Predictors of intraoperative electrosurgery-induced implantable cardioverter defibrillator (ICD) detection.

Hannah Friedman1, John V Higgins1, James D Ryan1, Thomas Konecny1, Samuel J Asirvatham1, Yong-Mei Cha2,3.   

Abstract

BACKGROUND: In the USA, the number of people needing implantable cardioverter defibrillators (ICDs) has grown dramatically. Many ICD recipients will need to undergo a surgical procedure at some point following ICD implantation. Most surgeries involve the use of electrocautery. Currently, the effects of electrocautery-induced electromagnetic interference (EI-EMI) on ICDs are poorly understood. The aim of this study was to study EI-EMI using prospectively collected clinical data.
METHODS: We analyzed prospectively collected ICD data from patients undergoing a surgical procedure at Mayo Clinic between 2011 and 2012. Information on clinical, device history, device interrogation pre- and post-surgery, and surgical information were collected for all patients. ICDs were programmed with detections on and therapies off. The patients were then categorized into two groups: those with EI-EMI inappropriate arrhythmia detection and those without detection. The stored electrograms were reviewed. Clinical and device parameters were analyzed to identify predictors of EI-EMI.
RESULTS: Of 103 patients studied, bipolar cautery did not induce EI-EMI (0/11 cases), whereas monopolar cautery resulted in noise detection in 11/92 procedures. Among 11 inappropriate episodes of detection, 10 had surgery at chest, neck, and upper extremity sites with cautery current across the ICD lead tip; 1 had abdominal surgery; and none had back or low extremity surgery. On average, the near-field electrogram amplitude values were greater than the far-field amplitude values.
CONCLUSIONS: EI-EMI does not occur when bipolar cautery or monopolar cautery is used below the hips with the dispersive ground pad applied to the lower extremities. In contrast to external EMI, EI-EMI may be larger on near-field than far-field electrograms.

Entities:  

Keywords:  Electrocautery; Implantable cardioverter defibrillator; Pacemaker

Mesh:

Year:  2016        PMID: 27665098     DOI: 10.1007/s10840-016-0184-8

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  11 in total

Review 1.  Interference in implanted cardiac devices, part II.

Authors:  Sergio L Pinski; Richard G Trohman
Journal:  Pacing Clin Electrophysiol       Date:  2002-10       Impact factor: 1.976

Review 2.  Implantable defibrillators and sudden cardiac death.

Authors:  Mark Josephson; Hein J J Wellens
Journal:  Circulation       Date:  2004-06-08       Impact factor: 29.690

3.  Effects of surgical and endoscopic electrocautery on modern-day permanent pacemaker and implantable cardioverter-defibrillator systems.

Authors:  Alan Cheng; Saman Nazarian; David D Spragg; Kenneth Bilchick; Harikrishna Tandri; Lynette Mark; Henry Halperin; Hugh Calkins; Ronald D Berger; Charles A Henrikson
Journal:  Pacing Clin Electrophysiol       Date:  2008-03       Impact factor: 1.976

Review 4.  Contemporary pacemaker and defibrillator device therapy: challenges confronting the general cardiologist.

Authors:  Mark H Schoenfeld
Journal:  Circulation       Date:  2007-02-06       Impact factor: 29.690

5.  Clinical treatment of life-threatening ventricular tachyarrhythmias with the automatic implantable defibrillator.

Authors:  M Mirowski; P R Reid; L Watkins; M L Weisfeldt; M M Mower
Journal:  Am Heart J       Date:  1981-08       Impact factor: 4.749

6.  Surgical management of the patient with an implanted cardiac device: implications of electromagnetic interference.

Authors:  J D Madigan; A F Choudhri; J Chen; H M Spotnitz; M C Oz; N Edwards
Journal:  Ann Surg       Date:  1999-11       Impact factor: 12.969

Review 7.  Cardiac pacemakers and electrocautery in ophthalmic surgery.

Authors:  Neena M Peter; Pura Ribes; Ramona Khooshabeh
Journal:  Orbit       Date:  2012-10-12

8.  Application of radiofrequency energy in surgical and interventional procedures: are there interactions with ICDs?

Authors:  Michael Fiek; Uwe Dorwarth; Ilka Durchlaub; Sabine Janko; Christian Von Bary; Gerhard Steinbeck; Ellen Hoffmann
Journal:  Pacing Clin Electrophysiol       Date:  2004-03       Impact factor: 1.976

9.  Electromagnetic Interference (EMI) and arrhythmic events in ICD patients undergoing gastrointestinal procedures.

Authors:  Danette Guertin; Osman Faheem; Thea Ling; Glenn Pelletier; David McComas; Ravi K Yarlagadda; Christopher Clyne; Jeffrey Kluger
Journal:  Pacing Clin Electrophysiol       Date:  2007-06       Impact factor: 1.976

10.  Perioperative management of patients with implantable cardioverter defibrillators.

Authors:  Patrick Donnelly; Nikhil Pal; Niall A Herity
Journal:  Ulster Med J       Date:  2007-05
View more
  1 in total

Review 1.  Perioperative Management of Patients with Cardiac Implantable Electronic Devices and Utility of Magnet Application.

Authors:  Tardu Özkartal; Andrea Demarchi; Maria Luce Caputo; Enrico Baldi; Giulio Conte; Angelo Auricchio
Journal:  J Clin Med       Date:  2022-01-28       Impact factor: 4.241

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.