Lung hyperaeration assessment by computed tomography: correction of reconstruction-induced bias.

Lorenzo Ball1, Claudia Brusasco2, Francesco Corradi2, Francesco Paparo3, Alessandro Garlaschi4, Peter Herrmann5, Michael Quintel5, Paolo Pelosi6.   

Abstract

BACKGROUND: Computed tomography (CT) reconstruction parameters, such as slice thickness and convolution kernel, significantly affect the quantification of hyperaerated parenchyma (VHYPER%). The aim of this study was to investigate the mathematical relation between VHYPER% calculated at different reconstruction settings, in mechanically ventilated and spontaneously breathing patients with different lung pathology.
METHODS: In this retrospective observational study, CT scans of patients of the intensive care unit and emergency department were collected from two CT scanners and analysed with different kernel-thickness combinations (reconstructions): 1.25 mm soft kernel, 5 mm soft kernel, 5 mm sharp kernel in the first scanner; 2.5 mm slice thickness with a smooth (B41s) and a sharp (B70s) kernel on the second scanner. A quantitative analysis was performed with Maluna® to assess lung aeration compartments as percent of total lung volume. CT variables calculated with different reconstructions were compared in pairs, and their mathematical relationship was analysed by using quadratic and power functions.
RESULTS: 43 subjects were included in the present analysis. Image reconstruction parameters influenced all the quantitative CT-derived variables. The most relevant changes occurred in the hyperaerated and normally aerated volume compartments. The application of a power correction formula led to a significant reduction in the bias between VHYPER% estimations (p < 0.001 in all cases). The bias in VHYPER% assessment did not differ between lung pathology nor ventilation mode groups (p > 0.15 in all cases).
CONCLUSIONS: Hyperaerated percent volume at different reconstruction settings can be described by a fixed mathematical relationship, independent of lung pathology, ventilation mode, and type of CT scanner.

Entities:  

Mesh:

Year:  2016        PMID: 27553378      PMCID: PMC4995787          DOI: 10.1186/s12871-016-0232-z

Source DB:  PubMed          Journal:  BMC Anesthesiol        ISSN: 1471-2253            Impact factor:   2.217


Background

Lung quantitative computed tomography (CT) is an effective method for the evaluation of lung aeration, especially in the mechanically ventilated patients, since the quantification of over and non-aerated lung might be useful for the optimization of the ventilatory strategy [1]. Tidal recruitment and hyperaeration have been proposed as the main mechanisms yielding Ventilator Induced Lung Injury [2, 3], and can also occur at low tidal volume ventilation [3]. Mechanical ventilation should be individually titrated in order to avoid lung over distension, maintaining an acceptable gas exchange [4, 5]. However, the potential future of CT assessment of hyperaeration in clinical and research applications is still hampered by technical issues; in particular: 1) the optimal attenuation threshold separating normally from over-inflated lung parenchyma in mechanically ventilated patients is matter of debate [6]; 2) the reconstruction parameters, such as slice thickness and convolution kernel, significantly affect CT assessment of overinflated areas in both chronic emphysema [7] and critically ill patients [8]. Recent studies found a relevant amount of hyperaerated lung parenchyma among healthy patients [9], as well as a dramatic effect of reconstruction parameters on the evaluation of severe emphysema [10]. Thus, a standardized and reproducible evaluation of hyperaeration by CT is warranted. A new method has been recently proposed to correct the discrepancy in hyperaeration assessment at different CT reconstruction parameters in spontaneously breathing patients with chronic emphysema [11]. The aim of the present study was to develop a technique to standardize hyperaeration assessment in spontaneously breathing and mechanically ventilated patients with various underlying lung conditions, using different reconstruction settings and two different CT scanners. We hypothesized that hyperaerated percent volume at different reconstruction settings could be described by a fixed mathematical relationship, independent of lung pathology and ventilation mode.

Methods

This was a retrospective observational study, reported according to the Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) statement [12].

Population

Patients of the Intensive Care Unit and the Emergency Department of a single university hospital in Genova, Italy, that underwent chest CT scan for clinical reasons between September 1st and November 31st 2012 were retrospectively screened for inclusion analysing the scanners’ database. The inclusion criteria were the following: 1) patient age ≥ 18 years; 2) standard helical thorax acquisition protocols used; 3) no contrast medium used and 4) no relevant motion artefacts.

