Patients' perspectives and preferences in the choice of inhalers: the case for Respimat(®) or HandiHaler(®).

Pieter Nicolaas Richard Dekhuijzen1, Federico Lavorini2, Omar S Usmani3.   

Abstract

Poor inhaler technique hampers the efficacy of drug therapy in asthma and chronic obstructive pulmonary disease. Not only does this affect individual patient care, but it also impacts on the wider health care economics associated with these conditions. Treatment guidelines recommend a systematic approach to drug class selection; however, standardization of inhaler selection is currently difficult owing to the complexity of the interaction between the inhaler device and the patient. Specifically, individual patient preference can influence how successful a treatment is overall. This article reviews inhaler devices from the patient perspective, with a particular focus on the dry powder inhaler HandiHaler(®) and Respimat(®) Soft Mist™ Inhaler. It discusses factors that influence device preference and treatment compliance and reviews tools that can aid health care professionals to better match inhaler devices to individual patients' needs.

Entities:  

Keywords:  Soft Mist™ Inhaler; asthma; chronic obstructive pulmonary disease; inhaler technique; tiotropium HandiHaler®; tiotropium Respimat®

Year:  2016        PMID: 27574405      PMCID: PMC4993394          DOI: 10.2147/PPA.S82857

Source DB:  PubMed          Journal:  Patient Prefer Adherence        ISSN: 1177-889X            Impact factor:   2.711


Introduction

Inhaled therapy remains the cornerstone of treatment for pulmonary disorders such as asthma and chronic obstructive pulmonary disease (COPD). It offers advantages over oral tablet administration, such as effective targeting of treatment to the lungs at a lower dose with a rapid onset of action.1–5 Despite the introduction of new pharmacological treatments and the availability of a range of different types of inhaler devices,1,2,5 asthma and COPD often remain suboptimally controlled in terms of symptom relief and reduction in exacerbation risk.6,7 Incorrect use of inhaler devices is common – real-world studies estimate that up to 94% of patients do not use their inhalers properly.6,8,9 This can lead to inadequate drug dosing, which, in turn, contributes to a worsening of quality of life and suboptimal disease control.8–11 With over 200 different drug–inhaler combinations available,2 matching patient characteristics and preferences to the most appropriate drug treatment and inhaler device remains challenging. Clinical respiratory practice in the majority of patients is to first select the class of medication, followed by the specific agent, and then the inhaler device itself. However, the importance of inhaler selection as part of the treatment decision is becoming increasingly recognized,2,9,12–14 and this article proposes that careful consideration of inhaler selection warrants greater priority among prescribers. This review examines the key considerations for selecting devices most suited to patients’ needs; it also discusses the importance of correct inhaler technique for optimizing treatment outcomes and evaluates patients’ perspectives and preferences when switched from a dry powder inhaler (DPI) to the Respimat® Soft Mist™ Inhaler (SMI; Boehringer Ingelheim International GmbH, Ingelheim am Rhein, Germany).

The case for monitoring inhaler suitability

Treatment guidelines emphasize that good symptom control with inhaled maintenance drugs is fundamental to the management of asthma and COPD.3,4 Nevertheless, guidelines have lacked clarity on specific guidance regarding appropriate inhaler selection for achieving optimum symptom control.2,14 It is encouraging, however, to see that the latest Global Initiative for Asthma and Global Initiative for Chronic Obstructive Lung Disease strategy documents both state the importance of inhaler device selection in day-to-day practice.3,4 It is well established that poor inhalation technique can severely impact the clinical efficacy of medications, resulting in impaired disease control, worsening quality of life, increased exacerbation and mortality risk, increased hospitalizations (Figure 1), and, in turn, increased health care expenditure.8–11 Although incorrect inhaler use may be due, in part, to a patient’s impaired dexterity (particularly with elderly COPD patients), evidence suggests that many patients receive inadequate or no training in the use of their device.2,8,9,15,16 In a systematic literature review of patients evaluated about their use of DPIs for either asthma or COPD, one-quarter of the patients had received no verbal instruction on inhaler technique.8 For those patients who had received instruction, the quality and duration of the instruction and review of technique were considered inadequate.8 Perhaps more concerning are the findings from several studies that many medical and nursing personnel do not have a good understanding of correct inhaler technique, despite their awareness of the importance of correct technique on treatment outcomes.9,15,17–19 Educational efforts therefore need to be focused on primary prescribers of inhaler devices so as to limit or avoid the consequences of poor inhaler technique by their patients. Written instruction is known to be ineffective as a training method; verbal instruction, technique assessment, and reassessment are all essential for achieving correct inhaler use.8 Reflecting this need for improved education, a call to action for improving education of inhaler technique at a policy level was published by the International Primary Care Respiratory Group13 and the Aerosol Drug Management Improvement Team.15
Figure 1

Association between asthma and COPD disease control and at least one critical inhaler error.

