Chest X-Ray appearances of a dual chamber ICD. Green arrow: battery and pulse generator. Red arrow: right atrial appendage lead (bradycardia sensing and pacing). White arrow: right ventricular lead (bradycardia sensing and pacing, anti-tachycardia pacing and defibrillation).
Perioperative ICD management recommendations7
Establish the device manufacturer and program from the patient-held card Arrange interrogation of the ICD, if not performed within the last six months If diathermy will be required, reprogram the ICD pre-operatively to monitor mode. Bipolar diathermy is preferred and low energy short bursts are desirable If monopolar diathermy is essential, low energy, short bursts are preferred. Diathermy cables and the grounding electrode should be remote from the ICD Arrange for ICD interrogation post-operatively |
Where possible follow elective surgery guidance If the device can not be switched to monitor mode pre-operatively
Restrict diathermy usage and where possible use bipolar diathermy Ensure that cardiopulmonary resuscitation facilities are available If an appropriate ICD shock occurs, correct any reversible causes If recurrent ICD shocks occur, follow standard CPR guidelines Arrange for ICD interrogation post-operatively |