G Krasopoulos1, F Falconieri2, U Benedetto1, J Newton1, R Sayeed1, R Kharbanda1, A Banning1. 1. Oxford Heart Centre, Oxford University Hospitals, Headley Way, Headington, Oxford, OX3 9DU, UK. 2. Oxford Heart Centre, Oxford University Hospitals, Headley Way, Headington, Oxford, OX3 9DU, UK. ffalconieri@hotmail.com.
Abstract
OBJECTIVE: Transcatheter aortic valve implantation (TAVI) has been adopted rapidly in Europe. TAVI national registries can augment understanding of technologies and represent real-world experience, providing further clinical insights. We undertook a meta-analysis of published European national TAVI registries to assess current results following TAVI in Europe. METHODS: Electronic databases were searched. The review focused on the comparison of the following TAVI strategies: transfemoral (TF) and transapical (TA) SAPIEN and CoreValve implantation. Individual event rates for outcomes of interest were pooled using a mixed effect model. RESULTS: Seven European national TAVI registries (UK, Swiss, Belgium, Italy, Spain, France, Germany) were identified, including a total of 9786 patients who received TF-SAPIEN (n = 2885), TA-SAPIEN (n = 2252) and CoreValve (n = 4649) implantation. Pooled incidence of 30-day mortality was 0.08% [95% Confidence Interval (CI): 0.05-0.11], 0.12% [95% CI: 0.07-0.19] and 0.06% [95% CI: 0.03-0.11] for TF-SAPIEN, TA-SAPIEN and CoreValve respectively (test for subgroup difference P = 0.18); there was high heterogeneity across European countries. Pooled incidence of stroke was comparable among the TAVI strategies (test for subgroup difference P = 0.79); the incidence of post-procedural moderate paravalvular leak ≥ 2 (P = 0.9) was similar across groups. CoreValve implantation was associated with an increased risk of pacemaker implantation (0.22 [95% CI: 0.19-0.26]; test for subgroup difference P < 0.0001). The lowest 30-day mortality was associated with TAVI performed in Spain (b coefficient -4.3; P = 0.03), in Italy (b coefficient -2.1; P < 0.0001), in UK (b coefficient -1.95; P = 0.01) and in France (b coefficient -2.8; P = 0.03). The German registry has the highest mortality for every TAVI strategy amongst all other European registries and especially for the TA-SAPIEN group. CONCLUSIONS: Transarterial TAVI approaches were associated with a low early mortality regardless of the type of device used. There was marked heterogeneity among European countries for early mortality.
OBJECTIVE: Transcatheter aortic valve implantation (TAVI) has been adopted rapidly in Europe. TAVI national registries can augment understanding of technologies and represent real-world experience, providing further clinical insights. We undertook a meta-analysis of published European national TAVI registries to assess current results following TAVI in Europe. METHODS: Electronic databases were searched. The review focused on the comparison of the following TAVI strategies: transfemoral (TF) and transapical (TA) SAPIEN and CoreValve implantation. Individual event rates for outcomes of interest were pooled using a mixed effect model. RESULTS: Seven European national TAVI registries (UK, Swiss, Belgium, Italy, Spain, France, Germany) were identified, including a total of 9786 patients who received TF-SAPIEN (n = 2885), TA-SAPIEN (n = 2252) and CoreValve (n = 4649) implantation. Pooled incidence of 30-day mortality was 0.08% [95% Confidence Interval (CI): 0.05-0.11], 0.12% [95% CI: 0.07-0.19] and 0.06% [95% CI: 0.03-0.11] for TF-SAPIEN, TA-SAPIEN and CoreValve respectively (test for subgroup difference P = 0.18); there was high heterogeneity across European countries. Pooled incidence of stroke was comparable among the TAVI strategies (test for subgroup difference P = 0.79); the incidence of post-procedural moderate paravalvular leak ≥ 2 (P = 0.9) was similar across groups. CoreValve implantation was associated with an increased risk of pacemaker implantation (0.22 [95% CI: 0.19-0.26]; test for subgroup difference P < 0.0001). The lowest 30-day mortality was associated with TAVI performed in Spain (b coefficient -4.3; P = 0.03), in Italy (b coefficient -2.1; P < 0.0001), in UK (b coefficient -1.95; P = 0.01) and in France (b coefficient -2.8; P = 0.03). The German registry has the highest mortality for every TAVI strategy amongst all other European registries and especially for the TA-SAPIEN group. CONCLUSIONS: Transarterial TAVI approaches were associated with a low early mortality regardless of the type of device used. There was marked heterogeneity among European countries for early mortality.
Dr Alain Cribier reported the first transcatheter aortic valve implantation (TAVI) procedure in 2002 [1]. Following the randomized trials Placement of Aortic Transcatheter Valves (PARTNER) A and B [2, 3], TAVI is now considered the standard of care for symptomatic patients with aortic valve stenosis that are either high-risk or have been turned down for conventional aortic valve replacement [4]. However, even within these groups, the patients enrolled into the trials were highly selected, and therefore may not reflect real-world patients requiring treatment.In Europe patients who undergo TAVI are recorded into national databases which capture the majority of the high risk or inoperable cases, due to co-morbidities and/or frailty patients. Their analysis can therefore provide further insight and evidence into the effectiveness of TAVI in the real-world clinical practice of symptomatic patients with aortic stenosis who are not candidates for conventional aortic valve replacement due to co-morbidities and/or frailty [4]. Furthermore, mixed national registries report country specific results that may be affected by variations in national health policy and local referral practice, device performance and definitions thus accounting for otherwise inexplicable differences in outcome and complications.The aim of this study is to obtain an insight into the role of TAVI in the treatment of high-risk patients with aortic stenosis in Europe by conducting a meta-analysis of European national registries, focusing on the three most commonly used TAVI procedures : the transfemoral and transapical balloon-expandable Edwards SAPIEN transcatheter heart valve (Edwards Lifesciences, Irvine, CA, USA) and the self-expanding Medtronic CoreValve (Medtronic, Minneapolis, MN, USA).
Material and methods
Design
The present review was performed according to the Cochrane Collaboration and PRISMA statements [5].
Eligibility criteria
Inclusion criteria for this meta-analysis were (1) All European national registries reporting outcomes of patients undergoing TAVI; (2) TAVI procedures should have been performed using transfemoral (TF SAPIEN) and/or transapical (TA SAPIEN) balloon-expandable Edwards SAPIEN transcatheter heart valve and/or self-expanding Medtronic CoreValve.Care was taken to ensure that studies selected did not result in duplication of data. Non-English language, review articles, and editorials were excluded. Studies that did not separate results for TF SAPIEN, TA SAPIEN and CoreValve were also excluded.
Search strategy
A literature search was done on the 1st of September 2014 using MEDLINE, EMBASE, and Web of Science to identify relevant articles. Search terms used the controlled vocabularies of MEDLINE and EMBASE alone or in combination with text words including “transcatheter aortic valve implantation”, “TAVI”, “registry”, “Europe”. References from the selected studies were manually searched to identify any other potentially suitable publications.Two reviewers independently screened all studies for inclusion. Disagreements were resolved by consensus. Agreement between reviewers regarding study inclusion was assessed using the Cohen k statistic.
Meta-analysed endpoints
All included studies were interrogated for the following endpoints: 30-day and one year mortality, incidence of stroke, incidence of pacemaker implantation and presence of post deployment moderate paravalvular leak (≥2, according to authors definition).