Notes: (A) Describes the relationship between risk of at least one critical inhaler error and Asthma Control Test questionnaire score. (B) Describes the relationship between risk of at least one critical inhaler error and self-reporting of some unscheduled health-care resources use in the last year. Data from Melani et al.57

Abbreviation: COPD, chronic obstructive pulmonary disease.

A meta-analysis of 59 randomized trials assessing respiratory treatments delivered by a metered dose inhaler (MDI), nebulizer, or DPI has suggested that when inhalers are used correctly, there is little difference in clinical efficacy between devices.20 Although inhaler selection remains an essential component of tailored treatment, the results from this meta-analysis highlight that considerable improvements in disease management are possible by investing in training of correct inhaler use. Ongoing monitoring of inhaler technique should be an integral component of routine management of COPD and asthma patients. An algorithm proposed by the Aerosol Drug Management Improvement Team gives prescribers a practical tool for assessing whether a change in device type is needed or a step-up in therapy is required (Figure 2).2,15 Although initially developed for asthma patients, this algorithm could also be adapted for the management of patients with COPD.2
Figure 2

ADMIT asthma therapy adjustment flow chart.

Note: Adapted from from Respir Med, 100/9, Crompton GK, Barnes PJ, Broeders M, et al, The need to improve inhalation technique in Europe: a report from the Aerosol Drug Management Improvement Team, 1479–1494, Copyright 2006, with permission from Elsevier.15

Abbreviation: ADMIT, Aerosol Drug Management Improvement Team.

Overview of handheld inhaler devices

Handheld inhalers are available as either pressurized MDIs (pMDIs), DPIs, or SMI (Respimat®, Boehringer Ingelheim, Germany; currently, the only SMI marketed).5 Nebulizers are beyond the scope of this review; however, the articles by Ibrahim et al21 and Lavorini et al5 provide detailed discussions of nebulizer technology. A description of the currently available handheld devices is provided, and a summary of the advantages and disadvantages of each type is given in Table 1.
Table 1

Major components, advantages, and disadvantages of inhaler devices

InhalerFormulationMetering systemAdvantagesDisadvantages
pMDIDrug suspended or dissolved in propellant (with surfactant and cosolvent)Metering valve and reservoirPortable and compactMultidose deviceRelatively inexpensiveCannot contaminate contentsAvailable for most inhaled medicationsDose indicator (new-generation devices)Contains propellantsNot breath actuatedMany patients cannot use it correctlyHigh oropharyngeal deposition
pMDI + spacerLow dependence on inspiratory flow rateEasier to coordinateLarge drug doses delivered more convenientlyLess oropharyngeal depositionLow dependence on inspiratory flow rateHigher lung deposition than a pMDILess portable than a pMDIPlastic spacers may acquire static chargeAdditional cost to a pMDI
BA-MDIDrug suspended in propellantMetering valve and reservoirPortable and compactMultidose deviceBreath actuated (no coordination needed)Cannot contaminate contentsContains propellants“Cold Freon” effectRequires moderate inspiratory flow to be triggered
DPIDrug blend in lactose, drug alone, drug/excipient particlesCapsules, blisters, multidose blister packs, reservoirsPortable and compactBreath actuated (no coordination needed)Does not contain propellantsDose indicator (new generation devices)Locking mechanism when empty (new generation devices)Requires a minimum inspiratory flowMay not be appropriate for emergency situationsMany patients cannot use it correctlyMost types are moisture sensitive
SMI (Respimat®)Aqueous solution or suspensionA spring-loaded system provides the energy required to force aqueous solution from a reservoir through the “uniblock” to produce the soft mistPortable and compactMultidose deviceLow dependence on inspiratory flow rate (slow moving aerosol)High fine particle fractionHigh lung depositionDoes not contain propellantsDose indicatorLocking mechanism when emptyNot breath actuated
NebulizersAqueous solution or suspensionNebule dispensed into reservoir chamber of nebulizerMay be used at any ageNo specific inhalation technique requiredVibrating mesh is portable and does not require an outside energy sourceMay dispense drugs not available with pMDIs or DPIsJet and ultrasonic nebulizers require an outside energy sourceTreatment times can be longPerformance varies between nebulizersJet nebulizers cannot aerosolize a certain volume of solutionRisk of bacterial contaminationNewer nebulizers are expensive

Note: Data from Lavorini22 and Lavorini et al.5

Abbreviations: BA-MDI, breath-actuated metered dose inhaler; DPI, dry powder inhaler; pMDI, pressurized metered dose inhaler; SMI, Soft Mist™ Inhaler.

