PURPOSE: To characterize mass and density changes of lung parenchyma in non-small cell lung cancer (NSCLC) patients following midtreatment resolution of atelectasis and to quantify the impact this large geometric change has on normal tissue dose. METHODS: Baseline and midtreatment CT images and contours were obtained for 18 NSCLC patients with atelectasis. Patients were classified based on atelectasis volume reduction between the two scans as having either full, partial, or no resolution. Relative mass and density changes from baseline to midtreatment were calculated based on voxel intensity and volume for each lung lobe. Patients also had clinical treatment plans available which were used to assess changes in normal tissue dose constraints from baseline to midtreatment. The midtreatment image was rigidly aligned with the baseline scan in two ways: (1) bony anatomy and (2) carina. Treatment parameters (beam apertures, weights, angles, monitor units, etc.) were transferred to each image. Then, dose was recalculated. Typical IMRT dose constraints were evaluated on all images, and the changes from baseline to each midtreatment image were investigated. RESULTS: Atelectatic lobes experienced mean (stdev) mass changes of -2.8% (36.6%), -24.4% (33.0%), and -9.2% (17.5%) and density changes of -66.0% (6.4%), -25.6% (13.6%), and -17.0% (21.1%) for full, partial, and no resolution, respectively. Means (stdev) of dose changes to spinal cord Dmax, esophagus Dmean, and lungs Dmean were 0.67 (2.99), 0.99 (2.69), and 0.50 Gy (2.05 Gy), respectively, for bone alignment and 0.14 (1.80), 0.77 (2.95), and 0.06 Gy (1.71 Gy) for carina alignment. Dose increases with bone alignment up to 10.93, 7.92, and 5.69 Gy were found for maximum spinal cord, mean esophagus, and mean lung doses, respectively, with carina alignment yielding similar values. 44% and 22% of patients had at least one metric change by at least 5 Gy (dose metrics) or 5% (volume metrics) for bone and carina alignments, respectively. Investigation of GTV coverage showed mean (stdev) changes in VRx, Dmax, and Dmin of -5.5% (13.5%), 2.5% (4.2%), and 0.8% (8.9%), respectively, for bone alignment with similar results for carina alignment. CONCLUSIONS: Resolution of atelectasis caused mass and density decreases, on average, and introduced substantial changes in normal tissue dose metrics in a subset of the patient cohort.
PURPOSE: To characterize mass and density changes of lung parenchyma in non-small cell lung cancer (NSCLC) patients following midtreatment resolution of atelectasis and to quantify the impact this large geometric change has on normal tissue dose. METHODS: Baseline and midtreatment CT images and contours were obtained for 18 NSCLCpatients with atelectasis. Patients were classified based on atelectasis volume reduction between the two scans as having either full, partial, or no resolution. Relative mass and density changes from baseline to midtreatment were calculated based on voxel intensity and volume for each lung lobe. Patients also had clinical treatment plans available which were used to assess changes in normal tissue dose constraints from baseline to midtreatment. The midtreatment image was rigidly aligned with the baseline scan in two ways: (1) bony anatomy and (2) carina. Treatment parameters (beam apertures, weights, angles, monitor units, etc.) were transferred to each image. Then, dose was recalculated. Typical IMRT dose constraints were evaluated on all images, and the changes from baseline to each midtreatment image were investigated. RESULTS:Atelectatic lobes experienced mean (stdev) mass changes of -2.8% (36.6%), -24.4% (33.0%), and -9.2% (17.5%) and density changes of -66.0% (6.4%), -25.6% (13.6%), and -17.0% (21.1%) for full, partial, and no resolution, respectively. Means (stdev) of dose changes to spinal cord Dmax, esophagus Dmean, and lungs Dmean were 0.67 (2.99), 0.99 (2.69), and 0.50 Gy (2.05 Gy), respectively, for bone alignment and 0.14 (1.80), 0.77 (2.95), and 0.06 Gy (1.71 Gy) for carina alignment. Dose increases with bone alignment up to 10.93, 7.92, and 5.69 Gy were found for maximum spinal cord, mean esophagus, and mean lung doses, respectively, with carina alignment yielding similar values. 44% and 22% of patients had at least one metric change by at least 5 Gy (dose metrics) or 5% (volume metrics) for bone and carina alignments, respectively. Investigation of GTV coverage showed mean (stdev) changes in VRx, Dmax, and Dmin of -5.5% (13.5%), 2.5% (4.2%), and 0.8% (8.9%), respectively, for bone alignment with similar results for carina alignment. CONCLUSIONS: Resolution of atelectasis caused mass and density decreases, on average, and introduced substantial changes in normal tissue dose metrics in a subset of the patient cohort.
