Literature DB >> 27882574

Inter- and intra-operator variability in the analysis of packed cell volume.

C R Breheny1, A Brown1, I Handel1, A G Gow1.   

Abstract

OBJECTIVES: To identify whether inter- and intra-operator variability occurs in the measurement of canine packed cell volume and, if so, at which stage these errors occur.
MATERIALS AND METHODS: Undergraduate veterinary students and veterinary surgeons were recruited to measure the packed cell volumes of three samples in duplicate. Measurements from each sample were confirmed by one author, and it was then ascertained whether the error was made in the capillary preparation or reading.
RESULTS: Data were obtained from 44 students and 11 vets. A total of 25% of students made errors associated with inadequate mixing; 23% students and 9% of vets made errors consistent with incorrect reading. There was also less intra-operator variation in values within the vet group (0·027 from the mean) in comparison to the student group (-0·21 from the mean). A total of 68·2% of students and 91% of vets filled the capillary tubes outwith World Health Organisation standards of two-thirds to three-quarters full. CLINICAL SIGNIFICANCE: Packed cell volume measurement is extremely useful when measuring erythroid mass, but it is crucial that the results upon which decisions are made are accurate and precise in order to manage these cases appropriately. Operator variation is a significant factor and must be addressed by proper training and following standard operating procedures.
© 2016 British Small Animal Veterinary Association.

Entities:  

Mesh:

Year:  2016        PMID: 27882574     DOI: 10.1111/jsap.12603

Source DB:  PubMed          Journal:  J Small Anim Pract        ISSN: 0022-4510            Impact factor:   1.522


  2 in total

1.  Interobserver reliability of canine urine specific gravity assessed by analog or digital refractometers.

Authors:  John C Rowe; Jessica A Hokamp; Jessica N Braatz; John R Freitag-Engstrom; Nicole L Stephens; Dennis J Chew; Catherine Langston; Adam J Rudinsky
Journal:  J Vet Diagn Invest       Date:  2021-03-01       Impact factor: 1.279

2.  Evaluation of a flash glucose monitoring system in nondiabetic dogs with rapidly changing blood glucose concentrations.

Authors:  Leigh A Howard; Jonathan A Lidbury; Nicholas Jeffery; Shannon E Washburn; Carly A Patterson
Journal:  J Vet Intern Med       Date:  2021-10-02       Impact factor: 3.333

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.