pMDIs

Introduced in the 1950s,22 pMDIs were the first of the handheld delivery devices. Technological developments in pMDIs have occurred over the past 60 years, but common components of all MDIs are a pressurized canister of drug in solution or suspension, a chamber for producing an aerosol, and a mouthpiece through which the aerosol is inhaled. Older pMDIs created a high-velocity spray of relatively large droplets.23,24 The change in propellants from chlorofluorocarbon to hydrofluoroalkane has led to a reduced-velocity spray that delivers a smaller particle size. These properties lead to reduced oropharyngeal deposition and improved distribution of drug particles throughout the airways with newer hydrofluoroalkane pMDIs compared with chlorofluorocarbon MDIs.23–27 In terms of operational design, two generations of pMDI device now exist in practice. First-generation pMDIs are simple, push-activated, pressure-powered aerosols that require precise coordination and inhalation by the patient. A valve-holding chamber or spacer with a one-way valve allows the aerosol to be inhaled with several intakes and can be added to reduce dependency on coordination. Although spacers are essential for children and those patients with very severe cognitive and physical impairment, they are inconvenient to carry and need to be cleaned between uses.16,28 Second-generation pMDIs require breath actuation rather than push activation to reduce dependency on the patient’s coordination of inhalation and actuation. Currently in development is a new generation of “intelligent” pMDIs that have electronic computation of delivery to compensate for inadequate inhalation and provide more reliable feedback on delivery.29

DPIs

DPIs rely on air drawn through the device to pick up powder from a container and carry it into the lungs within the same airstream.24 Some devices require insertion of drug capsules designed to release all the powdered content into the airstream. Other devices contain multiple doses taken from a reservoir or multiple prefilled blisters/cartridges of drug powder.24 These do not require the user to carry packets of separate drug capsules, but function is otherwise similar to single-capsule devices. Inhalation flow rate through the device is critical to the successful operation of all DPIs, and the inspiratory effort required to deliver the drug to the lungs depends on the device. Not all patients are capable of creating sufficiently controlled inspiratory effort every time, especially during an exacerbation.30 New DPI devices are beginning to address this important issue, including the dose protector of the NEXThaler® (Chiesi Limited, Manchester, UK), which prevents release of the dose until sufficient flow rate is achieved,31 and the electronic power-assisted DPIs (active DPIs) currently in development.5

Slow-moving SMIs

The SMI was introduced in 2007 by Boehringer Ingelheim as Respimat® SMI. It comprises design elements from the pMDI, but was designed to solve some disadvantages associated with pMDIs and DPIs, such as deposition of drug in the oropharynx.32 Respimat® uses spring power (rather than a pressurized container) to generate a low-velocity vapor cloud into the mouth from a liquid formulation of the drug, which is pushed through a specially designed nozzle. The mist generation is sustained for approximately 1.5 seconds but still requires a degree of coordination of inhalation and actuation (although less so than for pMDIs).32 Compared with pMDI devices, lung deposition with Respimat® is up to 50% higher and oropharyngeal deposition is lower.5 Typical scintigraphic images for Respimat® (Boehringer Ingelheim), Turbuhaler® (AstraZeneca, Lund, Sweden) DPI at slow and fast inhaled flow rates, and pMDI devices32,33 are shown in Figure 3. Respimat® also has the benefit of a dose indicator that provides the user with an estimate of doses remaining in the cartridge.
Figure 3

Typical scintigraphic images for Respimat®, Turbuhaler® DPI at slow and fast inhaled flow rates, and pMDI.

Notes: (A) Respimat® SMI; (B) Turbuhaler® DPI – slow; (C) Turbuhaler® DPI – fast; (D) pMDI. Reproduced from Pitcairn G, Reader S, Pavia D, Newman S. Deposition of corticosteroid aerosol in the human lung by Respimat Soft Mist inhaler compared to deposition by metered dose inhaler or by Turbuhaler dry powder inhaler. J Aerosol Med. 2005;18(3):264–272.33 The publisher for this copyrighted material is Mary Ann Liebert, Inc. publishers.

Abbreviations: DPI, dry powder inhaler; pMDI, pressurized metered dose inhaler; SMI, Soft Mist™ Inhaler.