Statistical analysis
Mixed effects meta-analysis was performed pooling all single registry proportions using the DerSimonian-Laird estimate for all outcomes according to the TAVI strategy used. The I
2 statistic was used to assess the heterogeneity across the reported results. I
2 values of 25 to 49%, 50 to 74%, and 75% or greater were used to indicative low, moderate, and high levels of heterogeneity [6]. Cochrane Q statistic was used as test for subgroup differences (random effects model). The multivariate meta-regression (mixed model) used was used to adjust the effect of different TAVI strategies for the following risk factors: patients risk profile according to mean Logistic Euroscore [7], the reporting European country, the total number of centres involved and the sponsorship of the registry by a TAVI valve manufacturer. R2 was used to estimate the amount of heterogeneity accounted for in the multivariate model. Publication bias was assessed using Begg & Mazumdar test. Trim-and-fill method was used for estimating and adjusting for the number and outcomes of missing studies. A p < 0.05 was used as the level of significance and 95% confidence intervals (95% CI) have been reported where appropriate.R version 3.1.0 (R Core Team (2014 http://www.R-project.org/.) and meta package (Guido Schwarzer (2014) http://CRAN.R-project.org/package=meta) were used for all statistical analyses.
Results
Studies selection
From 1,079 abstracts, 108 full-text articles were assessed for eligibility. After evaluating the full-text articles, 8 met our eligibility criteria and were selected for the systematic review and meta-analysis [8-15]. The studies are summarized in Table 1 and depicted in Fig. 1. A Cohen k statistic of 90% was obtained for the final selection process.
Table 1
European registries overview
Registry
Period
TAVI technology
Number
Logistic EuroSCORE
Funded by manufacturer
N centers
Belgian National TAVI Registry [8]
2009–2010
SAPIEN TF
99
29
no stated
15
SAPIEN TA
88
33
CoreValve
141
25
FRANCE 2 TAVI Registry [9]
2010–2011
SAPIEN TF
1540
22.2
Yes
34
SAPIEN TA
567
24.8
CoreValve
1043
21.3
German TAVI Registry [10]
2009–2010
SAPIEN TF
123
20.3
Yes
22
SAPIEN TA
113
20.3
CoreValve
1074
20.3
Italian Registry TA TAVI [11]
2008–2012
SAPIEN TA
774
25.6
Yes
21
Italian Multicenter CoreValve Registry [12]
2007–2009
CoreValve
659
23
Yes
14
Spain TAVI Registry [13]
2010–2011
SAPIEN TF
504
17
no stated
44
SAPIEN TA
302
19.2
CoreValve
610
16
SWISS TAVI Registry [14]
2011–2013
SAPIEN TF
232
20.2
Yes
8
CoreValve
324
20.2
UK TAVI Registry [15]
2007–2010
SAPIEN TF
387
17.7
Yes
30
SAPIEN TA
408
22.5
CoreValve
798
20.25
TAVI transcatheter aortic valve implantation, TF transfemoral, TA transapical, UK United Kingdom
Fig. 1
Outline of the systematic review process
European registries overviewTAVI transcatheter aortic valve implantation, TF transfemoral, TA transapical, UK United KingdomOutline of the systematic review processSeven European national TAVI registries (Belgium, France, Germany, Italy, Spain, Swiss, UK) were identified including 9786 patients which have received TF SAPIEN (n = 2885), TA SAPIEN (n = 2252) and CoreValve (n = 4649) implantation. All but Italian and Swiss registries reported on all three strategies. Two different and independent Italian TAVI registries reported on TA SAPIEN and CoreValve implantation separately. The Swiss registry reported on TF SAPIEN and CoreValve only. Mean logistic EuroSCORE ranged from 16% (Spain registry, CoreValve group) to 33% (Belgian registry, TA SAPIEN group).
