| Literature DB >> 27694916 |
Galina S Zamay1,2, Tatiana N Zamay1, Vasilii A Kolovskii3, Alexandr V Shabanov4, Yury E Glazyrin1,2, Dmitry V Veprintsev1, Alexey V Krat5, Sergey S Zamay4, Olga S Kolovskaya1, Ana Gargaun6, Alexey E Sokolov7, Andrey A Modestov5, Ivan P Artyukhov1, Nikolay V Chesnokov2, Marina M Petrova1, Maxim V Berezovski6, Anna S Zamay1,2.
Abstract
The development of an aptamer-based electrochemical sensor for lung cancer detection is presented in this work. A highly specific DNA-aptamer, LC-18, selected to postoperative lung cancer tissues was immobilized onto a gold microelectrode and electrochemical measurements were performed in a solution containing the redox marker ferrocyanide/ferricyanide. The aptamer protein targets were harvested from blood plasma of lung cancer patients by using streptavidin paramagnetic beads and square wave voltammetry of the samples was performed at various concentrations. In order to enhance the sensitivity of the aptasensor, silica-coated iron oxide magnetic beads grafted with hydrophobic C8 and C4 alkyl groups were used in a sandwich detection approach. Addition of hydrophobic beads increased the detection limit by 100 times. The detection limit of the LC-18 aptasensor was enhanced by the beads to 0.023 ng/mL. The formation of the aptamer - protein - bead sandwich on the electrode surface was visualized by electron microcopy. As a result, the electrochemical aptasensor was able to detect cancer-related targets in crude blood plasma of lung cancer patients.Entities:
Mesh:
Substances:
Year: 2016 PMID: 27694916 PMCID: PMC5046130 DOI: 10.1038/srep34350
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1Stability of the LC-18 based aptasensor in Tris-ClO4 buffer only (diamonds) and in blood plasma of a lung cancer patient (circles).
Figure 2(A) Plot of ΔI vs. LCPP (1) and LCPP bound with IH-SAB 4C–8C beads (2) ranging from 230 ng/mL to 0.0023 ng/mL. (B) Plot of ΔI vs. LCPP (1) and LCPP bound with IH-SAB 4C–8C beads (2) ranging from 230 ng/mL to 0.0023 ng/mL in logarithmic coordinates.
ΔI values in SWV for LCP and HP blood plasma samples and LCPP analyzed by LC-18 aptasensor.
| Sample group | Sample name | ΔIaptamer-target, μA | ΔIaptamer-IH-SAB 4C–8C, μA | Lung cancer type | Lung cancer stage | MTS | Other diseases | WBS* 109/L | NEU* 109/L | HGB* g/L | Age | Sex |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Lung cancer patients’ blood plasma | 1 | 24.52 | — | Small cell | IV | Brain | Duodenal ulcer, chronic bronchitis | 9.08 | 5.38 | 131 | 59 | M |
| 2 | 38.73 | — | Squamous | IIA | — | Chronic obstructive pulmonary disease | 8.5 | 5.82 | 119 | 67 | F | |
| 3 | 42.27 | — | Adenocarcinoma | IIIB | — | Chronic obstructive pulmonary disease, duodenal ulcer | 8.68 | 4.70 | 106 | 65 | M | |
| 4 | 53.46 | 61.17 | Adenocarcinoma | IIIA | — | Hypertonic disease | 7.04 | 5.21 | 125 | 52 | M | |
| 5 | 37.23 | 38.83 | Small cell | IIA | — | Cardiac ischemia, myocardial infarction stomach ulcer | 6.32 | 6.11 | 106 | 69 | M | |
| 6 | 46.40 | 61.91 | Squamous | IIIB | — | Chronic obstructive pulmonary disease | 17.74 | 4.49 | 139 | 66 | M | |
| 7 | 59.70 | 61.17 | Squamous | IIIB | — | Transient ischemic attack | 12.11 | 5.45 | 112 | 45 | M | |
| 8 | 25.67 | 28.89 | Adenocarcinoma | IB | — | Paroxysmal atrial fibrillation, gastrectomy for peptic ulcer disease | 10.19 | 6.79 | 146 | 67 | M | |