Inhaler techniques

Each type of device requires a different inhalation technique to achieve optimal drug delivery to the lungs (Table 2). Furthermore, specific instructions vary between devices. This can cause confusion among patients34 and highlights why the prescription of multiple inhaler types to one patient, or switching to different inhaler types, requires careful management and discussion with the patient.
Table 2

Correct techniques for using pMDI, DPI (specifically HandiHaler®), and SMI (Respimat®) devices

Inhaler typeCorrect techniqueErrors in technique
pMDIa1. Shake the inhaler well before use (three or four shakes)2. Remove the cap3. Breathe out, away from inhaler4. Place inhaler between teeth and close mouth around it5. Start to breathe in slowly. Press the top of inhaler once and keep breathing in slowly until a full breath has been taken6. Remove the inhaler and hold breath for about 10 seconds, then breathe out7. If a second puff is needed, wait 30 seconds, shake inhaler again, and repeat above steps8. Rinse mouth (after ICS) and record the number of doses takenInhaler not primedInhaler not shakenNo exhalationHolding mouthpiece away from mouth/poor seal around mouthpieceFiring device before start of inhalationFiring device at or after end of inhalationInhaling through noseStopping inhalation as device is firedFast inhalationNo breath-hold/short breath-hold
HandiHaler® (DPI)b,c1. Open the HandiHaler® device. Separate only one of the blisters from the blister card; then open the blister2. Insert the Spiriva® (Boehringer Ingelheim) capsule and close the mouthpiece firmly against the gray base until a click is heard3. Press the green piercing button once until it is flat (flush) against the base, then release4. Breathe out completely. Then, with the HandiHaler® in the mouth, breathe in deeply until lungs are full. Patient should hear or feel the Spiriva® capsule vibrate (rattle)5. Patient must inhale twice from the same Spiriva® capsule in order to take full daily doseIncorrect dose loadingFailure to pierce capsule or release piercing buttonBreathing out into deviceInhalation vents blockedPoor seal around mouthpiecePatients inhaling too forcefully (>40–50 L/min)dInsufficient flow “acceleration”Inhaling through noseNo breath-hold/short breath-holdInappropriate storage
Respimat® (SMI)e,f1. Keep cap closed. Turn base in direction of the arrows on label until a click is heard2. Flip cap open until it clicks into open position3. Close lips around the mouthpiece end. While taking in a slow, deep breath, press dose release button; keep breathing in slowly. Close cap4. Repeat for a total of two puffsFailure to prime device/load doseFailure to open mouthpiece capNo exhalationMouthpiece vents blockedFiring device before start of inhalationFiring device at or after end of inhalationInhaling through noseStopping inhalation as device is firedFast inhalationNo breath-hold/short breath-hold

Notes:

Instruction manual (Canada);

Instruction manual (USA);

HandiHaler® SmPC;58

Crucial error for some DPIs;

Instruction manual (USA);

Respimat® SmPC.59 Data from Newman.60,61

Abbreviations: DPI, dry powder inhaler; ICS, inhaled corticosteroids; pMDI, pressurized metered dose inhaler; SMI, Soft Mist™ Inhaler; SmPC, summary of product characteristics.

Requirements of the “ideal” inhaler

Characteristics of inhaler devices that affect patients’ perceptions of their COPD therapy include perceived efficacy, ease and convenience of use, how they will feel about using the device in public, their physician’s preference, availability of the drug or device preparations, loyalty to the brand of inhaler, cost, time it takes to learn how to use the device, device appearance (size, weight, etc), how to clean the device, and disposability/environmental issues.35 The ideal inhaler would be small and breath-activated; it would also deliver flow-independent drug deposition in the lung and be suitable for use in patients who have low inspiratory airflow.36 It should be lightweight and require no accessories, external assistance, or power source to enable its use. In addition, it should have the capability to count the number of delivered and/or residual doses; not allow degradation of the active ingredient; contain no additives, propellants, or excipients; and be able to quickly aerosolize the drug.37 The following are some of the factors associated with inhaler design and use that can influence how patients interact with their inhalers.

Aerosolizing

Conversion of the drug from a compacted to an inhalable form requires energy, which is either stored within the device (eg, pressurized canister, electronics, or spring) or provided by the airflow from the inhalation process. Drug delivery is affected by the inspiratory flow rate required to generate drug particles of a suitable size for inhalation (eg, 1–5.8 µm) and the fine particle fraction of the aerosol (defined as particles ≤5.8 µm).32,38 These factors vary between inhaler devices (Table 1). Aerosolizing by pMDIs and the SMI is less dependent on patient inspiratory effort than for DPIs that require a high inspiratory flow rate to ensure correct inhalation and deposition of the drug into the lungs.22 This has advantages for patients with impaired lung function. However, inspiratory flow rates vary between DPIs, depending on the internal resistance of the device, eg, the HandiHaler® (Boehringer Ingelheim International GmbH) is a flow-limiting device, with target flow between 20–30 L/min, compared with reported mean flow rates of 82 mL for the Turbuhaler® DPI (AstraZeneca) and 117 mL for the Diskhaler® DPI (GlaxoSmithKline plc, London, UK; reflecting higher internal resistance).39 To produce fine aerosol particles that are inhalable and maximize drug deposition in the lung,32,38 the target minimum inspiratory flow rates are <50 L/min for a high-resistance DPI, compared with 50–60 mL for a medium-resistance DPI and 90 L/min for a low-resistance DPI.5 Inhaler design also affects the extent to which the aerosol contains fine particles. A high fine particle fraction allows more of the drug dose to be deposited in the lung, rather than in the oropharynx.32 It has been shown that the fine particle fraction of the aerosol generated using the SMI is approximately 75%, which is more than twice the value reported for pMDIs or DPIs.38 However, other new generation inhalers, such as the NEXThaler® DPI (Chiesi Limited), have evolved to deliver extrafine drug particles that can increase drug deposition throughout the bronchial tree.5,40 All methods of aerosolizing require the use of carriers and excipients, whether they are gaseous, liquid, or solid. Although they are pharmacologically inert, they can produce a cold sensation that may cause the patient to stop inhaling.24 SMIs do not require propellants,41 and excipients are minimized. The delivery of the drug as a mist aerosol avoids the cool sensation from evaporating gases that can be experienced with pMDIs.16 This may be beneficial for some patients.