Obstructive lobar atelectasis, the collapse of lung
tissue due to
restricted airflow, commonly occurs in nonsmall cell lung
cancer
(NSCLC) patients
with centrally located tumors. Initial
atelectasis presentation rates for patients undergoing external beam radiotherapy have
been reported to range between 10% and 40%. During the course of radiotherapy, tumor regression or
progression can cause changes in atelectasis, either resolution or expansion. Studies
investigating anatomical variations during treatment found atelectasis changes in 10% to
30% of all NSCLCpatients.Resolution of atelectasis, in the case of full reaeration of whole lobe collapse,
appears in computed
tomography
(CT) scans as a
change from a uniform, high-intensity consolidated volume to a larger, lower-intensity
region of normal parenchyma. This can produce large geometric changes in treatment
anatomy which can cause baseline shifts in tumor position. These large geometric changes impact dose to the target and
critical structures and cannot be handled by treatment margins, instead requiring plan
adaptation. Anatomical variations
have been shown to have a greater impact on target dose than either respiratory
motion or baseline shifts (e.g., setup errors), highlighting the potential need for
adaptive radiotherapy in patients with atelectasis changes.Little has been reported about the characteristics of atelectasis changes during
radiotherapy, partly due to a lack of diagnostic-quality imaging during
treatment. The aforementioned studies relied on followup CTs taken months or years
after treatment or cone-beam
CT scans
which have relatively poor contrast resolution and electron density inaccuracies which make clear
identification of the atelectatic regions challenging. The purpose of this study is to
quantitatively characterize mass and density changes of obstructive lobar atelectasis
during treatment in NSCLCpatients using weekly helical CTs and to investigate the
dosimetric
impact of such changes on normal tissue structures in lieu of adaptive replanning. It is anticipated
that studying these mass changes will provide a better understanding of how atelectasis
resolution impacts dose calculation and image registration in a sizable subset of NSCLCpatients.
MATERIALS AND METHODS
Data acquisition
Pairs of baseline and midtreatment CT scans for eighteen patients were acquired on a
CT
simulator (Philips Brilliance Big Bore, Fitchburg, WI) under IRB-approved protocols.
The baseline scan was taken at or near the time of simulation, and the midtreatment
scan was taken during the course of radiotherapy, with a mean (stdev) of 46 (12) days
between the two scans. Five patients (28%) underwent breath hold scans, while the
remainder (72%) had free breathing 4DCT acquisitions. Four of the five breath hold
patients also had repeat scans taken during each weekly session, making available
three images at all time points. For all but four patients, the
planning CT
was used as the baseline scan. The remaining four patients had 4DCT planning
images, but breath hold midtreatment imaging. For
consistency, a breath hold image acquired close to the time of planning
CT
acquisition was used as the baseline scan. The 50% phase (end-of-expiration) of each
4D image was selected for use in this study as it is considered to
have minimum tissue motion and therefore likely to have the least sorting and
motion artifacts. Both images of every pair were of the same scan type. Voxel size
ranged from 1.17 × 1.17 × 2 mm3 to 1.37 × 1.37 × 3 mm3.
Images of each pair had identical voxel size for all but two
patients.Tumor staging
varied across patients between IB (5%), IIB (5%), IIIA (50%), and IIIB (40%) with a
mean (stdev) baseline tumor volume of 109.6 (89.2) ml across patients. Tumor locations were as
follows: left upper lobe (LUL) (17%), left lower lobe (LLL) (22%), right upper lobe
(RUL) (22%), right middle lobe (RML) (6%), and right lower lobe (RLL) (50%).
Treatment technique varied between 3D conformal and IMRT for the study cohort.
Dose was
delivered with conventional fractionation of 1.8–2.0 Gy per fraction, with a mean
(stdev) prescription dose of 63.2 (5.0) Gy. Initial collapse type was scored based on
whether the whole lobe (WL) (56%) or part of the lobe (PL) (44%), aside from
tumor, was
considered to be atelectatic. Patients had a mean (stdev) baseline atelectasis volume
of 232.42 (181.55) ml. 17% of patients experienced collapse in the LUL, 17% in the
LLL, 17% in the RUL, 11% in the RML, and 50% in the RLL. For some patients, the
tumor
and/or atelectasis was present in multiple lung lobes.
Atelectasis classification
Subjects were classified according to resolution of atelectatic tissue from baseline to
followup. Classification was based on the change in atelectasis volume: greater than
80% volume reduction was labeled as full (22%), between 80% and 20% volume reduction
was labeled as partial (50%), and a decrease in volume less than either 20% or 15 ml
was labeled as no resolution (28%).
