Eleanor A Powell1, Joel E Mortensen2.
Abstract
The BioMerieux NucliSENS easyMAG total nucleic acid extractor was evaluated for use on bacterial isolates in the clinical microbiology laboratory. Forty eight isolates were extracted, yielding quantifiable amounts of DNA for all isolates. The easyMAG is appropriate for DNA extraction from bacterial isolates and will be incorporated in the clinical laboratory.Entities:
Keywords: 16S rRNA sequencing; Bacterial sequencing; DNA extraction
Year: 2016 PMID: 27658494 PMCID: PMC5034424 DOI: 10.1186/s12941-016-0168-7
Source DB: PubMed Journal: Ann Clin Microbiol Antimicrob ISSN: 1476-0711 Impact factor: 3.944
Bacterial species included in validation
| Gram-positive cocci | Gram-positive bacilli |
| Non- |
|---|---|---|---|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
| ||
|
|
| ||
|
| |||
|
|
Concentration and A260/A280 of extracted total nucleic acid
| DNA concentration (ng/µL) | A260/A280 | |||
|---|---|---|---|---|
| Median | Range | Median | Range | |
| Gram-positive cocci | 8.5 | 7.4–14.5 | 1.35 | 1.05–1.70 |
| Gram-positive bacilli | 9.1 | 8.0–37.8 | 1.47 | 1.09–1.80 |
|
| 15.9 | 12.1–20.4 | 1.63 | 1.27–1.94 |
| Non- | 17.4 | 7.8–22.0 | 1.67 | 1.02–1.82 |