Literature DB >> 27582936

Biomechanical testing of a novel osteosynthesis plate for the ulnar coronoid process.

Johannes Kiene1, Jorn Bogun1, Nina Brockhaus1, Klaus Waizner1, Arndt-Peter Schulz1, Robert Wendlandt1.   

Abstract

BACKGROUND: The present study aimed to biomechanically evaluate a novel locking plate intended for osteosynthesis of coronoid fracture compared to mini L-plates and cannulated screws.
METHODS: Biomechanical tests were performed on a fracture model in synthetic bones. Three groups, each with eight implant-bone-constructs, were analyzed in quasi-static and dynamic tests. Finally, samples were tested destructively for maximum strength.
RESULTS: The mean (SD) highest stiffness was measured for the novel plate [693 (18) N/mm], followed by the mini L-plate [646 (37) N/mm] and the cannulated screws [249 (113) N/mm]. During the cycling testing of the novel plate and the mini L-plate, no failures occurred, although three of the eight samples of cannulated screws failed during the test. The mean (SD) maximum strength during the destructive testing was 1333 (234) N for the novel plate, 1338 (227) N for the mini-L-plate and 459 (56) N for the cannulated screws. No statistical differences were found during the destructive testing between the two plates (p = 0.999), although statistical differences were found between both plates and the cannulated screws (p = 0.000 each).
CONCLUSIONS: Osteosynthesis of the coronoid process using the novel plate is mechanically similar to the mini L-plate. Both plates were superior to osteosynthesis with cannulated screws.

Entities:  

Keywords:  Anatomic plate; coronoid process; fracture treatment; locking plate; ulna

Year:  2014        PMID: 27582936      PMCID: PMC4935064          DOI: 10.1177/1758573214532794

Source DB:  PubMed          Journal:  Shoulder Elbow        ISSN: 1758-5732


  37 in total

Review 1.  Evolution of the internal fixation of long bone fractures. The scientific basis of biological internal fixation: choosing a new balance between stability and biology.

Authors:  Stephan M Perren
Journal:  J Bone Joint Surg Br       Date:  2002-11

2.  Elbow joint stability following experimental osteoligamentous injury and reconstruction.

Authors:  Søren R Deutch; Steen L Jensen; Stein Tyrdal; Bo S Olsen; Otto Sneppen
Journal:  J Shoulder Elbow Surg       Date:  2003 Sep-Oct       Impact factor: 3.019

3.  Mechanical characterization of fourth generation composite humerus.

Authors:  P Grover; C Albert; M Wang; G F Harris
Journal:  Proc Inst Mech Eng H       Date:  2011-12       Impact factor: 1.617

4.  Surgical versus non-surgical treatment of ligamentous injuries following dislocation of the elbow joint. A prospective randomized study.

Authors:  P O Josefsson; C F Gentz; O Johnell; B Wendeberg
Journal:  J Bone Joint Surg Am       Date:  1987-04       Impact factor: 5.284

5.  Incidence of elbow dislocations in the United States population.

Authors:  Jason W Stoneback; Brett D Owens; Joshua Sykes; George S Athwal; Lauren Pointer; Jennifer Moriatis Wolf
Journal:  J Bone Joint Surg Am       Date:  2012-02-01       Impact factor: 5.284

6.  The effect of coronoid fractures on elbow kinematics and stability.

Authors:  Daphne M Beingessner; Cynthia E Dunning; Rebecca A Stacpoole; James A Johnson; Graham J W King
Journal:  Clin Biomech (Bristol, Avon)       Date:  2006-11-13       Impact factor: 2.063

7.  Mechanical comparison in cadaver specimens of three different 90-degree double-plate osteosyntheses for simulated C2-type distal humerus fractures with varying bone densities.

Authors:  Ivo Schuster; Jan Korner; Michael Arzdorf; Karsten Schwieger; Gerd Diederichs; Berend Linke
Journal:  J Orthop Trauma       Date:  2008-02       Impact factor: 2.512

8.  Biomechanical testing of a new plate system for the distal humerus compared to two well-established implants.

Authors:  Christine Voigt; Christina Rank; Klaus Waizner; Robert Wendlandt; Andreas Unger; Arndt P Schulz; Christian Jürgens; Felix Renken
Journal:  Int Orthop       Date:  2013-01-29       Impact factor: 3.075

Review 9.  Biomechanics of locked plates and screws.

Authors:  Kenneth A Egol; Erik N Kubiak; Eric Fulkerson; Frederick J Kummer; Kenneth J Koval
Journal:  J Orthop Trauma       Date:  2004-09       Impact factor: 2.512

10.  The influence of induced micromovement upon the healing of experimental tibial fractures.

Authors:  A E Goodship; J Kenwright
Journal:  J Bone Joint Surg Br       Date:  1985-08
View more
  3 in total

1.  Good outcome using anatomically pre-formed buttress plates for anteromedial facet fractures of the coronoid-a retrospective study of twenty-four patients.

Authors:  Alexander Klug; Stefan Buschbeck; Yves Gramlich; Johannes Buckup; Reinhard Hoffmann; Kay Schmidt-Horlohé
Journal:  Int Orthop       Date:  2019-06-07       Impact factor: 3.075

2.  No neurovascular damage after creation of an accessory anteromedial portal for arthroscopic reduction and fixation of coronoid fractures.

Authors:  Paolo Arrigoni; Davide Cucchi; Enrico Guerra; Francesco Luceri; Simone Nicoletti; Alessandra Menon; Pietro Randelli
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2018-04-02       Impact factor: 4.342

3.  Arthroscopic reduction and fixation of coronoid fractures with an exchange rod-a new technique.

Authors:  Kan Ouyang; Daping Wang; Wei Lu; Jianyi Xiong; Jian Xu; Liangquan Peng; Haifeng Liu; Hao Li; Wenzhe Feng
Journal:  J Orthop Surg Res       Date:  2017-01-18       Impact factor: 2.359

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.