Lobe segmentation
Accurate delineation of the boundary of atelectasis can be challenging, particularly
in cases where the atelectasis may be partially resolved. In addition to contouring
the atelectasis directly, individual lung lobes were delineated to reduce the impact of
delineation error since the lobe boundaries are generally more clearly defined. All
fives lobes (right upper, right middle, right lower, left upper, and left lower) were
delineated by individuals (NJ, LR) trained by an experienced radiation oncologist (EW) using a
commercial radiation oncology software suite (MIM Maestro v6.6.4, Cleveland, OH). All
delineations were reviewed by the same oncologist for accuracy and consistency. The
tumor,
atelectasis, and all five lung lobes were contoured in each image. For
patients with repeat scans during each weekly session, the contours for each scan
were drawn independently.Contours were converted to binary masks from which the tumor was removed, as
tumor
regression is known to occur and is not the focus of the study. To remove nonlung
tissue
inadvertently included in the lobe delineations, a combined lung binary mask was
created which was then eroded by 1 voxel in all dimensions a total of two times.
Eroded lobe masks were obtained by taking the union of the original lobe mask and the
eroded combined lung mask, effectively eroding the individual lobe from only the
exterior of the lung. An example of eroded lobe masks for one subject is shown in
Fig. 1.
FIG. 1.
Lobe segmentation example. Lobe labels for patient 6 are shown for (a)
baseline, where the lower right lobe (blue) is fully collapsed, and (b)
followup lobes, where the right lower lobe has fully reaerated revealing
healthy lung parenchyma and vessel structures. The right lower lobe
expands to fill the anterior pleural cavity, pushing the right middle lobe
(cyan) posterior to the shown coronal slice. Back-to-back 3D erosions of 1
voxel from the exterior of the lungs were performed to exclude extra-pleural
tissue. The gross tumor volume has
also been removed from the lobe masks.
Lobe segmentation example. Lobe labels for patient 6 are shown for (a)
baseline, where the lower right lobe (blue) is fully collapsed, and (b)
followup lobes, where the right lower lobe has fully reaerated revealing
healthy lung parenchyma and vessel structures. The right lower lobe
expands to fill the anterior pleural cavity, pushing the right middle lobe
(cyan) posterior to the shown coronal slice. Back-to-back 3D erosions of 1
voxel from the exterior of the lungs were performed to exclude extra-pleural
tissue. The gross tumor volume has
also been removed from the lobe masks.
Mass and density calculation
Images
were preprocessed to enable mass and density calculation. Variability in scanner
performance over time was removed by linearly calibrating the images according
to the average intensities of air outside the body and blood in the descending aorta,
as described by Staring et al. The voxel values of the calibrated images were then
in units of relative physical density, with air at approximately 0 mg/cc and blood at
about 1050 mg/cc. Relative density was calculated by averaging calibrated voxel
values in the region of interest, while relative mass was obtained by summing the
product of voxel intensity and voxel volume throughout the region of interest.
Relative mass and density change from baseline to midtreatment were calculated for
all lobes of each patient. For the four patients with repeat weekly scans, each scan
at the baseline time point was paired with a scan of the followup time point forming
three image pairs per patient for which mass and density changes were
calculated.
Treatment planning
Clinical treatment plans created with Philips pinnacle treatment planning
system (TPS) (Philips Radiation Oncology Systems, Fitchburg, WI) were available for
all patients. To investigate the dosimetric impact of atelectasis reaeration,
dose was
calculated on both the baseline and midtreatment scans using the exact clinical
treatment parameters (beam angles, beam weights, MLC segments, etc.). In cases where
the baseline image was not the clinical planning CT, the treatment plan
was first transferred to the baseline scan via rigid registration of bony anatomy
using the image fusion tools of MIM.
Midtreatment alignment
Two methods were used to align the baseline plan to the midtreatment
image in order to simulate different methods of daily patient
setup. In method one (bone-aligned), the midtreatment image was
rigidly registered to the baseline scan based on bony anatomy, and the resulting
fusion was adjusted with box-based alignment of a region covering the sternum and
spine. In the second method (carina-aligned), the default whole-body fusion was
performed again, but with subsequent manual translational adjustment to align the
carina region, mimicking carina-based volumetric image-guided
setup.
Dosimetric evaluation
Using the clinical beams (weights, apertures, and angles) and monitor units,
dose
was calculated and evaluated for the baseline plan, the bony-aligned midtreatment
plan, and the carina-aligned midtreatment plan. It is important to note that this
method assumes atelectasis resolution occurs before the first treatment fraction
and represents a worst-case estimate of dose changes. To quantify the dosimetric differences
in plan quality, a combination of RTOG 0617 and in-house normal tissue constraints was
evaluated (see Table I for the complete
list). Lungs were defined in three different ways: all lung
tissue,
lung
tissue
without gross tumor volume (GTV), and lung
tissue
without clinical target volume (CTV) in order to manage the effect of
tumor
regression on dose change assessment. The lung and GTV
delineations were made on each scan separately. The CTV was rigidly transferred
from the clinical planning image to baseline and followup scans after
image alignment was performed. Inclusion of lungs and
tumor
together is expected to minimize the influence of tumor regression on
dose
changes, while removal of the GTV and CTV from the lung is expected to
incorporate this effect.
