| Literature DB >> 27754382 |
Ryuichi Watanabe1, Chika Sugai2, Taichi Yamazaki3, Ryoji Matsushima4, Hajime Uchida5, Masahiro Matsumiya6, Akiko Takatsu7, Toshiyuki Suzuki8.
Abstract
ERETIC2 (Electronic Reference To access In vivo Concentrations 2) based on PULCON (Pulse Length-based Concentration determination) methodology is a quantitative NMR (qNMR) using an external standard. The performance of the PULCON method was assessed using maleic acid (MA). Quantification of the diarrhetic shellfish toxin and okadaic acid by PULCON was successfully consistent with that obtained by a conventional internal standard method, demonstrating that the PULCON method is useful for the quantification of invaluable marine toxins without any contaminations by an internal standard.Entities:
Keywords: ERETIC2; PULCON; external standard; okadaic acid; qNMR; quantitative NMR
Mesh:
Substances:
Year: 2016 PMID: 27754382 PMCID: PMC5086654 DOI: 10.3390/toxins8100294
Source DB: PubMed Journal: Toxins (Basel) ISSN: 2072-6651 Impact factor: 4.546
Effect of different deuterated solvents on NMR quantitation 1.
| Solvents | Recovery (%) | RSD (%, |
|---|---|---|
| D2O | 100.3 | 0.0 |
| CD3OD | 100.3 | 0.3 |
| Acetone- | 97.9 | 0.0 |
| DMSO- | 98.1 | 0.2 |
| Acetonitrile- | 97.1 | 0.2 |
| Average | 98.9 | - |
| Stdev | 1.5 | - |
| RSD (%) | 1.5 | - |
1 0.55 mM maleic acid in CD3OD was used as reference. Sample was used 2.86 mM maleic acid (n = 3).
Figure 1Relationship between receiver gain value and integral.
Effect of number of scans on NMR quantitation 1.
| Number of Scans | Recovery (%) |
|---|---|
| 1 | 100.7 |
| 8 | 101.4 |
| 16 | 101.4 |
| 32 | 101.4 |
| 64 | 100.7 |
| Average | 101.1 |
| Stdev | 0.4 |
| RSD (%) | 0.4 |
1 0.55 mM maleic acid in CD3OD was used as reference.
Repeatability (%) of maleic acid with intra-day and inter-day precision 1.
| Samples | Day 1 | Day 2 | Day 3 | Average | Stdev | RSD (%) |
|---|---|---|---|---|---|---|
| 1 | 99.7 | 100.3 | 99.5 | 99.8 | 0.4 | 0.4 |
| 2 | 99.5 | 100.5 | 99.5 | 99.8 | 0.6 | 0.6 |
| 3 | 99.7 | 100.0 | 99.2 | 99.6 | 0.4 | 0.4 |
| Average | 99.6 | 100.3 | 99.4 | - | - | - |
| Stdev | 0.2 | 0.3 | 0.2 | - | - | - |
| RSD (%) | 0.2 | 0.3 | 0.2 | - | - | - |
1 0.55 mM maleic acid in CD3OD was used as reference. Sample was used 3.69 mM maleic acid in CD3OD.
Figure 2Linear regression of concentration of maleic acid determined by either gravimetric and PULCON methods.
Figure 3Chemical structure and 1H-NMR spectrum of okadaic acid dissolved in CD3OD (800 MHz).
Quantitation results obtained by internal standard method and PULCON.
| Method | Tube | Signal | Integral Region | Quantitation (µg/g) | Average | Stdev | RSD (%) | ||
|---|---|---|---|---|---|---|---|---|---|
| Run1 | Run2 | Run3 | |||||||
| Internal standard method | 1 | 1 | 5.85–5.72 | 344.30 | 344.08 | 340.38 | 342.92 | 2.20 | 0.64 |
| 2 | 5.57–5.49 | 346.23 | 345.58 | 344.36 | 345.39 | 0.95 | 0.28 | ||
| 3 | 5.38–5.25 | 345.44 | 349.27 | 354.62 | 349.78 | 4.61 | 1.32 | ||
| 4 | 5.09–5.03 | 367.78 | 332.26 | 345.23 | 348.42 | 17.98 | 5.16 | ||
| 2 | 1 | 5.85–5.72 | 346.39 | 344.29 | 349.36 | 346.68 | 2.55 | 0.74 | |
| 2 | 5.57–5.49 | 362.29 | 360.85 | 357.01 | 360.05 | 2.73 | 0.76 | ||
| 3 | 5.38–5.25 | 365.55 | 361.48 | 355.51 | 360.85 | 5.05 | 1.40 | ||
| 4 | 5.09–5.03 | 361.77 | 355.68 | 355.76 | 357.74 | 3.49 | 0.98 | ||
| PULCON | 1 | 1 | 5.86–5.72 | 344.81 | 333.91 | 340.59 | 339.77 | 5.50 | 1.62 |
| 2 | 5.58–5.48 | 359.29 | 355.49 | 355.08 | 356.62 | 2.32 | 0.65 | ||
| 3 | 5.41–5.24 | 363.20 | 360.73 | 338.95 | 354.29 | 13.34 | 3.77 | ||
| 4 | 5.10–5.01 | 371.00 | 363.92 | 364.74 | 366.55 | 3.87 | 1.06 | ||
| 2 | 1 | 5.86–5.72 | 339.47 | 351.09 | 342.76 | 344.44 | 5.99 | 1.74 | |
| 2 | 5.58–5.48 | 351.92 | 360.04 | 357.88 | 356.61 | 4.21 | 1.18 | ||
| 3 | 5.41–5.24 | 353.67 | 359.63 | 342.35 | 351.88 | 8.78 | 2.49 | ||
| 4 | 5.10–5.01 | 359.73 | 365.08 | 366.31 | 363.71 | 3.50 | 0.96 | ||