Measuring dose from radiotherapy treatments in the vicinity of a cardiac pacemaker.

Samuel C Peet1, Rachael Wilks2, Tanya Kairn3, Scott B Crowe4.   

Abstract

This study investigated the dose absorbed by tissues surrounding artificial cardiac pacemakers during external beam radiotherapy procedures. The usefulness of out-of-field reference data, treatment planning systems, and skin dose measurements to estimate the dose in the vicinity of a pacemaker was also examined. Measurements were performed by installing a pacemaker onto an anthropomorphic phantom, and using radiochromic film and optically stimulated luminescence dosimeters to measure the dose in the vicinity of the device during the delivery of square fields and clinical treatment plans. It was found that the dose delivered in the vicinity of the cardiac device was unevenly distributed both laterally and anteroposteriorly. As the device was moved distally from the square field, the dose dropped exponentially, in line with out-of-field reference data in the literature. Treatment planning systems were found to substantially underestimate the dose for volumetric modulated arc therapy, helical tomotherapy, and 3D conformal treatments. The skin dose was observed to be either greater or lesser than the dose received at the depth of the device, depending on the treatment site, and so care should be if skin dose measurements are to be used to estimate the dose to a pacemaker. Square field reference data may be used as an upper estimate of absorbed dose per monitor unit in the vicinity of a cardiac device for complex treatments involving multiple gantry angles.
Copyright © 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Keywords:  Cardiac device; Out-of-field dose; Pacemaker; Radiotherapy

Mesh:

Year:  2016        PMID: 27881295     DOI: 10.1016/j.ejmp.2016.11.010

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  2 in total

1.  PAIDEIA: pacemaker and implanted cardioverter defibrillator management in radiation therapy-a survey by the Young Group of the Italian Association of Radiotherapy and Clinical Oncology (AIRO).

Authors:  Anna Rita Alitto; Silvia Chiesa; Pierfrancesco Franco; Michele Fiore; Lorenza Marino; Paolo Borghetti; Isacco Desideri; Daniela Greto; Alba Fiorentino
Journal:  Radiol Med       Date:  2019-12-12       Impact factor: 3.469

2.  The effect of stereotactic body radiotherapy (SBRT) using flattening filter-free beams on cardiac implantable electronic devices (CIEDs) in clinical situations.

Authors:  Hossein Aslian; Tomas Kron; Troy Watts; Cagla Akalanli; Nicholas Hardcastle; Peta Lonski; Atousa Montaseri; Barry Hay; James Korte; Kemal Berk; Francesco Longo; Mara Severgnini
Journal:  J Appl Clin Med Phys       Date:  2020-04-11       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.