Delivery

After the patient aerosolizes the drug into their mouth, their inhalation flow must carry the aerosolized drug into the lungs – a process highly influenced by the patient. Failure of the drug to be effectively delivered into the lungs can result in a diminished dose as well as the potential for local side effects (eg, sore throat) from oropharyngeal deposition of the drug.42,43 In addition, drug deposited in the oropharynx may be swallowed, leading to gastrointestinal absorption, which may contribute to systemic side effects. Ideally, drug reaching the lungs is targeted to the airways, where bronchodilatory and/or anti-inflammatory effects will provide the greatest benefit, with minimal alveolar deposition, since this is the main site of systemic absorption in the lungs.44 Advancements in inhaler design have greatly improved drug delivery and have allowed lower nominal doses of drug to be used to achieve control in asthma and COPD, thus minimizing the risk of overdosing.33 For example, the SMI Respimat® delivers a slow-moving spray cloud that is of a longer duration and smaller droplet size than the spray from a pMDI, which results in deposition of a higher proportion of the drug in the lungs compared with pMDIs and some DPIs.33

Dependability

Frequent cleaning of devices is often necessary16 since the drug can be deposited on the device and accumulation is inevitable. Nozzles can easily become blocked without regular maintenance. Microbial contamination is also possible within inhalers and spacer devices.45,46 Cleaning regimens must be as simple as possible to encourage compliance with this important component of inhaler use and to ensure operational reliability of the device. Perceived efficacy of an inhaler is an important factor influencing patients’ device preferences.35 Reassurance from device feedback interfaces such as dose counters has been shown to be important in improving treatment adherence, reminding patients when to replace cartridges, and preventing patients from using their inhalers beyond the recommended number of doses.5,11

Matching inhalers to the patient’s physical ability

The authors of this review and others have previously proposed an algorithm for the selection of a particular inhaler device based on the patient’s physical abilities (Figure 4).12 The algorithm considers three questions: 1) Is the patient capable of conscious inhalation? 2) Is the patient likely to reliably generate and control sufficient inspiratory flow? 3) Is the patient capable of hand–inhalation coordination?
Figure 4

Algorithm for choosing inhaler device according to the patient’s inspiratory flow and ability to coordinate inhaler actuation and inspiration.

Note: Reprinted from Respir Med, 107/12, Dekhuijzen PN, Vincken W, Virchow JC, et al, Prescription of inhalers in asthma and COPD: towards a rational, rapid and effective approach, 1817–1821, Copyright 2013, with permission from Elsevier.12

Abbreviations: BA-pMDI, breath-actuated pressurized metered dose inhaler; DPI, dry powder inhaler; pMDI, pressurized metered dose inhaler; SMI, Soft Mist™ Inhaler.

Using this algorithm, patients evaluated as having sufficient inspiratory flow with good coordination would be suited to a pMDI (with or without spacer), DPI, or SMI (ie, Respimat®, Boehringer Ingelheim); those with poor coordination may require a breath-actuated pMDI with spacer, DPI, or SMI.12 For patients with insufficient inspiratory flow but adequate coordination (eg, patients with severe COPD or asthma and recurrent exacerbations), a pMDI (with or without spacer) or SMI would be suitable.12 Options for patients with both insufficient inspiratory flow and poor inhaler coordination include a breath-actuated pMDI (with spacer), SMI, or nebulizer.12 According to the algorithm, the attributes of the SMI make it an appropriate choice for all patients who are capable of conscious inhalation, irrespective of their hand–inhalation coordination or ability to generate sufficient inspiratory flow.

