| Literature DB >> 27661352 |
Ming Tang, Xin Wang, Fan Xianguang, Wei Li, Yingjie Xu, Jing Que, Jian He, Yong Zuo.
Abstract
An integrated optical system with high sensitivity and resolution was presented. This was made possible by coupling a probe to a monochromator directly, using an aspheric lens and an achromatic lens in the coupling path, using a dichroic filter with high transmission and a steep transitional zone as a beam splitter, increasing the throughput of the monochromator, and optimizing its structural parameters. This optical system has a high sensitivity; the tested spectrogram of a glass rod demonstrated that the signal-to-noise ratio measured by the integrated optical system was almost 4 times as high as an independent probe and monochromator. The optical system also has the advantages of high resolution (4 cm<sup>-1</sup> or 0.28 nm), low cost, and portable size. This work lays a good groundwork for the development of a high sensitivity, high resolution, and low cost integrated portable Raman spectrometer.Year: 2016 PMID: 27661352 DOI: 10.1364/AO.55.007195
Source DB: PubMed Journal: Appl Opt ISSN: 1559-128X Impact factor: 1.980