Meta-analysis
30-day mortality
The pooled estimate for 30-day mortality (Fig. 2) was 0.08 [95% CI: 0.05–0.11], 0.12 [95% CI: 0.07–0.19] and 0.06 [95% CI: 0.03–0.11] for TF SAPIEN™, TA SAPIEN and CoreValve respectively (test for subgroup difference P = 0.18). High heterogeneity was present among registries for all three strategies: TF SAPIEN (I2 = 86.6%), TA SAPIEN (I2 = 93.3%) and CoreValve (I2 = 97%). In multivariate meta-regression the increased risk for 30-day mortality was independently associated with TA SAPIEN strategy (b coefficient 0.60; P = 0.001). A higher 30-day mortality was reported from registries with higher number of centres involved into the national TAVI program (b coefficient 0.14; P = 0.03). The lowest 30-day mortality was associated with TAVI performed in Spain (b coefficient −4.3; P = 0.03), in Italy (b coefficient −2.1; P < 0.0001), in UK (b coefficient −1.95; P = 0.01) and in France (b coefficient −2.8; P = 0.03). The German registry has the highest mortality for every TAVI strategy amongst all other European registries and especially for the TA SAPIEN group. Patients risk profile assessed by the mean Logistic EuroSCORE was not associated with observed 30-day mortality (P = 0.9, Fig. 3, left). Moderators included in the multivariate model accounted for 98% amount of heterogeneity with no significant residual heterogeneity (P = 0.25). No publication bias was detected (P = 0.31).
Fig. 2
Meta-analysis for proportions of 30-day mortality
Fig. 3
Bubble plot. Display the adjusted association between logistic EuroSCORE, 30-day mortality (left) and 1-year mortality (right)
Meta-analysis for proportions of 30-day mortalityBubble plot. Display the adjusted association between logistic EuroSCORE, 30-day mortality (left) and 1-year mortality (right)
Stroke
Pooled estimate for the incidence of stroke (Fig. 4) was 0.03 [95% CI: 0.03–0.04], 0.03 [95% CI: 0.02–0.05] and 0.03 [0.02–0.04] for TF SAPIEN™, TA SAPIEN and CoreValve respectively (test for subgroup difference P = 0.79). High heterogeneity was found among registries for TA SAPIEN (I2 = 81.2%) whilst TF SAPIEN and CoreValve were associate with low heterogeneity for stroke incidence (I2 = 0% and I2 = 32.6% respectively). The lowest incidence of stroke was associated with TAVIs performed in Italy (b coefficient −1.5; P = 0.02). Patients risk profile assessed by the mean Logistic EuroSCORE was not associated with the incidence of stroke (P = 0.74). Moderators included in the multivariate model accounted for 82.4% amount of heterogeneity with no significant residual heterogeneity (P = 0.32). No publication bias were detected (P = 0.27).
Fig. 4
Meta-analysis for proportions of stroke
Meta-analysis for proportions of stroke
Need for pacemaker implantation
Pooled estimate for incidence of pacemaker implantation (Fig. 5) was 0.08 [95% CI: 0.05–0.11], 0.07 [95% CI: 0.04–0.11] and 0.22 [95% CI: 0.19–0.26] for TF SAPIEN™, TA SAPIEN and CoreValve respectively (test for subgroup difference P < 0.0001). High heterogeneity was found among registries for TF SAPIEN (I2 = 80.9%), TA SAPIEN (I2 = 88.8%) and CoreValve (I2 = 83%) technologies. TF SAPIEN (b coefficient −1.2; P < 0.0001) and TA SAPIEN (b coefficient −1.08; P = 0.001) were independently associated with the lowest risk of pacemaker implantation. Patients risk profile assessed by the mean Logistic EuroSCORE was not associated with the rate of pacemaker implantation (P = 0.78). Moderators included in the multivariate model accounted for 90.5% amount of heterogeneity but significant residual heterogeneity was found (P = 0.02). No publication bias was detected (P = 0.08).
Fig. 5
Meta-analysis for proportions of pacemaker requirement
Meta-analysis for proportions of pacemaker requirement
Post implantation paravalvular leak
There was no significant difference in the incidence of moderate paravalvular leak (≥2): 0.07 [95% CI: 0.05–0.11], 0.06 [95% CI: 0.05–0.08] and 0.07 [95% CI: 0.04–0.12] for TF SAPIEN™, TA SAPIEN and CoreValve respectively (Fig. 6). (P = 0.9). High heterogeneity was found among registries for TF SAPIEN (I2 = 84.7%) and CoreValve (I2 = 94.4%) but not for TA SAPIEN (I2 = 29.1%). However, there was a significant publication bias (P = 0.002) and trim and fill method suggested an overall prevalence of aortic paravalvular leak of 0.10 [95% CI: 0.0767–0.131].
