Literature DB >> 27765832

IMAGING DOSE OF HUMAN ORGANS FROM kV-CBCT IN IMAGE-GUIDED RADIATION THERAPY.

Kihong Son1, Jin Sung Kim2, Hoyeon Lee1, Seungryong Cho1.   

Abstract

This study investigates dose distribution due to kV cone-beam computed tomography (CBCT) for the patients undergoing CBCT-based image-guided radiation therapy (IGRT). The kV-CBCT provides an efficient image-guidance tool for acquiring the latest volumetric image of a patient's anatomy, and has been being routinely used in clinics for an accurate treatment setup. Imaging radiation doses resulting from six different acquisition protocols of the on-board imager (OBI) were calculated using a Geant4 Application for Tomographic Emission (GATE) Monte Carlo simulation toolkit, and the absorbed doses by various organs were analyzed for the adult and pediatric numerical XCAT phantoms in this study. The calculated organ doses range from 0.1 to 24.1 mGy in the adult phantom, and from 0.1 to 36.8 mGy in the pediatric one. The imaging organ doses to the pediatric phantom turn out to be consistently higher than those to the adult phantom. It is believed that our results would provide reliable data to the clinicians for their making better decisions on CBCT scanning options and would also provide a platform for developing a new kV-CBCT scanning protocol in conjunction with a low-dose capability.
© The Author 2016. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Entities:  

Mesh:

Year:  2017        PMID: 27765832     DOI: 10.1093/rpd/ncw285

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  7 in total

Review 1.  Application of the 4-D XCAT Phantoms in Biomedical Imaging and Beyond.

Authors:  W Paul Segars; B M W Tsui; George S K Fung; Ehsan Samei
Journal:  IEEE Trans Med Imaging       Date:  2017-08-10       Impact factor: 10.048

2.  Dose assessment for daily cone-beam CT in lung radiotherapy patients and its combination with treatment planning.

Authors:  Hassan Albarakati; Price Jackson; Okan Gulal; Prabhakar Ramachandran; Glen Osbourne; Michael Liu; Tomas Kron; Moshi Geso
Journal:  Phys Eng Sci Med       Date:  2022-01-25

3.  Estimation of effective imaging dose and excess absolute risk of secondary cancer incidence for four-dimensional cone-beam computed tomography acquisition.

Authors:  Yuki Yuasa; Takehiro Shiinoki; Ryota Onizuka; Koya Fujimoto
Journal:  J Appl Clin Med Phys       Date:  2019-10-08       Impact factor: 2.102

4.  Evaluation of patient organ doses from kilovoltage cone-beam CT imaging in radiation therapy.

Authors:  Alper Özseven; Bahar Dirican
Journal:  Rep Pract Oncol Radiother       Date:  2021-04-14

5.  Technical note: No increase in effective dose from half compared to full rotation pelvis cone beam CT.

Authors:  Pascal Hauri; Roger A Hälg; Uwe Schneider
Journal:  J Appl Clin Med Phys       Date:  2017-08-02       Impact factor: 2.102

6.  Imaging dose and secondary cancer risk in image-guided radiotherapy of pediatric patients.

Authors:  Yvonne Dzierma; Katharina Mikulla; Patrick Richter; Katharina Bell; Patrick Melchior; Frank Nuesken; Christian Rübe
Journal:  Radiat Oncol       Date:  2018-09-05       Impact factor: 3.481

7.  Evaluation of IGRT-Induced Imaging Doses and Secondary Cancer Risk for SBRT Early Lung Cancer Patients In Silico Study.

Authors:  Yan-Hua Duan; Heng-Le Gu; Xiao-Hui Yang; Hua Chen; Hao Wang; Yan Shao; Xiao-Yang Li; Ai-Hui Feng; Yan-Chen Ying; Xiao-Long Fu; Kui Ma; Tao Zhou; Zhi-Yong Xu
Journal:  Technol Cancer Res Treat       Date:  2021 Jan-Dec
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.