A novel needle-type biosensor based on carbon nanotubes is reported. Thebiosensor was prepared by packing a mixture of multi-wall carbon nanotubes (MWCNTs),graphite powder and glucose oxidase (Gox) freeze-dried powder into a glass capillary of 0.5mm inner diameter. The resulting amperometric biosensor was characterizedelectrochemically using amperometry in the presence of hydrogen peroxide and in thepresence of glucose. The glucose biosensor sensitivity was influenced by the glucoseoxidase concentration within the MWCNTs mixture. The optimized glucose needle-typebiosensor displayed better sensitivity and stability, and a detected range of up to 20 mM.Based on its favorable stability, the needle biosensor was first time used in real-timemonitoring system as a kind of online glucose detector. The decay of current response isless than 10% after 24-hour continuous observation.Entities:
Keywords: Multiwall carbon nanotubes; glucose; needle-type biosensor
Year: 2008 PMID: 27879788 PMCID: PMC3663019 DOI: 10.3390/s8031712
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1.Influence of Gox contents upon the amperometric glucose response. ▲-MWCNTs mixture/Gox needle biosensor at 5:1 composition ratio, ●- MWCNTs mixture/Gox needle biosensor at 5:3 composition ratio, ■- MWCNTs mixture/Gox needle biosensor at 5:5 composition ratio.
Figure 2.Storage stability of needle-type biosensor. ▲- MWCNTs mixture/Gox needle-type biosensor (5:1 composition ratio), ●-MWCNTs mixture/Gox needle-type biosensor (5:3 composition ratio), ■- MWCNTs mixture/Gox needle-type biosensor (5:5 composition ratio). The data points represent biosensor's current response to 20 mM glucose, pH 7.0 PBS, measured for 100 second at 0.60 V on a given day. The response was normalized with respect to that on day 1.
Figure 3.24-hour current response of the needle-type biosensor. The needle-type MWCNTs mixtue/Gox biosensor used in this experiment has 5:3 composition ratio of MWCNTs mixtue and Gox.