The impact of patient table on size-specific dose estimate (SSDE).

Choirul Anam1,2, Freddy Haryanto3, Rena Widita3, Idam Arif3, Geoff Dougherty4, Donald McLean5.   

Abstract

The use of parameters water equivalent diameter (D W ) and size-specific dose estimate (SSDE) are becoming increasingly established as a recognised method to relate patient dose from a CT examination to the dose indicator volume CT dose index (CTDIVOL). However, the role of the attenuation due to the patient table in these estimations requires careful consideration and is the subject of this study. The aim of this study is to investigate the impact of a minimal part of the patient table when calculating the D W and SSDE. We investigated 164 patients who had undergone CT examinations for the pelvis, abdomen, thorax and head. We subsequently calculated D W and SSDE using two methods: one using a small circular region of interest (ROI) including a minimal part of the patient table and the other using a ROI fitted to the patient border alone. The results showed that the water equivalent diameter calculated with the table included in the ROI (D W,t ) is greater, compared to that without the consideration of the patient table (D W,nt ), by 1.5-6.2% depending on the anatomy being imaged. On the other hand, the SSDE calculated with inclusion of the patient table (SSDEt) is smaller than otherwise (SSDEnt) by 1.0-5.5% again depending on the anatomy being imaged. The effect of the patient table on D W and SSDE in the thorax CT examination was statistically significant, but its effect on D W and SSDE in the other examinations of head, pelvis and abdomen was relatively small and not statistically significant.

Entities:  

Keywords:  Patient dose; Patient table; Size-specific dose estimates (SSDE); Volume CT dose index (CTDIVOL); Water equivalent diameter (D W )

Mesh:

Substances:

Year:  2016        PMID: 27832459     DOI: 10.1007/s13246-016-0497-z

Source DB:  PubMed          Journal:  Australas Phys Eng Sci Med        ISSN: 0158-9938            Impact factor:   1.430


  6 in total

1.  Multi-slice CT examinations of adult patients at Sudanese hospitals: radiation exposure based on size-specific dose estimates (SSDE).

Authors:  Einas H Bashier; I I Suliman
Journal:  Radiol Med       Date:  2018-01-30       Impact factor: 3.469

2.  Evaluation of AAPM Reports 204 and 220: Estimation of effective diameter, water-equivalent diameter, and ellipticity ratios for chest, abdomen, pelvis, and head CT scans.

Authors:  Christiane S Burton; Timothy P Szczykutowicz
Journal:  J Appl Clin Med Phys       Date:  2017-11-27       Impact factor: 2.102

3.  Using a Somatosensory Controller to Assess Body Size for Size-Specific Dose Estimates in Computed Tomography.

Authors:  Jay Wu; Ruo-Ping Han; Yan-Lin Liu
Journal:  Biomed Res Int       Date:  2018-05-31       Impact factor: 3.411

4.  Comparison of Application Value of Different Radiation Dose Evaluation Methods in Evaluating Radiation Dose of Adult Thoracic and Abdominal CT Scan.

Authors:  Jimin He; Guanwei Dong; Yi Deng; Jun He; ZhiGang Xiu; Fanzi Feng
Journal:  Front Surg       Date:  2022-03-25

5.  An algorithm for automated modulation transfer function measurement using an edge of a PMMA phantom: Impact of field of view on spatial resolution of CT images.

Authors:  Choirul Anam; Toshioh Fujibuchi; Wahyu Setia Budi; Freddy Haryanto; Geoff Dougherty
Journal:  J Appl Clin Med Phys       Date:  2018-10-19       Impact factor: 2.102

6.  Comparison of central, peripheral, and weighted size-specific dose in CT.

Authors:  Choirul Anam; Dwi Adhianto; Heri Sutanto; Kusworo Adi; Mohd Hanafi Ali; William Ian Duncombe Rae; Toshioh Fujibuchi; Geoff Dougherty
Journal:  J Xray Sci Technol       Date:  2020       Impact factor: 1.535

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.