| Literature DB >> 27751595 |
Myung-Won You1, Kyung Won Kim2, Junhee Pyo3, Jimi Huh1, Hyoung Jung Kim4, So Jung Lee4, Seong Ho Park1.
Abstract
We aimed to evaluate the correlation between liver stiffness measurement using transient elastography (TE-LSM) and hepatic venous pressure gradient and the diagnostic performance of TE-LSM in assessing clinically significant portal hypertension through meta-analysis. Eleven studies were included from thorough literature research and selection processes. The summary correlation coefficient was 0.783 (95% confidence interval [CI], 0.737-0.823). Summary sensitivity, specificity and area under the hierarchical summary receiver operating characteristic curve (AUC) were 87.5% (95% CI, 75.8-93.9%), 85.3 % (95% CI, 76.9-90.9%) and 0.9, respectively. The subgroup with low cut-off values of 13.6-18 kPa had better summary estimates (sensitivity 91.2%, specificity 81.3% and partial AUC 0.921) than the subgroup with high cut-off values of 21-25 kPa (sensitivity 71.2%, specificity 90.9% and partial AUC 0.769). In summary, TE-LSM correlated well with hepatic venous pressure gradient and represented good diagnostic performance in diagnosing clinically significant portal hypertension. For use as a sensitive screening tool, we propose using low cut-off values of 13.6-18 kPa in TE-LSM. Copyright ÂEntities:
Keywords: Chronic liver disease; Fibroscan; Liver stiffness measurement; Portal hypertension; Transient elastography
Mesh:
Year: 2016 PMID: 27751595 DOI: 10.1016/j.ultrasmedbio.2016.07.025
Source DB: PubMed Journal: Ultrasound Med Biol ISSN: 0301-5629 Impact factor: 2.998