A gate evaluation of the sources of error in quantitative 90Y PET.

Jared Strydhorst1, Thomas Carlier2, Arnaud Dieudonné3, Maurizio Conti4, Irène Buvat1.   

Abstract

PURPOSE: Accurate reconstruction of the dose delivered by 90Y microspheres using a postembolization PET scan would permit the establishment of more accurate dose-response relationships for treatment of hepatocellular carcinoma with 90Y. However, the quality of the PET data obtained is compromised by several factors, including poor count statistics and a very high random fraction. This work uses Monte Carlo simulations to investigate what impact factors other than low count statistics have on the quantification of 90Y PET.
METHODS: PET acquisitions of two phantoms-a NEMA PET phantom and the NEMA IEC PET body phantom-containing either 90Y or 18F were simulated using gate. Simulated projections were created with subsets of the simulation data allowing the contributions of random, scatter, and LSO background to be independently evaluated. The simulated projections were reconstructed using the commercial software for the simulated scanner, and the quantitative accuracy of the reconstruction and the contrast recovery of the reconstructed images were evaluated.
RESULTS: The quantitative accuracy of the 90Y reconstructions were not strongly influenced by the high random fraction present in the projection data, and the activity concentration was recovered to within 5% of the known value. The contrast recovery measured for simulated 90Y data was slightly poorer than that for simulated 18F data with similar count statistics. However, the degradation was not strongly linked to any particular factor. Using a more restricted energy range to reduce the random fraction in the projections had no significant effect.
CONCLUSIONS: Simulations of 90Y PET confirm that quantitative 90Y is achievable with the same approach as that used for 18F, and that there is likely very little margin for improvement by attempting to model aspects unique to 90Y, such as the much higher random fraction or the presence of bremsstrahlung in the singles data.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27782704     DOI: 10.1118/1.4961747

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Radioembolization Dosimetry with Total-Body 90Y PET.

Authors:  Gustavo Costa; Benjamin Spencer; Negar Omidvari; Cameron Foster; Michael Rusnak; Heather Hunt; Denise T Caudle; Rex T Pillai; Catherine Tram Vu; Emilie Roncali
Journal:  J Nucl Med       Date:  2021-11-18       Impact factor: 11.082

2.  Comparison of TOF-PET and Bremsstrahlung SPECT Images of Yttrium-90: A Monte Carlo Simulation Study.

Authors:  Akihiko Takahashi; Kazuhiko Himuro; Shingo Baba; Yasuo Yamashita; Masayuki Sasaki
Journal:  Asia Ocean J Nucl Med Biol       Date:  2018
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.