TABLE I.
Dose
constraint metric changes.
Bone aligned
Carina aligned
Structure
Metric
Units
Mean
Stdev
Min
Max
Mean
Stdev
Min
Max
Spinal cord
Dmax
Gy
0.67
2.99
−2.78
10.93
0.14
1.80
−2.94
4.29
Esophagus
Dmean
Gy
0.99
2.69
−3.72
7.92
0.77
2.95
−4.56
7.07
Heart
V40
% Vol.
1.64
5.64
−16.15
9.72
1.59
3.62
−10.51
6.28
Heart
V60
% Vol.
0.96
2.14
−2.87
4.38
0.83
1.59
−3.58
2.64
Lungs
Dmean
Gy
0.50
2.05
−2.89
5.69
0.06
1.71
−3.35
4.56
Lungs
V20
% Vol.
1.17
3.42
−3.22
11.31
0.28
2.56
−4.15
6.55
Lungs
V30
% Vol.
0.87
2.86
−3.59
6.44
0.19
2.55
−4.22
6.97
Lungs-CTV
Dmean
Gy
0.61
1.72
−1.80
5.61
0.18
1.31
−1.87
4.02
Lungs-CTV
V20
% Vol.
1.40
3.14
−2.87
11.43
0.49
2.15
−2.99
5.86
Lungs-CTV
V30
% Vol.
1.11
2.41
−2.66
6.42
0.41
2.04
−2.94
6.17
Lungs-GTV
Dmean
Gy
0.99
2.23
−1.70
5.94
0.55
2.09
−1.83
7.39
Lungs-GTV
V20
% Vol.
1.90
3.59
−2.72
11.35
1.00
3.09
−2.83
10.56
Lungs-GTV
V30
% Vol.
1.65
3.24
−2.50
9.35
0.97
3.19
−2.77
11.34
Note: Shown are changes in dose metrics from baseline to followup
for bone and carina followup alignments. Mean and standard deviation were
taken over all subjects. Also shown are the largest increases and
decreases in metric values experienced by any subject.
Dose
constraint metric changes.Note: Shown are changes in dose metrics from baseline to followup
for bone and carina followup alignments. Mean and standard deviation were
taken over all subjects. Also shown are the largest increases and
decreases in metric values experienced by any subject.
Analysis
Analyses were performed with R 3.2.1. To determine if mass and density changes
differed between healthy lobes (contralateral and pathology-free ipsilateral) and
atelectatic lobes, a Wilcoxon rank sum test was used. An F-test was used to determine
if variance in dose changes was significantly different between patients showing
no resolution and those experiencing partial or full resolution. All tests used a
0.95 confidence level and were unpaired and two-sided. Additionally, mean and
standard deviation of all changes were calculated.
RESULTS
Mass change
Changes in lobe mass from baseline to followup, as a percentage of the baseline
value, are shown in Fig. 2. The mean (stdev) of
mass change for all healthy contralateral (n = 41) and healthy
ipsilateral (n = 29) lobes was −3.7% (12.2%) and 0.0% (23.0%),
respectively. There was no significant difference between the two healthy lobe groups
(p = 1). For atelectatic lobes, changes of −2.8% (36.6%), −24.4%
(33.0%), and −9.2% (17.5%) were found for full resolution (n = 4),
partial resolution (n = 9), and no resolution (n =
5) cases, respectively. Mass change was not significantly different from healthy
lobes for full resolution (p = 0.9) or no resolution
(p = 0.4) lobes. However, partial resolution mass change showed
significant difference from that of healthy lobes (p = 0.005). For
patients with multiple scans per session, intrapatient standard deviation of mass
change taken across the three image pairs was 4.7% for healthy lobes and
3.5% for atelectatic lobes, on average.
FIG. 2.
Box plots of percent change in relative mass from baseline to followup are
shown for atelectatic lobes (left), healthy ipsilateral lobes (center), and
contralateral lobes (right). Lobes containing atelectasis are subdivided by
resolution type.
Box plots of percent change in relative mass from baseline to followup are
shown for atelectatic lobes (left), healthy ipsilateral lobes (center), and
contralateral lobes (right). Lobes containing atelectasis are subdivided by
resolution type.
Density change
Changes in lobe density are shown in Fig. 3.
Lobes containing atelectasis experienced changes in density, from baseline to
followup, of −66.0% (6.5%), −25.6% (13.6%), and −17.0% (21.1%) for full, partial, and
no resolution, respectively. Density changes for healthy ipsilateral and
contralateral lobes were −3.5% (23.3%) and −5.2% (12.0%), respectively. There was no
significant difference in density change between healthy ipsilateral and
contralateral (p = 0.9) or between no resolution and healthy lobes
(p = 0.3). Significant differences were present between full
resolution and healthy lobes (p = 0.0008) and partial resolution and
healthy lobes (p = 0.0006). Among the four patients with multiple
image pairs, average intrapatient standard deviation of density
change was 3.5% for healthy lobes and 0.6% for atelectatic lobes.
