Impact of Implantation Technique and Plaque Morphology on Strut Embedment and Scaffold Expansion of Polylactide Bioresorbable Scaffold - Insights From ABSORB Japan Trial.

Yohei Sotomi1, Yoshinobu Onuma, Jouke Dijkstra, Jeroen Eggermont, Shengnan Liu, Erhan Tenekecioglu, Yaping Zeng, Taku Asano, Robbert J de Winter, Jeffrey J Popma, Ken Kozuma, Kengo Tanabe, Patrick W Serruys, Takeshi Kimura.   

Abstract

BACKGROUND: The optimal implantation technique for the bioresorbable scaffold (Absorb, Abbott Vascular) is still a matter of debate. The purpose of the present study was to evaluate the effect of implantation technique on strut embedment and scaffold expansion.Methods and 
Results: Strut embedment depth and scaffold expansion index assessed by optical coherence tomography (OCT) (minimum scaffold area/reference vessel area) were evaluated in the ABSORB Japan trial (OCT subgroup: 87 lesions) with respect to implantation technique using either quantitative coronary angiography (QCA) or OCT. Strut embedment was assessed at the strut level (n=667), while scaffold expansion was assessed at the lesion level (n=81). The mean embedment depth was 63±59 µm. Balloon sizing and inflation pressure had no direct effect on strut embedment. Plaque morphology affected strut embedment [nonatherosclerotic (58.9±54.3 µm), fibroatheroma (73.3±59.6 µm), fibrous plaque (59.7±51.1 µm), and fibrocalcific plaque (-3.1±61.6 µm, negative value means malapposition), P <0.001]. The balloon-artery ratio positively correlated with the expansion index. This relationship was stronger when the OCT-derived reference vessel diameter (RVD) was used as a reference for balloon selection rather than the QCA-derived one [predilatation (Pearson correlation r: QCA: 0.167 vs. OCT: 0.552), postdilatation (QCA: 0.316 vs. OCT: 0.717)].
CONCLUSIONS: Underlying plaque morphology influenced strut embedment, whereas implantation technique had no direct effect on it. Optimal balloon sizing based on OCT-derived RVD might be recommended. However, the safety of such a strategy should be investigated in a prospective trial. (Circ J 2016; 80: 2317-2326).

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27725525     DOI: 10.1253/circj.CJ-16-0818

Source DB:  PubMed          Journal:  Circ J        ISSN: 1346-9843            Impact factor:   2.993


  6 in total

1.  Neointimal characteristics comparison between biodegradable-polymer and durable-polymer drug-eluting stents: 3-month follow-up optical coherence tomography light property analysis from the RESTORE registry.

Authors:  Tomoaki Kobayashi; Yohei Sotomi; Satoshi Suzuki; Yuma Hamanaka; Shimpei Nakatani; Jouke Dijkstra; Yoshinobu Onuma; Patrick W Serruys; Yasushi Sakata; Atsushi Hirayama; Yoshiharu Higuchi
Journal:  Int J Cardiovasc Imaging       Date:  2019-10-31       Impact factor: 2.357

2.  Comparable neointimal healing in patients with stable coronary lesions and acute coronary syndrome: 3-month optical coherence tomography analysis.

Authors:  Yuma Hamanaka; Yohei Sotomi; Tomoaki Kobayashi; Takashi Omatsu; Jouke Dijkstra; Yasushi Sakata; Atsushi Hirayama; Akio Hirata; Yoshiharu Higuchi
Journal:  Int J Cardiovasc Imaging       Date:  2021-02-16       Impact factor: 2.357

3.  Strut protrusion and shape impact on endothelial shear stress: insights from pre-clinical study comparing Mirage and Absorb bioresorbable scaffolds.

Authors:  Erhan Tenekecioglu; Yohei Sotomi; Ryo Torii; Christos Bourantas; Yosuke Miyazaki; Carlos Collet; Tom Crake; Solomon Su; Yoshinobu Onuma; Patrick W Serruys
Journal:  Int J Cardiovasc Imaging       Date:  2017-04-01       Impact factor: 2.357

4.  Cardiovascular Disease in Patients with End-Stage Renal Disease on Hemodialysis.

Authors:  Jiro Aoki; Yuji Ikari
Journal:  Ann Vasc Dis       Date:  2017-12-25

5.  Impact of diabetes mellitus on the early-phase arterial healing after drug-eluting stent implantation.

Authors:  Takayuki Ishihara; Yohei Sotomi; Takuya Tsujimura; Osamu Iida; Tomoaki Kobayashi; Yuma Hamanaka; Takashi Omatsu; Yasushi Sakata; Yoshiharu Higuchi; Toshiaki Mano
Journal:  Cardiovasc Diabetol       Date:  2020-12-02       Impact factor: 9.951

6.  Five-year comparative study of thin-strut rapamycin-eluting bioabsorbable scaffold with metallic drug-eluting stent in porcine coronary artery.

Authors:  Yaokun Liu; Bo Zheng; Bin Zhang; Robert Ndondo-Lay; Fangfang Nie; Naijie Tang; Yongsheng Miao; Jianping Li; Yong Huo
Journal:  Front Cardiovasc Med       Date:  2022-07-22
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.