Joel S Beckett1, Gary R Duckwiler2, Satoshi Tateshima2, Viktor Szeder2, Reza Jahan2, Nestor Gonzalez2, Fernando Vinuela2.
Abstract
Due to technical limitations, small, distal, and tortuous intracranial pathology is sometimes out of reach of the current armamentarium of microcatheters designed for intracranial coil embolization. The Marathon microcatheter (Medtronic, Minneapolis, Minnesota, USA), designed specifically for the delivery of Onyx, is longer and more flexible than most coil delivery catheters. We report on nine patients (three with arteriovenous fistula, three with arteriovenous malformation, two with intracranial aneurysm, and one with tumor) where Marathon was used to deliver commercially available platinum coils. We also conducted laboratory compatibility testing and conclude that the Marathon can be used as a coil delivery catheter for Barricade coils (Blockade Medical, Irvine, California, USA) with diameter less than 0.012 in.Entities:
Keywords: Coil embolization; Marathon microcatheter; endovascular
Mesh:
Substances:
Year: 2016 PMID: 27789619 PMCID: PMC5305147 DOI: 10.1177/1591019916667722