FIG. 3.
Box plots of percent change in relative density from baseline to followup are
shown for atelectatic lobes (left), healthy ipsilateral lobes (center), and
contralateral lobes (right). Lobes containing atelectasis are subdivided by
resolution type.
Box plots of percent change in relative density from baseline to followup are
shown for atelectatic lobes (left), healthy ipsilateral lobes (center), and
contralateral lobes (right). Lobes containing atelectasis are subdivided by
resolution type.
Volume change
Changes in lobe volume are shown in Fig. 4.
Lobes containing atelectasis experienced changes in volume, from baseline to
followup, of +206% (167%), +22.5% (79.1%), and +14.1% (29.5%) for full, partial, and
no resolution, respectively. Volume changes for healthy ipsilateral and contralateral
lobes were +8.6% (33.1%) and +3.7% (21.6%), respectively. There was no significant
difference in volume change between healthy ipsilateral and contralateral
(p = 0.8), between no resolution and healthy lobes
(p = 0.5), or between partial resolution and healthy lobes
(p = 0.8). A significant difference in mean volume change was
present between full resolution and healthy lobes (p = 0.001).
FIG. 4.
Box plots of percent change in relative volume from baseline to followup are
shown for atelectatic lobes (left), healthy ipsilateral lobes (center), and
contralateral lobes (right). Lobes containing atelectasis are subdivided by
resolution type.
Box plots of percent change in relative volume from baseline to followup are
shown for atelectatic lobes (left), healthy ipsilateral lobes (center), and
contralateral lobes (right). Lobes containing atelectasis are subdivided by
resolution type.
Organ at risk dose
Dose and
volume changes from baseline to followup were analyzed across all patients using
available treatment plans. Changes in dose constraint metrics are shown in Table I. While mean dose changes were less
than the typical dose per fraction of 2 Gy, very large dose and volume changes
occurred for a subset of patients. In particular, dose increases from
baseline to followup of up to 10.93, 7.92, and 5.69 Gy were found for maximum spinal
cord dose,
mean esophagus dose, and mean lung-GTV dose, respectively, when
the subject was aligned via bone. Maximum changes were slightly reduced for carina
alignment. Histograms of changes found for each dose metric are shown in
Fig. 5 for both followup alignments. Across all
patients, the percentage of changes exceeding 1 Gy/1%, 2 Gy/2%, 5 Gy/5%, and 10
Gy/10% were 63%, 38%, 12%, and 2%, respectively, for bone alignment and 62%, 28%, 5%,
and 1% for carina alignment. The number of patients with at least one change larger
than 5 Gy/5% was 44% for bone alignment and 22% for carina alignment.
FIG. 5.
Histograms of change in dose and volume metrics from baseline to followup are shown
for bone and carina alignment for all evaluated dose constraints.
Bin widths were set to 1 Gy for dose metrics and 2% for volume constraints. The
red dashed box surrounds heart constraints, while the blue dotted box
encompasses lung constraints.
Histograms of change in dose and volume metrics from baseline to followup are shown
for bone and carina alignment for all evaluated dose constraints.
Bin widths were set to 1 Gy for dose metrics and 2% for volume constraints. The
red dashed box surrounds heart constraints, while the blue dotted box
encompasses lung constraints.Change in constraint metric value relative to baseline value was assessed in relation
to atelectasis resolution type. The mean (stdev) relative changes in metrics across
all patients for full, partial, and no resolution were 7.9% (18.6%), 18.6% (63.2%),
and 1.9% (11.5%) for bone alignment, respectively. Deviations were similar for carina
alignment with mean (stdev) relative changes of 13.0% (19.1%), 5.5% (27.8%), and
−1.0% (10.6%) for full, partial, and no resolution, respectively. There was a
significant difference between variances of changes between patients experiencing
some degree of resolution (full or partial) and those showing no change in
atelectasis for both alignments (p ≪ 0.05).Table II lists dose constraints
investigated along with the number of plans meeting the constraints for each plan
type. Also shown is the number of metrics changing from being met in baseline to
unmet in followup and vice-versa. For bone-aligned followup plans, four subjects had
at least one OAR metric improve, e.g., change from being unmet in baseline to being
met in followup, and seven subjects had at least one metric worsen. With carina
alignment, four subjects had at least one metric improve and six subjects had at
least one metric worsen, though some patients differed between the two groups. It
should be noted that lungs were defined to include atelectasis in this study, whereas
lung
delineations of the original clinical treatment plans excluded atelectatic
tissue,
explaining the lack of baseline plans meeting lung constraints.
TABLE II.
Dose
constraints.
