Alexandra Meidani1, Chara Tzavara2.
Abstract
PURPOSE: The main aim of this study was to compare the efficacy, safety, and predictability of femtosecond laser-assisted in situ keratomileusis performed by two different laser suites in the treatment of myopia for up to 6 months.Entities:
Keywords: IntraLase FS 60 kHz; LASIK; Visx Star S4 IR; WaveLight EX500; WaveLight FS200; myopia
Year: 2016 PMID: 27601880 PMCID: PMC5003563 DOI: 10.2147/OPTH.S110626
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Characteristics of the two study groups
| Group 1
| Group 2
| ||
|---|---|---|---|
| Mean (SD) | Mean (SD) | ||
| Sex, male, n (%) | 6 (42.9) | 7 (50.0) | 0.705 |
| Age (years) | 34.0 (8.3) | 31.8 (9.3) | 0.511 |
| K (D) | 43.8 (1.3) | 43.3 (1.6) | 0.193 |
| CCT (μm) | 558.9 (37.6) | 541.4 (58.8) | 0.092 |
Notes:
Fisher’s exact test;
Student’s t-test. Group 1: FS 60 kHz + Visx Star S4 IR (Abbott, Medical Optics, Inc., Santa Ana, CA, USA); Group 2: WaveLight FS200 (Alcon Laboratories, Inc., Fort Worth, TX, USA) + WaveLight EX500 (Alcon Laboratories, Inc.).
Abbreviations: CCT, corneal thickness; K, keratometry; SD, standard deviation.
Comparison of efficacy between the two study groups
| Group | UDVA 1 week (decimal)
| UDVA 1 month (decimal)
| UDVA 6 months (decimal)
| |||||
|---|---|---|---|---|---|---|---|---|
| Mean | SD | Mean | SD | Mean | SD | |||
| Group 1 | 1.00 | 0.15 | 1.01 | 0.15 | 1.06 | 0.15 | 0.001 | 0.285 |
| Group 2 | 1.04 | 0.12 | 1.04 | 0.12 | 1.08 | 0.13 | 0.008 | |
| 0.304 | 0.293 | 0.602 | ||||||
Notes:
P-value for time effect;
effects reported include differences between the two groups in the degree of change in the particular variable (repeated measurements ANOVA);
P-value for group effect. Group 1: FS 60 kHz + Visx Star S4 IR (Abbott, Medical Optics, Inc., Santa Ana, CA, USA); Group 2: WaveLight FS200 (Alcon Laboratories, Inc., Fort Worth, TX, USA) + WaveLight EX500 (Alcon Laboratories, Inc.).
Abbreviations: ANOVA, analysis of variance; SD, standard deviation; UDVA, uncorrected distance visual acuity.
Figure 1Efficacy and safety outcomes at 6 months (56 eyes).
Notes: Six-month postoperative (56 eyes) cumulative UDVA, group 1 (A); 6-month postoperative cumulative UDVA, group 2 (B); change in corrected distance visual acuity in both groups (C); attempted versus achieved SE in group 1 (D) and in group 2 (E). Group 1: FS 60 kHz + Visx Star S4 IR (Abbott, Medical Optics, Inc., Santa Ana, CA, USA); Group 2: WaveLight FS200 (Alcon Laboratories, Inc., Fort Worth, TX, USA) + WaveLight EX500 (Alcon Laboratories, Inc.).
Abbreviations: CDVA, corrected distance visual acuity; SE, spherical equivalent; UDVA, uncorrected distance visual acuity; postop, postoperative.
Efficacy comparison of the two study groups
| Parameter | Group 1
| Group 2
| |
|---|---|---|---|
| % | % | ||
| 1 week | |||
| 20/16 or better | 21.4 | 28.6 | 0.758 |
| 20/20 or better | 78.6 | 92.9 | 0.252 |
| 1 month | |||
| 20/16 or better | 25.0 | 32.1 | 0.768 |
| 20/20 or better | 78.6 | 92.9 | 0.252 |
| 6 months | |||
| 20/16 or better | 46.4 | 46.4 | >0.999 |
Notes:
Pearson chi-square test. Group 1: FS 60 kHz + Visx Star S4 IR (Abbott, Medical Optics, Inc., Santa Ana, CA, USA); Group 2: WaveLight FS200 (Alcon Laboratories, Inc., Fort Worth, TX, USA) + WaveLight EX500 (Alcon Laboratories, Inc.).
Abbreviation: UDVA, uncorrected distance visual acuity.
Mean MRSE values at baseline and during follow-up for both study groups
| Group | MRSE at baseline
| MRSE at 1 month follow-up
| MRSE at 6 months follow-up
| |||||
|---|---|---|---|---|---|---|---|---|
| Mean | SD | Mean | SD | Mean | SD | |||
| Group 1 | −3.76 | 1.67 | 0.09 | 0.32 | 0.05 | 0.26 | <0.001 | 0.132 |
| Group 2 | −4.57 | 2.2 | 0.13 | 0.26 | 0.07 | 0.24 | <0.001 | |
| 0.206 | 0.802 | 0.495 | ||||||
Notes:
P-value for change from baseline to 6 months;
effects reported include differences between the two groups in the degree of change in the particular variable (repeated measurements ANOVA);
P-value for group effect. Group 1: FS 60 kHz + Visx Star S4 IR (Abbott, Medical Optics, Inc., Santa Ana, CA, USA); Group 2: WaveLight FS200 (Alcon Laboratories, Inc., Fort Worth, TX, USA) + WaveLight EX500 (Alcon Laboratories, Inc.).
Abbreviations: ANOVA, analysis of variance; MRSE, mean refractive spherical equivalent; SD, standard deviation.
Figure 2Predictability and stability at 6 months.
Notes: Six-month refractive accuracy (56 eyes) in both groups (A), refractive astigmatism in group 1 (B) and in group 2 (C), refractive stability in group 1 (D) and in group 2 (E). Group 1: FS 60 kHz + Visx Star S4 IR (Abbott, Medical Optics, Inc., Santa Ana, CA, USA); Group 2: WaveLight FS200 (Alcon Laboratories, Inc., Fort Worth, TX, USA) + WaveLight EX500 (Alcon Laboratories, Inc.).
Abbreviations: preop, preoperative; postop, postoperative.
Postoperative astigmatism at 1 and 6 months for the two study groups
| Group 1
| Group 2
| ||||
|---|---|---|---|---|---|
| n | % | n | % | ||
| Postoperative astigmatism at 1 month | |||||
| ≤0.25 | 17 | 60.7 | 16 | 57.1 | >0.999 |
| 0.26–0.5 | 10 | 35.7 | 11 | 39.3 | |
| 0.51–0.75 | 1 | 3.6 | 1 | 3.6 | |
| Postoperative astigmatism at 6 months | |||||
| ≤0.25 | 23 | 82.1 | 26 | 92.9 | 0.422 |
| 0.26–0.5 | 4 | 14.3 | 2 | 7.1 | |
| 0.51–0.75 | 1 | 3.6 | 0 | 0 | |
Notes:
Fisher’s exact test. Group 1: FS 60 kHz + Visx Star S4 IR (Abbott, Medical Optics, Inc., Santa Ana, CA, USA); Group 2: WaveLight FS200 (Alcon Laboratories, Inc., Fort Worth, TX, USA) + WaveLight EX500 (Alcon Laboratories, Inc.).