | |
| Susceptible | > = 29 mm | 58 (50.4%) | 7 (6.1%) | 65 (56.5%) |
| Non-Susceptible | <29 mm | 4 (3.5%) | 46 (40.0%) | 50 (43.5%) |
| Totals | 62 (53.9%) | 53 (46.1%) | 115 (100.0%) | |
Fig. 1Plot of disk-diffusion zone diameter (mm) against agar-dilution minimum inhibition concentration (MIC, μg/ml), with CLSI threshold cutoffs for susceptibility. A total of 115 isolates were tested. a. Ceftriaxon (CRO), b. Cefixime (CFX), c. Cefpodoxime (CPD). x-axis: Agar-dilution (MIC;μg/ml). y-axis: Disk-diffusion zone diameter (mm)
Fig. 2Plot of disk-diffusion zone diameter (mm) against itself at two different points in time, with fitted line and 95% confidence band, for CFX. x-axis: Disk-diffusion zone diameter, time A (mm). y-axis: Repeat Disk-diffusion zone diameter, time B (mm)