Fig. 6
Meta-analysis for proportions of Incidence of paravalvular leak ≥ 2
Meta-analysis for proportions of Incidence of paravalvular leak ≥ 2
1 year mortality
Pooled estimate for incidence of 1 year mortality (Fig. 7) was 0.15 [95% CI: 0.14–0.16], 0.23 [95% CI: 0.19–0.28] and 0.17 [95% CI: 0.15–0.19] for TF SAPIEN™, TA SAPIEN and CoreValve respectively (test for subgroup difference P = 0.0008). High heterogeneity was found among registries for TA SAPIEN (I2 = 81.9%) and CoreValve (I2 = 59.3%) but not for TF SAPIEN (I2 = 0%). At multivariate meta-regression, mean Logistic EuroSCORE was moderately associated with all-cause mortality at one year (b coefficient 0.07; P = 0.06, Fig. 3 right). Different TAVI strategies did not impact on the 1-year mortality (TF SAPIEN versus CoreValve: b coefficient −0.1428; P = 0.07) and TA SAPIEN versus CoreValve: b coefficient 0.1515; P = 0.3). Moderators included in the multivariate model accounted for 100% amount of heterogeneity and no significant residual heterogeneity was found (P = 0.48). No publication bias was detected (P = 0.25).
Fig. 7
Meta-analysis for proportions of 1-year mortality
Meta-analysis for proportions of 1-year mortality
Discussion
This meta-analysis of all published European TAVI-registries provides an insight to the real world practice of TAVI in the western world. The results support the superiority of the transarterial approach, regardless of the type of implantation device used. CoreValve was associated with > 2 fold increased risk for pacemaker implantation. There was a marked heterogeneity among European countries for all short-term outcomes investigated. Logistic EuroSCORE, which is widely adopted in Europe to select patients at high-risk for surgical AVR, failed to predict 30-day mortality but it was moderately associated with 1-year mortality.The present meta-analysis found that the overall European TAVI clinical practice has a 30-day mortality of 8% which is higher than the mortality reported by the randomized trials (PARTNER trial A: 3.4% for TF-TAVI, PARTNER trial B: 5% for TF-TAVI and STACCATO trial [16]: 5.8% in TA-TAVI). The inclusion of unselected patients with higher risk profile (mainly severe peripheral disease) in the TA SAPIEN group might partially explain the inferior outcomes and the higher 30-day mortality (12%) for this group.Particularly with regards to the German registry, there is a substantial difference in mortality across all TAVI strategies and in particular in relation to the TA SAPIEN group, which is difficult to explain in isolation. This may be due to the fact that the German registry is reporting the initial period of introduction of the technique which may not be a true representation of the current status. Isolated published series have independently reported lower mortality for TA-TAVI [17, 18].It was also evident that, the more concentrated the program was to a limited number of centres, the lower the reported 30-day mortality was for the corresponded country, suggesting that TAVI should be offered by a restricted and highly specialised centres.Logistic EuroSCORE was not associated with 30-day mortality in the present analysis. It is widely recognized that the logistic EuroSCORE is not an ideal tool for measuring the pre-procedural risk of TAVI as the predicted mortality is grossly overestimated [19]. The present analysis found that early mortality after TAVI varied significantly across European countries, regardless of the type of strategy used. These discrepancies might be partially explained by national differences in patient selection, not accounted for by the logistic EuroSCORE but also by other factors like team integration and learning curve, use of general versus local