No. of plans meeting
constraint
No. improved
No. worsened
Structure
Metric
Limit
BL
FUBone
FUCarina
FUBone
FUCarina
FUBone
FUCarina
Spinal cord
Dmax
50.5 Gy
17
17
17
0
0
0
0
Esophagus
Dmean
34 Gy
12
12
10
0
0
0
2
Heart
V40
50%
17
16
16
0
0
1
1
Heart
V60
30%
17
17
17
0
0
0
0
Lungs
Dmean
20 Gy
7
10
10
3
3
0
0
Lungs
V20
30%
6
5
7
0
1
1
0
Lungs
V30
20%
4
3
3
0
0
1
1
Lungs-CTV
Dmean
20 Gy
15
12
12
0
0
3
3
Lungs-CTV
V20
30%
12
10
11
1
1
3
2
Lungs-CTV
V30
20%
7
7
7
1
1
1
1
Lungs-GTV
Dmean
20 Gy
9
10
10
1
1
0
0
Lungs-GTV
V20
30%
7
6
8
0
1
1
0
Lungs-GTV
V30
20%
4
3
3
0
0
1
1
Note: Dosimetric constraints used to evaluate dose changes are
shown along with the number of plans out of eighteen total meeting each
constraint in baseline, bone-aligned followup, and carina-aligned followup.
Volume constraints are given in units of % structure volume. A constraint is
defined as Improved if its limit was unmet in baseline and became met in
followup, whereas Worsened signifies a constraint which was met in baseline
but was unmet in followup.
Dose
constraints.Note: Dosimetric constraints used to evaluate dose changes are
shown along with the number of plans out of eighteen total meeting each
constraint in baseline, bone-aligned followup, and carina-aligned followup.
Volume constraints are given in units of % structure volume. A constraint is
defined as Improved if its limit was unmet in baseline and became met in
followup, whereas Worsened signifies a constraint which was met in baseline
but was unmet in followup.Figure 6 illustrates how atelectasis resolution
had varying impact on changes seen in lung
dose metrics.
For patient 5, full resolution of atelectasis occurred which, in combination with
significant tumor regression, caused the lung to expand and normal
tissue to
fall into the high dose region near the target. This resulted in increased
dose to
healthy lung
and changes in lung metrics from baseline. Patient 12 also experienced full
resolution of atelectasis. In this case, however, the atelectatic tissue was located mostly
outside the high dose region and thus did not cause a large increase in healthy
tissue
dose. For
reference, atelectatic lobe mass and density changes were, respectively, −29.4% and
−59.9% for patient 5 and −39.1% and −64.7% for patient 12.
FIG. 6.
Dose
distributions for two subjects, patient 5 (top) and patient 12 (bottom), are
shown for baseline (left) and bone-aligned followup (right). Contours are shown
for the lobe experiencing atelectasis resolution (blue) and the GTV (red).
Dose
is displayed as a percentage of prescription dose, 66 Gy for
patient 5 and 58 Gy for patient 12. While both subjects experienced full
resolution of whole lobe collapse, patient 5 had large dose changes while
patient 12 showed modest differences due to the majority of atelectasis being a
greater distance from the high dose region.
Dose
distributions for two subjects, patient 5 (top) and patient 12 (bottom), are
shown for baseline (left) and bone-aligned followup (right). Contours are shown
for the lobe experiencing atelectasis resolution (blue) and the GTV (red).
Dose
is displayed as a percentage of prescription dose, 66 Gy for
patient 5 and 58 Gy for patient 12. While both subjects experienced full
resolution of whole lobe collapse, patient 5 had large dose changes while
patient 12 showed modest differences due to the majority of atelectasis being a
greater distance from the high dose region.
GTV coverage
Changes in GTV coverage were investigated in addition to OAR dose changes.
Dose
variation to GTV was analyzed rather than CTV, as it is not clear if the midtreatment
CTV should be constructed by re-expansion of the GTV or by tracking the movement of
tissue
within the CTV region. Means
(stdev) of changes in volume of GTV receiving at least the prescription
dose,
V, were −5.5% (13.5%) and −5.3% (15.5%) for bone and
carina alignments, respectively, and are reported in units of % GTV volume.
V changes ranged from −44.31% to +6.6% for bone
alignment and from −57.3% to +10.3% for carina alignment. Due to the variation in
prescription dose among patients of this study, the following dose changes are reported
in units of % prescription dose. The mean (stdev) change in maximum GTV dose was 2.5% (4.2%) and
ranged from −2.9% to +13.9% for bone alignment and was 2.3% (3.8%) and ranged from
−2.1% to +13.3% for carina alignment. Similarly, mean (stdev) change in minimum GTV
dose was
0.8% (8.9%) and ranged from −19.3% to +19.0% for bone alignment and was 0.2% (9.0%)
and ranged from −17.1% to +19.5% for carina alignment. Tumor regression
typically occurs during radiotherapy and can have an impact on target coverage. For
reference, the mean (stdev) of GTV volume reduction at the time of the midtreatment
imaging was 39.2% (26.7%) for the patients of this study.
