Literature DB >> 2759644

Performance degradation of the coherent averaging technique by a molding effect.

F Lippens, P Savard, M Shenasa.   

Abstract

In coherent averaging, we show that when temporal alignment is performed by matched filtering, the SNR improvement may reach a stable limit after a certain number of averaged events. The source of this degradation lies in the noise that can be contained in the template of the matched filter. This effect depends on the SNR and the morphology of the signal and it can be reduced by a multipass averaging procedure or by a template updating technique.

Mesh:

Year:  1989        PMID: 2759644     DOI: 10.1109/10.30812

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  1 in total

1.  Fast and reliable QRS alignment technique for high-frequency analysis of signal-averaged ECG.

Authors:  O J Escalona; R H Mitchell; D E Balderson; D W Harron
Journal:  Med Biol Eng Comput       Date:  1993-07       Impact factor: 2.602

  1 in total