anaesthesia and level of postoperative care. The presence of heterogeneity in early mortality across European countries highlights the urgent need for standardization of patient selection criteria for TAVI.The incidence of stroke after TAVI in PARTENR A & B and the STACCATO trials was 5.5, 6.7 and 8.8% respectively. Our meta-analysis has revealed the European incidence of stroke after TAVI is 3.0%, regardless of the TAVI strategy used. This result is encouraging considering the risk of peri-operative stroke following surgical AVR in elderly patients, which is ranging from 3 to 7% [20].This analysis has also confirmed that CoreValve is associated with an increased rate of pacemaker implantation when compared to the SAPIEN valve (8%). However, the rate of pacemaker implantation in patients receiving SAPIEN valve was higher than those reported in the PARTNER A (3.8%) and B (3.4%) trials. This could be a reflection of the increased anatomic complexity for TAVI in the real-world practice.This meta-analysis has reported an overall 7% incidence of moderate paravalvular leak (≥2) is better than that the one reported in PARTNER A (12.2%) and B (11.8%) trials. The standardisation of techniques and the increasing familiarity with the procedures with better understanding of deployment strategies may have influence the reduction in the incidence of moderate paravalvular leak, which is the most important complication that negatively influences the long term outcome after TAVI [21]. However, there is an absence of standardized in definition and quantification of paravalvular leak after TAVI (i.e. angiography versus echocardiography, qualitative or semi-quantitative) which needs to be addressed.Finally, the logistic EuroSCORE was the only moderator marginally associated with 1-year mortality, suggesting that mid-term outcome after TAVI is more “patient-related” rather than “procedure-related”.This manuscript has a number of limitations; there is a marked heterogeneity in the reported data, which reflect differences in practice, data collection and data analysis. Patient level data were not disclosed and as such, any attempt to analyse further the observed differences is not possible. We understand and acknowledge that meta-analysis of data from different registries remains inferior to RCT. Finally; there is an inherited lack of surgical comparator that makes it difficult for the reader to formulate clear opinion on the subject.
Conclusions
In conclusion, this meta-analysis of European TAVI registries is reporting higher 30-day mortality than that reported by the published randomized trials, in particular after transapical TAVIs. This result may reflect the inclusion of patients with a higher risk profile. The evidence suggests that transarterial TAVI should be the preferred valve implantation strategy, whenever possible. The significant heterogeneity in early mortality across European countries suggests that there in an urgent need for standardization of patient selection process and procedural aspects in order to optimise outcomes and guarantee high standards of care, across European countries. Despite the fact that in Europe patients are currently selected for TAVI on the basis of their surgical risk measured by the logistic EuroSCORE (≥20%), such a risk stratification system is ineffective in predicting early mortality after TAVI. European registries of TAVI and surgical AVR (with clear differentiation amongst conventional, minimally invasive and suture-less minimally invasive surgical AVR) may be used to generate a well-powered TAVI-weighted risk score.
Authors: F Roques; S A Nashef; P Michel; E Gauducheau; C de Vincentiis; E Baudet; J Cortina; M David; A Faichney; F Gabrielle; E Gams; A Harjula; M T Jones; P P Pintor; R Salamon; L Thulin Journal: Eur J Cardiothorac Surg Date: 1999-06 Impact factor: 4.191