DISCUSSION
This work has investigated mass, density, and OAR dose changes in radiotherapy
subjects with obstructive lobar atelectasis using baseline and midtreatment fan-beam
CT scans.For a subset of patients, some dose constraints were unmet by the baseline plan. As described in
Sec. 3.D, the large number of unmet baseline
lung
constraints was likely due to the difference in lung delineations between
this study, which included atelectatic tissue as part of the lung, and clinical treatment
planning, where collapsed lung is excluded. Five patients exceeded the mean esophagus
dose limit in
their baseline plans, though this was not surprising as the esophagus dose limit is a soft
constraint which is sometimes exceeded when necessary to obtain adequate target
coverage.Predicting dosimetric change for a particular patient is challenging due to the
multifactorial nature including amount and location of atelectasis at baseline, amount
and shape of atelectasis resolution, location of organs at risk relative to target
position, etc. For example, the largest of all dose changes occurred in a
patient with partial resolution (mean esophagus dose increase >7 Gy), due
to location of the esophagus relative to high dose gradients. Thus, the goal of the dosimetric portion of this
study was to investigate the range and magnitude of such changes, on average, rather
than try to predict the amount of dosimetric change for particular groups.Midtreatment alignment based on carina, rather than bone, in the presence of atelectasis
resolution causes less dose differences and slightly more dose constraints to improve
during followup compared to their baseline status. Yet in both cases, while not always
causing metrics to exceed clinical limits, large max and min differences from baseline
occurred in lung
dose/volume
metrics. Heart
V40 showed a maximum increase of over 9% for bone alignment
and was only slightly reduced to 6% when the subject was aligned via carina.While dose
changes from baseline to followup were small when averaged across patients, a few
subjects experienced large dose increases to critical structures from the intended
dose/volumes
of the baseline plan. Maximum dose to spinal cord varied substantially with a standard deviation
across patients of 3 Gy for bone alignment. Likewise, mean dose to the esophagus had
large variations across patients, with standard deviations of about 2.7 Gy for both
alignment methods. Mean lung
dose was
expected to be relatively insensitive to changes in dose distribution yet
increased in some subjects by over 4–7 Gy, depending on the lung definition. A high
percentage of constraints were violated, and these unmet limits were spread over all
patients rather than just a small subset, highlighting the need for adaptive
planning.Lung
tissue was
defined in three ways for the purposes of constraint evaluation. Inclusion of all
lung
tissue and
tumor within
the lung borders
minimized the effects of tumor regression. Removal of the GTV from lungs assessed
dose to
tissue
appearing as healthy lung in a CT
image.
Defining the lungs in this way included tissue adjacent to the
tumor volume,
which received high dose. RTOG 0617 evaluates lung constraints on the lungs minus CTV. By doing
so, dose
intended to be delivered to microscopic disease extension was excluded, leaving only
healthy lung
tissue where no
dose was
desired.The constraint metric changes reported in Table I
were consistent with the various lung definitions. Lung-GTV had the largest changes, reflective of
the tumor
regression which had occurred between baseline and followup. As the tumor shrunk, an increased
volume of lung
tissue was
included in the high-dose region increasing the difference from baseline lung metrics. The smallest
differences between baseline and followup resulted from the lung and lung-CTV
definitions since these included parenchyma that were at a greater distance from the
high dose
region.Relative mass and density changes were calculated on a lobe-by-lobe basis in order to
reduce delineation uncertainty, as lobe fissures were more easily discerned. Healthy
lobes and lobes which showed expansion or no change in atelectasis did not have
significant mass change. Lobes with partial resolution of atelectasis by the
midtreatment time point showed, on average, a decrease in mass. When full resolution of
atelectasis occurred, mixed results were observed where some patients had increase in
mass and others had decrease in mass, yet the mass change on average showed only a
slight decrease of 2.8%. Lack of clarity in results for full resolution lobes is likely
a consequence of limited numbers experiencing complete resolution of atelectasis
(n = 4 for full resolution).Given that resolution of atelectasis occurs through reaeration of collapsed
lung
tissue, we
hypothesized that no mass change would occur during atelectasis resolution. Partial
resolution results reported here demonstrate that there is reduction in overall lobe
mass as the lobe transitions from a consolidated collapsed state back to healthy
parenchyma. One possible explanation for decrease in mass is the additional presence of
edema and/or infiltrate in atelectatic lung, which may resolve following re-aeration. Density
change results were in alignment with expectations. Healthy lobes and those experiencing
no change in atelectasis showed no significant changes in tissue density. Atelectatic
lobes had decreases in density proportional to the degree of re-aeration, with full
resolution lobes showing larger density decreases than partially resolving lobes.