CT image acquisition and analysis

CT scans were collected from two different scanners, each patient underwent a single CT scan with either scanner A or B. Scanner A was a LightSpeed 16 (GE Medical Systems, Milwaukee, US), located in the emergency department, set with 120 kVp, field of view according to clinical needs (range 360–500), pitch factor 1.75, collimation 16 × 0.625 mm. The images of scanner A were reconstructed in axial DICOM series with three different kernel-thickness combinations (reconstructions): 1.25 mm soft kernel, 5 mm soft kernel, 5 mm sharp kernel. Smooth and sharp filters were the GE proprietary Lung and Body filters, respectively. Scanner B was a Somatom Emotion 6 (Siemens, Munich, Germany), located in the radiology department, and settings were: 110 kVp, Pitch Factor 0.85, Slice Collimation 6 × 2.0 mm. The images of scanner B were reconstructed at 2.5 mm slice thickness with a smooth (B41s) and a sharp (B70s) kernel. Scans were executed in accordance to the routine practice of our Radiology Unit: during breath-hold at full inspiration in spontaneously breathing patients, or during uninterrupted ventilation in mechanically ventilated patients. Image segmentation was performed with Maluna® software (University of Mannheim, Göttingen, Germany) by means of a semi-automatic method with further manual refinement. Big vessels and airways larger than 5 mm were excluded from segmentation. Scans were segmented on all slices of each reconstruction by three physicians (LB, CB, and FC) and then revised by a radiologist (AG). For each reconstruction, quantitative analysis was performed to obtain Total Lung Volume (TLV), mean lung attenuation in Hounsfield Units (HU), Total Lung Weight and the following aeration compartments [13] as % of TLV: hyper-aerated volume (VHYPER%, -1000 to -901 HU), normally aerated volume (VNORMAL%, -900 to -501 HU), poorly aerated volume (VPOOR%, -500 to -101 HU), non-aerated volume (VNON %, -100 to +100 HU). CT variables calculated with the different CT reconstruction parameters were compared in pairs. For each reconstruction pair, VHYPER% was calculated at the two reconstructions and the mathematical relationship between the two was analysed by using a quadratic and power function according to the corrected Akaike information criterion (AICc), as previously described [11]. Patients were divided in three pathology groups according to the main finding of the CT scan: no pathological findings in lung parenchyma of a spontaneously breathing or mechanical ventilated patient (Healthy), hyperaerated lung parenchyma and findings compatible with chronic obstructive pulmonary disease (COPD), evidence of lung injury or aeration loss (Injury). Patients with COPD exacerbation were included in the Injury group due to the relevant amount of non-aerated lung volume.

Statistical analysis

Data are presented as mean ± standard deviation where not otherwise specified; absolute volumes, weights and mean HU were rounded to the nearest integer. Normality was assessed by D’Agostino–Pearson omnibus test and comparisons were made by repeated measures t-test and repeated measures ANOVA with Tukey post-hoc or Wilcoxon tests and Friedman test with Dunn post-hoc, accordingly. Independent samples were compared by t-test or Kruskal-Wallis test. Bias between measurements of VHYPER% with the two reconstructions was compared before and after the application of such correction by Bland-Altman method. Expecting a very strong correlation (R ≥ 0.8) between VHYPER calculated at different reconstructions [11], we needed to enrol at least 9 subjects from each scanner to achieve 90 % power (1-β) to observe such correlation at an alpha level of 0.05. Statistical analysis was performed with SPSS version 21 (IBM Corp.), function fit and graphs with Prism 6 (GraphPad Software Inc., San Diego, CA). Statistical significance was considered for p < 0.05, all tests were two-tailed.

Results

In the observed period, 178 chest CT scans were performed. Among them, 135 were excluded: 5 patients were minors, 28 scans used non-standard acquisition protocols, 89 required the use of contrast medium and 13 reported unacceptable motion artefacts. As a result, 43 scans were included in the final analysis: 33 from scanner A and 10 from scanner B. The resulting population is described in Table 1.
Table 1

Patient data

Patient Data
Age, y (IQR)62 (47–71)
Sex (%)27 Male (62.8), 16 Female (37.2)
Lung CT indication (%)Respiratory Failure27 (62.8)
Polytrauma6 (14.0)
Other10 (23.3)
Mode of ventilation (%)Mechanically Ventilated15 (34.9)
Spontaneously Breathing28 (65.1)
Main lung CT finding (%)Injury23 (53.4)
COPD10 (23.3)
Healthy10 (23.3)
Patient data Image reconstruction parameters influenced all the quantitative CT-derived variables (see Table 2). The most relevant changes occurred in the VHYPER% and VNORMAL% volume compartments, while differences in Total Lung Volume, Lung Density, Lung Weight, VPOOR% and VNON % were limited in magnitude. As shown in Fig. 1, differences in VHYPER% were associated to opposite changes in VNORMAL% and the sum of the two compartments was not influenced by reconstruction parameters (p > 0.05 for all reconstruction pairs).
Table 2

Mean results of CT quantitative analysis at different slice thickness-kernel combinations for GE LightSpeed scanner and Siemens Emotion

GE LightSpeed
ReconstructionTotal Volume (ml)EI950%a VHYPER %a VNORMAL %a VPOOR %a VNON %Density (HU)a Lung Weight (g)a
1.25 mm Body 4021 ± 14941.0 ± 1.812.6 ± 14.372.7 ± 15.56.9 ± 5.07.0 ± 13.0−721 ± 1441013 ± 435
5.00 mm Body 4037 ± 15000.3 ± 0.78.2 ± 11.877.3 ± 15.67.2 ± 5.36.6 ± 13.2b −713 ± 1461049 ± 464
5.00 mm Lung 3943 ± 1451b 4.4 ± 4.320.3 ± 16.064.6 ± 13.97.6 ± 5.26.6 ± 11.7−728 ± 139961 ± 374
Siemens Somatom
ReconstructionTotal Volume (ml)EI950%a VHYPER %a VNORMAL %a VPOOR %VNON %Density (HU)a Lung Weight (g)a
2.50 mm B41s 3224 ± 12640.3 ± 0.45.0 ± 5.673.4 ± 14.814.3 ± 11.47.0 ± 7.0−663 ± 136960 ± 252
2.50 mm B70s 3203 ± 12762.9 ± 2.312.0 ± 8.466.2 ± 11.614.7 ± 11.36.6 ± 6.1−671 ± 130933 ± 256

