Literature DB >> 27797105

Gradient and shim pre-emphasis by inversion of a linear time-invariant system model.

S Johanna Vannesjo1, Yolanda Duerst1, Laetitia Vionnet1, Benjamin E Dietrich1, Matteo Pavan1, Simon Gross1, Christoph Barmet1,2, Klaas P Pruessmann1.   

Abstract

PURPOSE: The goal of this contribution is to enhance the fidelity and switching speed of gradient and shim fields by advancing pre-emphasis toward broadband and full cross-term correction. THEORY AND METHODS: The proposed approach is based on viewing gradient and shim chains as linear, time-invariant (LTI) systems. Pre-emphasis is accomplished by inversion of a broadband digital system model. In the multiple-channel case, it amounts to a matrix of broadband filters that perform concerted self- and cross-term correction. This approach is demonstrated with gradients and shims up to the third order in a 7 Tesla whole-body MR system.
RESULTS: Pre-emphasis by LTI model inversion is first verified by studying settling speeds and response behavior without and with the correction. It is then demonstrated for rapid shim updating, achieving substantially enhanced fidelity of field dynamics and shim settling within approximately 1 ms. In single-shot echo-planar imaging (EPI) acquisitions in vivo, this benefit is shown to translate into enhanced geometric fidelity.
CONCLUSIONS: The fidelity of gradient and shim dynamics can be greatly enhanced by pre-emphasis based on inverting a general LTI system model. Permitting shim settling on the millisecond scale, broadband multiple-channel pre-emphasis promises to render higher-order shimming fully versatile at the level of MRI sequence design. Magn Reson Med 78:1607-1622, 2017.
© 2016 International Society for Magnetic Resonance in Medicine. © 2016 International Society for Magnetic Resonance in Medicine.

Keywords:  GIRF; LTI; SIRF; dynamic shimming; higher-order shimming; pre-emphasis

Mesh:

Year:  2016        PMID: 27797105     DOI: 10.1002/mrm.26531

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  4 in total

Review 1.  Spinal cord MRI at 7T.

Authors:  Robert L Barry; S Johanna Vannesjo; Samantha By; John C Gore; Seth A Smith
Journal:  Neuroimage       Date:  2017-07-03       Impact factor: 6.556

Review 2.  In vivo B0 field shimming methods for MRI at 7T.

Authors:  Jason P Stockmann; Lawrence L Wald
Journal:  Neuroimage       Date:  2017-06-07       Impact factor: 6.556

3.  Dynamic B0 shimming for multiband imaging using high order spherical harmonic shims.

Authors:  Hoby P Hetherington; Chan Hong Moon; Michael Schwerter; Nadim Joni Shah; Jullie W Pan
Journal:  Magn Reson Med       Date:  2020-08-28       Impact factor: 4.668

4.  Thermal variation in gradient response: measurement and modeling.

Authors:  Jennifer Nussbaum; Benjamin E Dietrich; Bertram J Wilm; Klaas P Pruessmann
Journal:  Magn Reson Med       Date:  2021-12-21       Impact factor: 3.737

  4 in total