Although not statistically significant, median mass and density changes for healthy
lobes were less than zero and warrant further investigation. Calculation of the
cumulative dose
requires deformable image
registration which is challenged by large geometric changes,
particularly when accompanied by mass and density variations such as those observed in
this study. Understanding mass and density changes of atelectasis can help design new
registration algorithms geared more toward accurate modeling and registration of these
changes.Previous studies have investigated the impact of atelectasis resolution on
dose but with
conflicting conclusions. One study implementing a traffic light protocol system to catch
large setup errors on a perfraction basis found 6% of lung
cancerpatients
experienced atelectasis resolution; all were flagged by the protocol, and 75% were
severe enough to require inspection by an oncologist. Another investigation into dosimetric impact of
atelectasis resolution and tumor regression found both changes to have minimal effect on
dose yet
stated that replanning may be necessary. However, their study only investigated tissue density changes,
ignoring geometry changes, which likely explains the difference in results compared to
the current work. Additionally, a study into the benefits of adaptive radiotherapy
estimated that 30% of their subjects with atelectasis would gain no benefit from plan
adaptation.Despite the differences in conclusions, the degree of atelectasis resolution among
patient cohorts was similar to previous inquiries. Møller reported 110 ml median atelectasis volume change, whereas the
subjects in this study had average and median atelectasis volume reductions of 144 ml
and 73 ml, respectively. The study by Grams et al. simulated atelectasis resolution by replacing
atelectatic tissue with intensities of lung parenchyma to create
pseudo CT
images on
which dose was
evaluated. The authors acknowledge that this would not realistically simulate
re-aeration as such changes are usually accompanied by surrounding tissue deformation. Our
study addressed this limitation by calculating dose on actual midtreatment
images. Not only were tissue heterogeneities taken into account, but also real
tissue
deformations and displacements were present. The analysis by Møller et
al. also relied on pseudo
CT
images
for dose
investigation and did not evaluate dose metrics for OARs. By using clinically relevant
dose/volume
metrics, the critical structure dose investigation of the current study provided an accurate
assessment of the dosimetric impact of atelectasis resolution.The dose change
analysis presented in this study assumes the observed tissue changes occur prior
to the start of treatment and remain through the entire treatment course. This
dose
calculation method provides a worst-case scenario of dose change and
overestimates dose differences when atelectasis resolution occurs later in the
treatment course. A more precise measure of dosimetric change would involve deformable registration
of more frequent imaging and accumulation of the dose; however, this is
challenging due to the inaccuracy of registration in presence of large changes. Random
interfraction variations may reduce and/or increase dosimetric changes, and
their combined effect may wash out over the full duration of treatment. A major
limitation of the study was small patient number due to the limited number of
NSCLCpatients presenting with atelectasis and having repeat CT scans. Contour
delineation variability, another important concern, was minimized by delineation of
lung lobes
rather than the atelectasis directly and through the use of a single observer when
possible. Lobe fissures are relatively easy to identify, whereas determining
correspondences between subvolumes of a lobe is challenging at best and becomes nearly
impossible when partial lobar atelectasis is present. While several individuals provided
assistance with organ delineation, a single, experienced radiation oncologist reviewed all
contours and altered them when necessary. The small deviations of mass and density
changes across repeat image pairs suggest delineation variability had a negligible impact
on our results. Differentiating tumor from atelectasis was also challenging for a subset of
patients, as little to no contrast differences were visible in the CT scans. Clinical GTV
contours, which utilized positron emission tomography to assist in GTV delineation, were
used as a guide to minimize this uncertainty.
CONCLUSIONS
Mass and density of lung parenchyma appeared to decrease on average by midtreatment
regardless of the degree to which atelectasis resolved. Re-aeration of collapsed
lung had an
impact on normal tissue
dose due to mass
and density changes, with up to 44% of patients having 5 Gy/5% or larger variations in
at least one clinical dose constraint metric.
CONFLICT OF INTEREST DISCLOSURE
Elisabeth Weiss receives research support from Philips Healthcare and the National
Institutes of Health, has a licensing agreement with Varian Medical Systems, and
receives royalties from UpToDate. Gary Christensen maintains research grants from
National Institutes of Health and has received a gift from Roger Koch to support
research. Geoffrey Hugo receives research support from Philips Healthcare and the
National Institutes of Health and has a licensing agreement with Varian Medical
Systems.
Authors: Michael P Grams; Luis E Fong de Los Santos; Lindsay C Brown; Charles S Mayo; Sean S Park; Yolanda I Garces; Kenneth R Olivier; Debra H Brinkmann Journal: Pract Radiat Oncol Date: 2014-04-03
Authors: Karl Wurstbauer; Heinz Deutschmann; Peter Kopp; Manfred Kranzinger; Florian Merz; Olaf Nairz; Michael Studnicka; Felix Sedlmayer Journal: Int J Radiat Oncol Biol Phys Date: 2009-11-10 Impact factor: 7.038