Components of patient preference and satisfaction with inhalers

As with all treatment decisions, it is important that the patient be involved in the choice of inhaler device. Patients are more likely to use a device effectively if they are comfortable with it and can use it even when they are incapable of achieving high inspiratory flow rates.30 Many studies have investigated patients’ preferences for different inhaler types.31,35,41,47–50 While acknowledging that many preference studies are sponsored by the device or respiratory drug companies, such studies can provide valuable insights into the features that patients consider to be most important in a device. These include simplicity and convenience (eg, size, durability), and experience of use (eg, taste, side effects).9,35

Simplicity and convenience

Ease of use is an important factor in adherence with any device, but it is particularly important with drug delivery devices for maintenance treatment.5 Use of electronics and other automations can reduce dependency on the patient for accurate technique, but it adds to the cost and complexity in terms of potential faults and possible cleaning issues. In several surveys of device preference, ease of use has been ranked by patients as one of the most important features;47,48,51,52 however, perception of ease of use varies among patients. Patient perception of the level of convenience that is offered by the device might take into account its size, the instructions for use, its durability, the ease of cleaning the device, how comfortable it is to use (eg, for arthritic hands), and its portability.48 Patients want procedures to be as simple as possible, and one study found that 18% of patients stopped inhaled therapy spontaneously due to perceived complexity.53 Complexity and confusion between required techniques may ensue when mixed types of inhaler devices are prescribed for the delivery of different drug therapies.34

Experience

Overall experience with an inhaler may be influenced by aspects unrelated to its physical form or functionality. Side effects from the drug, excipient, or inhaler use can affect the patient’s experience and willingness to continue using the device. Common issues with inhalers are the presence of an aftertaste, an effect on the throat (such as cold, dry, or sore), and cough. Poor delivery and/or slow onset of action can reinforce a patient’s decision to voluntarily discontinue their regular medication. Indeed, lack of perceived benefit has been reported to account for 30% of patients with COPD intentionally discontinuing their therapy.11

Apprehension regarding switching device

Patients may be apprehensive when switching to a new inhaler device because they may need to learn new skills and put these into practice with only limited training.16

Experience with SMI versus other inhaler types

A number of studies comparing the patients’ experience of using SMI with other inhaler devices illustrate the influence of some of the factors already described in this section. Hodder and Price35 have reviewed patient preference studies of Respimat®. The device was shown to be consistently well accepted and preferred by patients with COPD and asthma, with almost three-quarters of patients expressing a preference for Respimat® over a pMDI or Turbuhaler® DPI (AstraZeneca; Figure 5).35 In another trial of device switching in patients with COPD, 21 of 29 patients (72%) considered Respimat® to be easier to use than HandiHaler®; no patient thought HandiHaler® was easier to use than Respimat®.41
Figure 5

Proportions of patients indicating preference for Respimat® SMI versus alternative inhaler devices in three studies using the Patient Satisfaction and Preference Questionnaire.

Notes: Respimat® SMI versus (A) pMDI, (B) Turbuhaler®, (C) Diskus®. Republished with permission of Dove Medical Press, from Patient preferences for inhaler devices in chronic obstructive pulmonary disease: experience with Respimat Soft Mist inhaler, Hodder R, Price D, Int J Chron Obstruct Pulmon Dis, 4, copyright 2012;35 permission conveyed through Copyright Clearance Center, Inc.

Abbreviations: pMDI, pressurized metered dose inhaler; SMI, Soft Mist™ Inhaler.

Features relating to ease of handling (irrespective of patient age) are commonly reported as reasons for patients preferring Respimat® over other devices.35 In the two aforementioned studies in which patients rated their satisfaction with Respimat® compared with a pMDI or Turbuhaler® DPI, Respimat® consistently scored significantly higher on ease of handling during use. Durability of the device and numerous aspects of device performance were also rated better with Respimat® than with pMDI or Turbuhaler®. Size was the only parameter on which Respimat® scored significantly worse than pMDI or Turbuhaler® in these studies; however, ease of carrying the device was not rated significantly worse, and there was no negative impact of size on ease of handling, which, as noted above, was rated higher for Respimat® in both studies.48,54 One survey of patients with COPD suggested that Respimat® had a significantly milder aftertaste than HandiHaler®, but there were no significant differences between the incidence of surveyed adverse events, including dry mouth and throat-related side effects.55 In another survey, Respimat® was associated with increased incidence of cough within the first 4 weeks of switching from HandiHaler®; however, this was transient and most patients overcame having a cough as they got used to using Respimat®.41 Satisfaction with performance of tiotropium Respimat®, including feeling of inhalation and reliable working of the inhaler, was shown in a patient preference study to be consistently better than that of a DPI (Turbuhaler®, AstraZeneca).48 In a survey of patients with COPD that evaluated their handling experiences and preferences 8 weeks after switching from tiotropium HandiHaler® to Respimat®, 46% of patients stated that they preferred Respimat®.55 In a follow-up study, 2–3 years later, the proportion of patients who expressed a preference for Respimat® had significantly increased to 80% (P<0.0001).55 The different responses obtained during the first and second surveys may suggest that the change in inhaler device was stressful for the patients. This highlights the importance of involving patients in their treatment decisions to ensure that they are comfortable with, and understand the need for, any change in treatment.

