Yasufumi Asao1, Yohei Hashizume2, Takahiro Suita2, Ken-Ichi Nagae2, Kazuhiko Fukutani2, Yoshiaki Sudo2, Toshikazu Matsushita2, Shuichi Kobayashi2, Mariko Tokiwa3, Iku Yamaga3, Elham Fakhrejahani3, Masae Torii3, Masahiro Kawashima3, Masahiro Takada3, Shotaro Kanao4, Masako Kataoka4, Tsuyoshi Shiina5, Masakazu Toi3.
Abstract
We have constructed a prototype photoacoustic mammography system (PAM-02) capable of simultaneously acquiring photoacoustic (PA) and ultrasound (US) images. Each PA, US, and fused PA/US image can be acquired over a wide area of the breast using the scanning module of a US transducer, a PA detector, and optical prisms. The resolution of the PA images exhibits improvement from 2 to 1 mm compared to images acquired using our previous prototype. The maximum scan area of PAM-02 is 90 mm along the horizontal axis and 150 mm along the vertical axis. In a phantom experiment, the available depth was at least 45 mm. A representative example of the application of the PAM-02 prototype in clinical research at Kyoto University is presented and shows S-factor images, which are considered an approximation parameter related to hemoglobin saturation of tumor-related blood vessels. We confirmed the applicability of the system for anatomical and biological research.Entities: