Intraocular pressure measurement after DSAEK by iCare, Goldmann applanation and dynamic contour tonometry: A comparative study.

A Achiron1, O Blumenfeld2, H Avizemer2, L Karmona2, G Leybowich2, V Man3, E Bartov2, Z Burgansky-Eliash2.   

Abstract

PURPOSE: Corneal thickness inevitably increases following Descemet's stripping automated endothelial keratoplasty (DSAEK), owing to the addition of a donor graft. The current study compares different devices in assessing post-DSAEK intraocular pressure (IOP).
METHODS: We compared IOP values measured by the Goldmann tonometry (GAT), iCare rebound tonometry (iCare) and Pascal dynamic contour tonometry (PDCT) in eyes following DSAEK. Agreement between measurements was calculated with correlation analysis and Bland-Altman plots. Effects of keratometry, central, thickness (CCT), endothelial cell density (ECD) and axial length on IOP measurements were assessed with Pearson's correlation.
RESULTS: Twenty eyes of 20 patients (mean age 74.3±14.4, 14 females) post-DSAEK were included in this study. There was a high concordance between the IOP readings obtained by the three devices: a strong and significant correlation was found between GAT and PDCT (r=0.94, P<0.001) GAT and iCare (r=0.86, P<0.001) and iCare with PDCT (r=0.81, P<0.001). However, the iCare measurements were significantly and consistently lower than that obtained with GAT (ΔIOP=1.68±2.0, P=0.002, 95% CI: 0.7-2.6) and with PDCT (ΔIOP=1.61±2.5, P=0.01, 95% CI: 0.4-2.8). CCT, ECD, CCT, AXL, corneal curvature or astigmatism did not influence IOP measurement by any instrument.
CONCLUSIONS: IOP measurement with three different techniques (applanation, rebound and dynamic contour) showed good correlations, despite an increased corneal thickness following DSAEK. However, the iCare, which is based on a rebound tonometry showed significant lower IOP then the two other methods. This should be taken into account when evaluating patients post DSAEK. Copyright Â
© 2016 Elsevier Masson SAS. All rights reserved.

Entities:  

Keywords:  Aplanation de Goldmann; DSAEK; Dynamic contour tonometry; Goldmann applanation; Intraocular pressure; Pression intraoculaire; Tonomètre dynamique de contour; iCare

Mesh:

Year:  2016        PMID: 27871779     DOI: 10.1016/j.jfo.2016.09.009

Source DB:  PubMed          Journal:  J Fr Ophtalmol        ISSN: 0181-5512            Impact factor:   0.818


  4 in total

1.  Risk Factors for the Development of Ocular Hypertension After Keratoplasty: A Systematic Review.

Authors:  Ilona Liesenborghs; Johannes S A G Schouten; Tos T J M Berendschot; Henny J M Beckers; Rudy M M A Nuijts; Nienke Visser; Carroll A B Webers
Journal:  Cornea       Date:  2020-03       Impact factor: 3.152

Review 2.  Challenges in the complex management of post-keratoplasty glaucoma.

Authors:  Lisa-Marie Anders; Zisis Gatzioufas; Matthias C Grieshaber
Journal:  Ther Adv Ophthalmol       Date:  2021-07-20

3.  Incidence and Risk Factors of Ocular Hypertension/Glaucoma After Descemet Stripping Automated Endothelial Keratoplasty.

Authors:  Mohamed Elalfy; Sundas Maqsood; Shady Soliman; Sherif Momtaz Hegazy; Ahmed Abdou Hannon; Zisis Gatzioufas; Damian Lake; Samer Hamada
Journal:  Clin Ophthalmol       Date:  2021-05-25

4.  Demographic and clinical profile, surgical outcome, and quality of life in patients who underwent bilateral lamellar corneal grafts.

Authors:  Sreelakshmi P Amar; Rajesh Sinha; Nidhi Kalra; Tushar Agarwal; Namrata Sharma; Jeewan S Titiyal
Journal:  Indian J Ophthalmol       Date:  2021-07       Impact factor: 1.848

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.