| Literature DB >> 27871969 |
V E Ozden1, N Saglam2, G Dikmen3, I R Tozun3.
Abstract
Oxidized zirconium (Oxinium™; Smith & Nephew, Memphis, TN, USA) articulated with polyethylene in total hip arthroplasty (THA) appeared to have the potential to reduce wear dramatically. The thermally oxidized metal zirconium surface is transformed into ceramic-like hard surface that is resistant to abrasion. The exposure of soft zirconium metal under hard coverage surface after the damage of oxidized zirconium femoral head has been described. It occurred following joint dislocation or in situ succeeding disengagement of polyethylene liner. We reported three cases of misuse of Oxinium™ (Smith & Nephew, Memphis, TN, USA) heads. These three cases resulted in catastrophic in situ wear and inevitable failure although there was no advice, indication or recommendation for this use from the manufacturer.Entities:
Keywords: Hip arthroplasty; Oxidized zirconium on ceramic; Oxinium™
Mesh:
Substances:
Year: 2016 PMID: 27871969 DOI: 10.1016/j.otsr.2016.10.004
Source DB: PubMed Journal: Orthop Traumatol Surg Res ISSN: 1877-0568 Impact factor: 2.256