Literature DB >> 27846630

A First Evaluation of OMNI®, A New Device for Continuous Renal Replacement Therapy.

Pierre Schläpfer1, Jean-Daniel Durovray, Valery Plouhinec, Cristiano Chiappa, Rinaldo Bellomo, Antoine Schneider.   

Abstract

BACKGROUND: Omni® (B. Braun, Germany) is a new-generation, continuous renal replacement therapy (CRRT) machine designed to improve user interface, minimize downtime and optimize renal dose delivery. It was never tested in humans.
METHODS: We used Omni® to provide CRRT in 10 critically ill patients. We collected therapy data, metabolic parameters and evaluated user's satisfaction with a survey.
RESULTS: CRRT was delivered using Omni® in CVVH-heparin (6 patients) and CVVHD-citrate (4 patients) modes for a total duration of 617.7 h. No adverse event was observed. The mean filter life was 22.8 (CVVH-heparin) and 33.5 (CVVHD-citrate) h. Alarms-related downtime corresponded to 5.9% of total therapy time. Delivered renal dose was 96.6% of prescribed. Satisfactory metabolic control and fluid removal were achieved. Overall, users evaluated interface, design and usability as excellent.
CONCLUSION: CRRT in CVVH-heparin and CVVHD-citrate modes was provided using Omni® in a safe and efficient way for 10 critically ill patients. Video Journal Club 'Cappuccino with Claudio Ronco' at http://www.karger.com/?doi=451053.
© 2016 S. Karger AG, Basel.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27846630     DOI: 10.1159/000451053

Source DB:  PubMed          Journal:  Blood Purif        ISSN: 0253-5068            Impact factor:   2.614


  2 in total

1.  ACUsmart Continuous Renal Replacement Therapy Platform: Multicenter Pilot Study for Technical and Clinical Assessment (A.M.P. Study).

Authors:  Mauro Neri; Anna Lorenzin; Massimo de Cal; Alessandra Brendolan; Nicola Marchionna; Sara Samoni; Monica Zanella; Silvia De Rosa; Francesca Martino; Zaccaria Ricci; Javier Maynar; Georg Auzinger; Gianluca Villa; Didier Payen; Michael Joannidis; Claudio Ronco
Journal:  Blood Purif       Date:  2019-02-01       Impact factor: 2.614

Review 2.  Methods for dose quantification in continuous renal replacement therapy: Toward a more precise approach.

Authors:  Gianluca Villa; Sergio Fabbri; Sara Samoni; Matteo Cecchi; Antonio Fioccola; Caterina Scirè-Calabrisotto; Gaia Mari; Diego Pomarè Montin; Stefano Romagnoli
Journal:  Artif Organs       Date:  2021-08-18       Impact factor: 3.094

  2 in total