Identifying the noncompliant patient

Improved compliance and adherence to therapy is a quest that is common to many chronic conditions. It is particularly important among patients with COPD in order to slow disease progression, maintain quality of life, and reduce mortality. Tools such as patient questionnaires and decision algorithms that can be easily adapted and implemented within daily practice may help physicians to select the devices most suited to their patients’ needs and abilities; they may also help to identify patients who have the potential for poor compliance or incorrect inhalation technique. Involving the patient in their treatment decision-making process can change the physician–patient relationship from one of noncompliance to one of participative adherence to treatment.47 But how can clinicians identify patients who are likely to be noncompliant with their inhaler therapy? Identification of patients most likely to be noncompliant with therapy may enable more effective targeting of educational and behavioral resources, including continued assessment of inhaler technique, to foster improved compliance and maximize treatment efficacy. Sanduzzi et al11 evaluated several methods that can assess whether a patient will be compliant with their prescribed inhaler; they proposed that simple, brief questionnaires based on the Morisky scale56 were most effective (Table 3).11 These questionnaires address patients’ attitudes toward their medications and their health, the regularity of their medical check-ups, and their home life situation.11 Similar preference questionnaires could be refined for ease of use in consultation rooms to gain a better understanding of individual patient preference and prescribing needs.
Table 3

Proposed psychosocioeconomic questions to ask patients with COPD to estimate compliance and to assess adherence phenotype and status

Questions for a treatment naïve patient prior to prescribing inhaled therapyQuestions for patients already under treatment to confirm adherence to inhaled therapy
How many different medications do you take every day?Who do you live with?Are you worried?aWould you stop taking a medication if it gave you side effects or if it was too complicated to take?What do you think about inhaled medications?Do you see any pulmonologists for check-up examinations?How often did you forget to take your inhaled therapy in the past month?Are you more likely to forget taking your medicines in the morning or the evening?Do you find your therapy dangerous?Do you prefer inhaled drugs to oral tablets?Do you think your prescription is appropriate for your disease?

Notes: Each question has a defined set of four responses. Responses are scored using the Morisky scale to give a general indicator of risk.11 See Sanduzzi et al11 for the complete questionnaires.

Question relates to the patient’s health condition, and economic and family situations. Data from Sanduzzi et al.11

Abbreviation: COPD, chronic obstructive pulmonary disease.

Conclusion

This review highlights the importance of understanding patient capabilities and preferences, and matching these to the most appropriate type of inhaler device; it also stresses the need to adopt ongoing training and monitoring of inhaler technique. With the large array of devices currently available to the prescribing physician, it is increasingly important to understand the factors that influence a patient’s preference and the likelihood of the inhaler being used correctly (eg, ease of use and convenience). Respimat® represents a novel class of inhaler (SMI) that overcomes some of the drawbacks associated with pMDIs and DPIs, such as nontargeted deposition of drug. It might be expected that SMIs in development will perform in a similar maner to Respimat® with regards to what benefits patients. Patient satisfaction and preference studies have consistently shown Respimat® to be well accepted by patients, with higher levels of satisfaction and preference versus HandiHaler® and other device types. Whether the observed preference for Respimat® ultimately translates into improved adherence to therapy and improved outcomes remains to be evaluated in clinical studies.
  56 in total

1.  Accuracy of three electronic monitors for metered-dose inhalers.

Authors:  Steven M Julius; James M Sherman; Leslie Hendeles
Journal:  Chest       Date:  2002-03       Impact factor: 9.410

2.  Results of a programme to improve house staff use of metered dose inhalers and spacers.

Authors:  M Lee-Wong; P H Mayo
Journal:  Postgrad Med J       Date:  2003-04       Impact factor: 2.401

Review 3.  Pulmonary drug delivery. Part II: the role of inhalant delivery devices and drug formulations in therapeutic effectiveness of aerosolized medications.

Authors:  N R Labiris; M B Dolovich
Journal:  Br J Clin Pharmacol       Date:  2003-12       Impact factor: 4.335

4.  Respimat Soft Mist inhaler versus hydrofluoroalkane metered dose inhaler: patient preference and satisfaction.

Authors:  Wolfgang Schürmann; Sören Schmidtmann; Petra Moroni; Dan Massey; Mahmud Qidan
Journal:  Treat Respir Med       Date:  2005

Review 5.  Device selection and outcomes of aerosol therapy: Evidence-based guidelines: American College of Chest Physicians/American College of Asthma, Allergy, and Immunology.