| 9 | 46.84 | 52.21 | Squamous and adenocarcinoma (mixed type) | IA | — | Hypertonic disease, duodenal ulcer, chronic bronchitis | 7.87 | 5.02 | 116 | 76 | M | |
| Healthy persons’ blood plasma | 10 | 18.37 | — | — | — | — | — | 6.61 | 4.25 | 142 | 62 | M |
| 11 | 9.85 | — | — | — | — | — | 4.01 | 2.95 | 160 | 30 | M | |
| 12 | 21.10 | — | — | — | — | — | 7.08 | 5.31 | 125 | 36 | F | |
| 13 | 18.51 | — | — | — | — | — | 5.15 | 2.37 | 135 | 60 | F | |
| Concentration of purified proteins (ng/mL) harvested from blood plasma obtained from patient LC6 | 230 | 21.90 | 24.23 | Squamous | IIIB | — | Chronic obstructive pulmonary disease | 17.74 | 4.49 | 139 | 66 | M |
| 23 | 12.31 | 19.05 | ||||||||||
| 2,3 | 7.48 | 18.32 | ||||||||||
| 0,23 | −1.15 | 8.46 | ||||||||||
| 0,023 | −3.03 | 4.88 | ||||||||||
| 0,0023 | −3.44 | 1.3 | ||||||||||
| Concentration of purified proteins (ng/mL) harvested from blood plasma obtained from patient LC8 | 230 | 17,4 | 11,01 | Adenocarcinoma | IB | — | Paroxysmal atrial fibrillation, gastrectomy for peptic ulcer disease | 10.19 | 6.79 | 146 | 67 | M |
| 23 | 13,36 | 12,96 | ||||||||||
| 2,3 | 1,74 | 11,34 | ||||||||||
| 0,23 | −5,25 | 2,68 | ||||||||||
| 0,023 | −4,57 | 1,67 | ||||||||||
| 0,0023 | −7,76 | −2,4 | ||||||||||
| Concentration of purified proteins (ng/mL) harvested from blood plasma obtained from patient LC9 | 230 | 18,4 | 22,57 | Squamous and adenocarcinoma (mixed type) | IA | — | Hypertonic disease, duodenal ulcer, chronic bronchitis | 7.87 | 5.02 | 116 | 76 | M |
| 23 | 6,55 | 11,89 | ||||||||||
| 2,3 | 8,25 | 19,15 | ||||||||||
| 0,23 | 9,19 | 9,8 | ||||||||||
| 0,023 | 2,53 | 7,6 |
*WBS – level of white blood cells in blood, NEU – level of neutrophils in blood, HGB – level of hemoglobin in blood.
Figure 3Scheme depicting an aptamer-based sensor for plasma protein detection.
(1) 5′-thiolated DNA aptamer LC-18 is self-assembled onto a gold microelectrode. (2) Cancer-related blood plasma protein binds to the immobilized aptamer forming a complex. (3) IH-SAB 4C–8C beads bind to the protein enhancing SWV signal.
Comparison of proposed biosensor with other reported methodologies for detection of proteins in blood plasma, serum or whole blood.
| Sensing target | Detection limit | Biosensor type | Reference |
|---|---|---|---|
| Angiogenin | 1 pM | SWV based aptamer biosensor | |
| Thrombin | 6.4 nM | MB based aptamer biosensor | |
| Tumor necrosis factor-alpha | 10 ng/mL | MB based aptamer biosensor | |
| Platelet-derived growth factor | 1.25 ng/mL (50 pM) | MB based aptamer biosensor | |
| Adenosine | 60 μM | Turn-on luminescent aptamer biosensor | |
| Histone H2B Neutrophil defensin | 2.3 ng/mL (without IH-SAB 4C–8C) 0.023 ng/mL (with IH-SAB 4C–8C) | SWV based aptamer biosensor | Our work |
Figure 4ΔI values in SWV of LCP and HP blood plasma samples measured by LC-18 aptasensor.
Figure 5(A) Mapping of (a) iron ions and (b) gold ions of the electrodes; (с) contrast EM; (d) overlay of iron and gold ions and contrast EM of the working electrodes surfaces of LC-18-based aptasensors with the formation of targets from the following samples with IH-SAB 4C–8C beads: LCP2 blood plasma sample, LCPP at 23 ng/mL concentration, HP11 blood plasma sample. (B) The specters of iron and gold ions from the working electrodes surfaces incubated with LCP2 blood plasma sample, LCPP at 23 ng/mL, HP3 blood plasma sample. The percentage ratios of iron compared to gold are 3.48%, 2.14%, 0.64% for LCP, LCPP and HP respectively.