Design of a multimodal (1H/23Na MR/CT) anthropomorphic thorax phantom.

Wiebke Neumann1, Florian Lietzmann2, Lothar R Schad2, Frank G Zöllner2.   

Abstract

OBJECTIVES: This work proposes a modular, anthropomorphic MR and CT thorax phantom that enables the comparison of experimental studies for quantitative evaluation of deformable, multimodal image registration algorithms and realistic multi-nuclear MR imaging techniques.
METHODS: A human thorax phantom was developed with insertable modules representing lung, liver, ribs and additional tracking spheres. The quality of human tissue mimicking characteristics was evaluated for 1H and 23Na MR as well as CT imaging. The position of landmarks in the lung lobes was tracked during CT image acquisition at several positions during breathing cycles. 1H MR measurements of the liver were repeated after seven months to determine long term stability.
RESULTS: The modules possess HU, T1 and T2 values comparable to human tissues (lung module: -756±148HU, artificial ribs: 218±56HU (low CaCO3 concentration) and 339±121 (high CaCO3 concentration), liver module: T1=790±28ms, T2=65±1ms). Motion analysis showed that the landmarks in the lung lobes follow a 3D trajectory similar to human breathing motion. The tracking spheres are well detectable in both CT and MRI. The parameters of the tracking spheres can be adjusted in the following ranges to result in a distinct signal: HU values from 150 to 900HU, T1 relaxation time from 550ms to 2000ms, T2 relaxation time from 40ms to 200ms.
CONCLUSION: The presented anthropomorphic multimodal thorax phantom fulfills the demands of a simple, inexpensive system with interchangeable components. In future, the modular design allows for complementing the present set up with additional modules focusing on specific research targets such as perfusion studies, 23Na MR quantification experiments and an increasing level of complexity for motion studies.
Copyright © 2016. Published by Elsevier GmbH.

Entities:  

Keywords:  Anthropomorphes Phantom; Anthropomorphic phantom; Multimodale Bildgebung; Multinukleare MR-Bildgebung; Quantifikation; Thorax-Phantom; multi-nuclear MRI; multimodal imaging; quantification; thorax phantom

Mesh:

Year:  2016        PMID: 27596568     DOI: 10.1016/j.zemedi.2016.07.004

Source DB:  PubMed          Journal:  Z Med Phys        ISSN: 0939-3889            Impact factor:   4.820


  3 in total

1.  Risk assessment of copper-containing contraceptives: the impact for women with implanted intrauterine devices during clinical MRI and CT examinations.

Authors:  Wiebke Neumann; Tanja Uhrig; Matthias Malzacher; Verena Kossmann; Lothar R Schad; Frank G Zoellner
Journal:  Eur Radiol       Date:  2018-11-19       Impact factor: 5.315

Review 2.  COVID-19 Pandemic Spurs Medical Telerobotic Systems: A Survey of Applications Requiring Physiological Organ Motion Compensation.

Authors:  Lingbo Cheng; Mahdi Tavakoli
Journal:  Front Robot AI       Date:  2020-11-09

3.  A novel 3D printed mechanical actuator using centrifugal force for magnetic resonance elastography: Initial results in an anthropomorphic prostate phantom.

Authors:  Wiebke Neumann; Andreas Bichert; Jonas Fleischhauer; Antonia Stern; Roxana Figuli; Manfred Wilhelm; Lothar R Schad; Frank G Zöllner
Journal:  PLoS One       Date:  2018-10-08       Impact factor: 3.240

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.