Linhui Yu1, Ying Wu2, Jeff F Dunn3, Kartikeya Murari4.
Abstract
We propose a single fiber optical system for monitoring tissue oxygen saturation (sO2) based on continuous-wave reflectance spectroscopy in the visible wavelengths. The system is designed for measurements in deep brain structures by stereotaxically implanting the 200 μm-core fiber probe into the tissue of interest. Monte Carlo (MC) simulations were used to estimate the measurement tissue volume between 0.02-0.03 mm3. Experiments in an optical phantom indicated the system had a root mean squared error (RMSE) of 4.21% compared with a commercial fluorescence-based tissue oxygen partial pressure measuring system. Finally, we used the system for continuously monitoring tissue sO2 from a highly-localized volume in anesthetized mice.Entities:
Keywords: (170.1470) Blood or tissue constituent monitoring; (170.3660) Light propagation in tissues; (170.3890) Medical optics instrumentation; (170.6510) Spectroscopy, tissue diagnostics
Year: 2016 PMID: 27896007 PMCID: PMC5119607 DOI: 10.1364/BOE.7.004685