Authors: Corrado Tamburino; Davide Capodanno; Angelo Ramondo; Anna Sonia Petronio; Federica Ettori; Gennaro Santoro; Silvio Klugmann; Francesco Bedogni; Francesco Maisano; Antonio Marzocchi; Arnaldo Poli; David Antoniucci; Massimo Napodano; Marco De Carlo; Claudia Fiorina; Gian Paolo Ussia Journal: Circulation Date: 2011-01-10 Impact factor: 29.690
Authors: Alain Cribier; Helene Eltchaninoff; Assaf Bash; Nicolas Borenstein; Christophe Tron; Fabrice Bauer; Genevieve Derumeaux; Frederic Anselme; François Laborde; Martin B Leon Journal: Circulation Date: 2002-12-10 Impact factor: 29.690
Authors: Manel Sabaté; Sergio Cánovas; Eulogio García; Rosana Hernández Antolín; Luis Maroto; José María Hernández; Juan H Alonso Briales; Antonio J Muñoz García; Enrique Gutiérrez-Ibañes; Jorge Rodríguez-Roda Journal: Rev Esp Cardiol (Engl Ed) Date: 2013-10-04
Authors: Paolo Buja; Massimo Napodano; Corrado Tamburino; Anna Sonia Petronio; Federica Ettori; Gennaro Santoro; Gian Paolo Ussia; Silvio Klugmann; Francesco Bedogni; Angelo Ramondo; Francesco Maisano; Antonio Marzocchi; Arnaldo Poli; Valeria Gasparetto; David Antoniucci; Antonio Colombo; Giuseppe Tarantini Journal: Am J Cardiol Date: 2013-01-01 Impact factor: 2.778
Authors: Josep Rodés-Cabau; John G Webb; Anson Cheung; Jian Ye; Eric Dumont; Christopher M Feindel; Mark Osten; Madhu K Natarajan; James L Velianou; Giuseppe Martucci; Benoît DeVarennes; Robert Chisholm; Mark D Peterson; Samuel V Lichtenstein; Fabian Nietlispach; Daniel Doyle; Robert DeLarochellière; Kevin Teoh; Victor Chu; Adrian Dancea; Kevin Lachapelle; Asim Cheema; David Latter; Eric Horlick Journal: J Am Coll Cardiol Date: 2010-01-22 Impact factor: 24.094
Authors: James M Brown; Sean M O'Brien; Changfu Wu; Jo Ann H Sikora; Bartley P Griffith; James S Gammie Journal: J Thorac Cardiovasc Surg Date: 2009-01 Impact factor: 5.209
Authors: Peter Wenaweser; Stefan Stortecky; Dik Heg; David Tueller; Fabian Nietlispach; Volkmar Falk; Giovanni Pedrazzini; Raban Jeger; Oliver Reuthebuch; Thierry Carrel; Lorenz Räber; Franz W Amann; Enrico Ferrari; Stefan Toggweiler; Stephane Noble; Marco Roffi; Jürg Gruenenfelder; Peter Jüni; Stephan Windecker; Christoph Huber Journal: EuroIntervention Date: 2014-12 Impact factor: 6.534
Authors: Daniel J Blackman; Paul D Baxter; Chris P Gale; Neil E Moat; Philip A Maccarthy; David Hildick-Smith; Uday Trivedi; David Cunningham; Mark A DE Belder; Peter F Ludman Journal: J Interv Cardiol Date: 2013-12-30 Impact factor: 2.279
Authors: Adam S Evans; Menachem M Weiner; Shahzad Shaefi; Prakash A Patel; Matthew M Townsley; Abirami Kumaresan; Jared W Feinman; Ashley V Fritz; Archer K Martin; Toby B Steinberg; J Ross Renew; Jane L Gui; Brian Radvansky; Himani Bhatt; Sudhakar Subramani; Archit Sharma; Jacob T Gutsche; John G Augoustides; Harish Ramakrishna Journal: J Cardiothorac Vasc Anesth Date: 2019-11-09 Impact factor: 2.628
Authors: Gabby Elbaz-Greener; Feng Qiu; John G Webb; Kayley A Henning; Dennis T Ko; Andrew Czarnecki; Idan Roifman; Peter C Austin; Harindra C Wijeysundera Journal: J Am Heart Assoc Date: 2019-06-05 Impact factor: 5.501
Authors: Mauro Chiarito; Alessandro Spirito; Johny Nicolas; Alexandra Selberg; Giulio Stefanini; Antonio Colombo; Bernhard Reimers; Annapoorna Kini; Samin K Sharma; George D Dangas; Roxana Mehran Journal: J Clin Med Date: 2022-07-30 Impact factor: 4.964
Authors: Matthias Lutz; David Messika-Zeitoun; Tanja K Rudolph; Eberhard Schulz; Jeetendra Thambyrajah; Guy Lloyd; Alexander Lauten; Norbert Frey; Jana Kurucova; Martin Thoenes; Cornelia Deutsch; Peter Bramlage; Richard Paul Steeds Journal: Open Heart Date: 2020-09