| Literature DB >> 27818728 |
Hyunkyung Kang1, Hang Sik Roh1, Hyemin Song1, Kwangmoon Lee1, Seung-Tae Chung1, Sang-Ja Ban1, In Pil Mo2, Beum-Soo An3, Chi-Young Ahn1.
Abstract
The potency of influenza vaccine is determined based on its hemagglutinin (HA) content. In general, single radial immunodiffusion (SRID) assay has been utilized as the standard method to measure HA content. However, preparation of reagents for SRID such as antigen and antibody takes approximately 2~3 months, which causes delays in the development of influenza vaccine. Therefore, quantification of HA content by other alternative methods is required. In this study, we measured HA contents of H1N1 antigen and H1N1 influenza vaccine by reverse phase-high performance liquid chromatography (RP-HPLC) methods. The presence of HA1 and HA2 was investigated by silver staining and Western blot assay. In addition, accuracy and repeatability of HA measurement by RP-HPLC were evaluated. Comparison of HA concentration by SRID and RP-HPLC revealed a precise correlation between the two methods. Our results suggest that RP-HPLC assay can replace SRID in the event of a pandemic flu outbreak for rapid vaccine development.Entities:
Keywords: Hemagglutinin; Influenza; RP- HPLC; Vaccine
Year: 2016 PMID: 27818728 PMCID: PMC5080855 DOI: 10.5487/TR.2016.32.4.269
Source DB: PubMed Journal: Toxicol Res ISSN: 1976-8257
Fig. 1Chromatograms of influenza H1N1 antigen and seasonal influenza monovalent H1N1 vaccine. RP-HPLC chromatograms of H1N1 antigens (A) and H1N1 vaccine (B) are represented. Fractions from RP-HPLC chromatograms were designated with numbers according to the eluted time.
Fig. 2Silver staining (A and C) and Western blot (B and D) of RP-HPLC fractions from standard H1N1 antigen (A and B) and seasonal influenza H1N1 vaccine (C and D). Elutes separated from RP-HPLC chromatograms were subjected to silver staining and Western blot. Specific H1N1 anti-HA serum was used to label HA protein.
Fig. 3Plots of calibration represent HA1 peak areas from H1N1 antigen (A) and seasonal influenza H1N1 vaccine (B). Different amounts of HA1 in a concentration-dependent manner were processed for RP-HPLC, and the peak area from 12 min was analyzed. The peak area from RP-HPLC was highly correlated with the concentration of HA1.
Repeatability of the HA1 peak area and retention time
| No. | Area | Retention time (min) | |
|---|---|---|---|
| H1N1 antigens | 1 | 207907 | 12.02 |
| 2 | 209547 | 12.03 | |
| 3 | 204499 | 12.03 | |
| 4 | 208902 | 12.04 | |
| 5 | 205727 | 12.05 | |
| 6 | 209614 | 12.04 | |
| Average | 207316 | 12.03 | |
| RSD (%) | 1.03 | 0.08 | |
|
| |||
| H1N1 vaccines | 1 | 183050 | 12.05 |
| 2 | 183960 | 12.05 | |
| 3 | 188573 | 12.05 | |
| 4 | 192404 | 12.04 | |
| 5 | 189547 | 12.03 | |
| 6 | 193271 | 12.05 | |
| Average | 188468 | 12.05 | |
| RSD (%) | 2.24 | 0.06 | |
RSD: Relative Standard Deviation.
Precision within laboratory
| Vaccine (lot) | Amount of HA (1st) | Amount of HA (2nd) | Amount of HA (3rd) | Average | RSD (%) |
|---|---|---|---|---|---|
| v121092 | 426.5 ± 3.6 | 400.7 ± 4.3 | 422.58 ± 8.2 | 416.6 ± 13.0 | 3.1 |
| v121093 | 439.5 ± 1.8 | 415.3 ± 2.2 | 419.64 ± 5.3 | 424.8 ± 11.5 | 2.7 |
| v121094 | 425.4 ± 1.9 | 393.8 ± 2.9 | 415.48 ± 5.8 | 411.6 ± 14.2 | 3.4 |
| v121096 | 401.6 ± 6.9 | 360.9 ± 7.8 | 397.86 ± 2.1 | 368.8 ± 20.1 | 5.2 |
| v121097 | 407.1 ± 4.7 | 381.1 ± 2.1 | 398.54 ± 3.6 | 395.6 ± 11.8 | 2.9 |
| v121098 | 394.1 ± 1.4 | 361.6 ± 2.2 | 380.82 ± 4.5 | 378.8 ± 14.2 | 3.7 |
| v121099 | 374.2 ± 2.2 | 363.8 ± 3.6 | 362.13 ± 4.9 | 366.7 ± 6.6 | 1.8 |
| v121100 | 396.6 ± 3.0 | 379.3 ± 4.7 | 415.95 ± 2.8 | 397.3 ± 16.0 | 4.0 |
| v121101 | 403.0 ± 1.6 | 379.5 ± 2.4 | 422.95 ± 4.5 | 401.8 ± 18.7 | 4.6 |
| v121106 | 379.7 ± 3.8 | 359.4 ± 3.8 | 390.68 ± 4.7 | 376.6 ± 14.1 | 3.7 |
| v121107 | 394.7 ± 3.3 | 372.1 ± 4.1 | 407.09 ± 4.2 | 391.3 ± 15.5 | 3.9 |
| v121108 | 396.0 ± 6.3 | 372.2 ± 3.9 | 413.79 ± 2.9 | 394.0 ± 18.3 | 4.6 |
RSD: Relative Standard Deviation.
HA content results obtained by SRID and RP-HPLC methods from seasonal influenza H1N1 vaccines
| Lot No | HA1 by RP HPLC | HA1 by SRID | RP HPLC/SRID |
|---|---|---|---|
| v121092 | 416 ± 13.0 | 223 | 1.87 |
| v121093 | 424.8 ± 11.5 | 239 | 1.78 |
| v121094 | 411.6 ± 14.2 | 229 | 1.80 |
| v121096 | 368.8 ± 20.1 | 206 | 1.79 |
| v121097 | 395.6 ± 11.8 | 215 | 1.84 |
| v121098 | 378.8 ± 14.2 | 207 | 1.83 |
| v121099 | 366.7 ± 6.6 | 211 | 1.74 |
| v121100 | 397.3 ± 16.0 | 211 | 1.88 |
| v121101 | 401.8 ± 18.7 | 224 | 1.79 |
| v121106 | 376.6 ± 14.1 | 195 | 1.93 |
| v121107 | 391.3 ± 15.5 | 202 | 1.94 |
| v121108 | 394.0 ± 18.3 | 198 | 1.99 |
|
| |||
| Average | 393.6 | 213.3 | 1.85 |