Authors:  Myrna B Dolovich; Richard C Ahrens; Dean R Hess; Paula Anderson; Rajiv Dhand; Joseph L Rau; Gerald C Smaldone; Gordon Guyatt
Journal:  Chest       Date:  2005-01       Impact factor: 9.410

Review 6.  What plays a role in the choice of inhaler device for asthma therapy?

Authors:  J Christian Virchow
Journal:  Curr Med Res Opin       Date:  2005       Impact factor: 2.580

7.  Effective delivery of particles with the HandiHaler dry powder inhalation system over a range of chronic obstructive pulmonary disease severity.

Authors:  S Chodosh; J S Flanders; S Kesten; C W Serby; D Hochrainer; T J Witek
Journal:  J Aerosol Med       Date:  2001

Review 8.  Suboptimal medical therapy in COPD: exploring the causes and consequences.

Authors:  S D Ramsey
Journal:  Chest       Date:  2000-02       Impact factor: 9.410

9.  Inappropriate inhaler use: assessment of use and patient preference of seven inhalation devices. EDICI.

Authors:  J Lenney; J A Innes; G K Crompton
Journal:  Respir Med       Date:  2000-05       Impact factor: 3.415

10.  Meeting the needs of patients with COPD: patients' preference for the Diskus inhaler compared with the Handihaler.

Authors:  A C Moore; S Stone
Journal:  Int J Clin Pract       Date:  2004-05       Impact factor: 2.503

View more
  16 in total

1. 

Authors:  Lee Fidler; Samantha Green; Kimberly Wintemute
Journal:  CMAJ       Date:  2022-06-13       Impact factor: 16.859

2.  ELLIPTA Versus DISKUS plus HandiHaler in COPD: A Randomized, Open-Label, Crossover Study in a Clinical Trial Setting.

Authors:  Edward M Kerwin; Selwyn Spangenthal; Michael Zvarich; Valerie Millar; Renu Jain; Kathryn Collison; Raj Sharma
Journal:  Chronic Obstr Pulm Dis       Date:  2020-04

3.  Therapeutic Success of Tiotropium/Olodaterol, Measured Using the Clinical COPD Questionnaire (CCQ), in Routine Clinical Practice: A Multinational Non-Interventional Study.

Authors:  Arschang Valipour; Sergey Avdeev; Adam Barczyk; Valentina Bayer; Zvi Fridlender; Mariela Georgieva; Ondřej Kudela; Alexey Medvedchikov; Ramona Miron; Maria Sanzharovskaya; Virginija Šileikienė; Jurij Šorli; Marc Spielmanns; Zsuzsanna Szalai
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2021-03-10

Review 4.  Matching Inhaler Devices with Patients: The Role of the Primary Care Physician.

Authors:  Alan Kaplan; David Price
Journal:  Can Respir J       Date:  2018-05-23       Impact factor: 2.409

Review 5.  The Nurse Practitioners' Perspective on Inhaler Education in Asthma and Chronic Obstructive Pulmonary Disease.

Authors:  Jane Scullion
Journal:  Can Respir J       Date:  2018-08-05       Impact factor: 2.409

6.  Assessment of physical functioning and handling of tiotropium/olodaterol Respimat® in patients with COPD in a real-world clinical setting.

Authors:  Karl-Otto Steinmetz; Birgit Abenhardt; Stefan Pabst; Michaela Hänsel; Anke Kondla; Valentina Bayer; Roland Buhl
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2019-07-04

7.  Maintenance inhaler preference, attribute importance, and satisfaction in prescribing physicians and patients with asthma, COPD, or asthma-COPD overlap syndrome consulting for routine care.

Authors:  Bo Ding; Mark Small; Gina Scheffel; Ulf Holmgren
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2018-03-16

8.  Misuse of inhalers among COPD patients in a community hospital in Taiwan.

Authors:  Chi-Yen Liang; Yi-Jen Chen; Shew-Meei Sheu; Ching-Fang Tsai; Wei Chen
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2018-04-23

9.  Inhaler device feature preferences among patients with obstructive lung diseases: A systematic review and meta-analysis.

Authors:  Maryam Navaie; Carole Dembek; Soojin Cho-Reyes; Karen Yeh; Bartolome R Celli
Journal:  Medicine (Baltimore)       Date:  2020-06-19       Impact factor: 1.817

Review 10.  Is there room for further innovation in inhaled therapy for airways disease?

Authors:  Martyn F Biddiscombe; Omar S Usmani
Journal:  Breathe (Sheff)       Date:  2018-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.