Literature DB >> 27838945

Development, validity and reliability of a new pressure air biofeedback device (PAB) for measuring isometric extension strength of the lumbar spine.

Andries W Pienaar1, Justhinus G Barnard1.   

Abstract

This study describes the development of a new portable muscle testing device, using air pressure as a biofeedback and strength testing tool. For this purpose, a pressure air biofeedback device (PAB®) was developed to measure and record the isometric extension strength of the lumbar multifidus muscle in asymptomatic and low back pain (LBP) persons. A total of 42 subjects (age 47.58 years, ±18.58) participated in this study. The validity of PAB® was assessed by comparing a selected measure, air pressure force in millibar (mb), to a standard criterion; calibrated weights in kilograms (kg) during day-to-day tests. Furthermore, clinical trial-to-trial and day-to-day tests of maximum voluntary isometric contraction (MVIC) of L5 lumbar multifidus were done to compare air pressure force (mb) to electromyography (EMG) in microvolt (μV) and to measure the reliability of PAB®. A highly significant relationship were found between air pressure output (mb) and calibrated weights (kg). In addition, Pearson correlation calculations showed a significant relationship between PAB® force (mb) and EMG activity (μV) for all subjects (n = 42) examined, as well as for the asymptomatic group (n = 24). No relationship was detected for the LBP group (n = 18). In terms of lumbar extension strength, we found that asymptomatic subjects were significantly stronger than LBP subjects. The results of the PAB® test differentiated between LBP and asymptomatic subject's lumbar isometric extension strength without any risk to the subjects and also indicate that the lumbar isometric extension test with the new PAB® device is reliable and valid.

Entities:  

Keywords:  PAB®; air pressure; isometric strength; low back pain (LBP); reliability

Mesh:

Year:  2016        PMID: 27838945     DOI: 10.1080/03091902.2016.1253794

Source DB:  PubMed          Journal:  J Med Eng Technol        ISSN: 0309-1902


  5 in total

1.  The reliability and validity of a designed setup for the assessment of static back extensor force and endurance in older women with and without hyperkyphosis.

Authors:  Taybeh Roghani; Minoo Khalkhali Zavieh; Abbas Rahimi; Saeed Talebian; Farideh Dehghan Manshadi; Alireza Akbarzadeh Baghban; Nicole King; Wendy Katzman
Journal:  Physiother Theory Pract       Date:  2018-01-25       Impact factor: 2.279

2.  Practicable performance-based outcome measures of trunk muscle strength and their measurement properties: A systematic review and narrative synthesis.

Authors:  Shouq Althobaiti; Alison Rushton; Ahmad Aldahas; Deborah Falla; Nicola R Heneghan
Journal:  PLoS One       Date:  2022-06-17       Impact factor: 3.752

3.  Inter-rater reliability of pressure biofeedback unit among individuals with and without chronic low back pain.

Authors:  Muhammad Khan; Hamayun Zafar; Syed Amir Gilani
Journal:  Pak J Med Sci       Date:  2022 Mar-Apr       Impact factor: 2.340

4.  Dynamics of Changes in Isometric Strength and Muscle Imbalance in the Treatment of Women with Low back Pain.

Authors:  Jacek Wilczyński; Alicja Kasprzak
Journal:  Biomed Res Int       Date:  2020-01-11       Impact factor: 3.411

5.  Muscle Strength and Neuromuscular Control in Low-Back Pain: Elite Athletes Versus General Population.

Authors:  María Moreno Catalá; Arno Schroll; Gunnar Laube; Adamantios Arampatzis
Journal:  Front Neurosci       Date:  2018-07-03       Impact factor: 4.677

  5 in total