Repeated measures pairwise comparisons: aall pairwise comparisons between reconstructions significant (p < 0.05), bsignificantly different compared pairwise to both other reconstructions (p < 0.05)

Fig. 1

Differences between volume compartments assessment in different reconstruction pairs. GE LightSpeed: 1B (1.25 mm Body smooth kernel) 5B (5 mm Body smooth kernel) 5 L (5 mm Lung sharp kernel). Siemens Emotion: 2.5 mm sharp (B70s) and 2.5 mm smooth (B41s)

Mean results of CT quantitative analysis at different slice thickness-kernel combinations for GE LightSpeed scanner and Siemens Emotion Repeated measures pairwise comparisons: aall pairwise comparisons between reconstructions significant (p < 0.05), bsignificantly different compared pairwise to both other reconstructions (p < 0.05) Differences between volume compartments assessment in different reconstruction pairs. GE LightSpeed: 1B (1.25 mm Body smooth kernel) 5B (5 mm Body smooth kernel) 5 L (5 mm Lung sharp kernel). Siemens Emotion: 2.5 mm sharp (B70s) and 2.5 mm smooth (B41s) For each reconstruction pair, the relationship between VHYPER% calculated in the two reconstructions was adequately described by both quadratic and power functions, and the preferred model according to the AICc was the power function (y = a · xb) in all cases, as shown in Table 3. The application of the correction formula led to a significant reduction in the bias between reconstructions in VHYPER% estimation (p < 0.001 in all cases), as shown in Fig. 2. Figure 3 illustrates the effect of reconstruction of a representative subject with injured lung, whose CT was acquired with scanner A and reconstructed with three different kernel-thickness combinations. The bias between reconstructions in VHYPER% assessment did not differ between lung pathology nor ventilation mode groups (see Table 4).
Table 3

Results of function fit. Parameters reported as mean ± standard error

Reconstruction 1Reconstruction 2Best ModelabR2
GE LightSpeed5 mm body1.25 mm body power function 3.66 ± 0.120.69 ± 0.010.997
5 mm body5 mm lung power function 11.60 ± 0.580.41 ± 0.020.977
1. 25 mm body5 mm lung power function 5.64 ± 0.390.58 ± 0.020.982
Siemens Emotion2.5 mm B70s2.5 mm B41s power function 6.58 ± 0.850.48 ± 0.060.951
Fig. 2

Fit results (left panels), agreement before (middle panels) and after (right panels) application of correction. Each panel row represents one reconstruction pair. GE LightSpeed 16: 1B (1.25 mm Body smooth kernel) 5B (5 mm Body smooth kernel) 5 L (5 mm Lung sharp kernel). Siemens Emotion 6: 2.5 mm sharp (B70s) and 2.5 mm smooth (B41s). Dot shapes indicates ventilation mode: spontaneously breathing (squares) and mechanically ventilated subjects (circles)

Fig. 3

Effects of changes of slice thickness and kernel in distribution and amount of VHYPER% (red mask). Hyperaeration calculated with -950 HU as threshold is also shown (yellow mask)

Table 4

Bias between reconstructions in VHYPER% after correction, in the different lung pathology and ventilation mode groups

Reconstruction 1Reconstruction 2InjuredCOPDHealthyp
GE LightSpeed5 mm body1.25 mm body0.2 ± 0.6−0.1 ± 0.9−0.3 ± 0.70.27
5 mm body5 mm lung0.2 ± 2.40.4 ± 1.7−1.6 ± 2.80.18
1. 25 mm body5 mm lung0.3 ± 1.90.5 ± 2.0−0.8 ± 2.70.39
Siemens Emotion2.5 mm B70s2.5 mm B41s0.2 ± 2.1−0.3 ± 1.4−1.8 ± 0.80.45
Reconstruction 1Reconstruction 2Spontaneously BreathingMechanically Ventilatedp
GE LightSpeed5 mm body1.25 mm body0.1 ± 0.8−0.1 ± 0.70.42
5 mm body5 mm lung−0.4 ± 2.50.2 ± 2.30.54
1. 25 mm body5 mm lung−0.3 ± 2.10.8 ± 2.00.15
Siemens Emotion2.5 mm B70s2.5 mm B41s−0.5 ± 2.20.3 ± 0.40.38
Results of function fit. Parameters reported as mean ± standard error Fit results (left panels), agreement before (middle panels) and after (right panels) application of correction. Each panel row represents one reconstruction pair. GE LightSpeed 16: 1B (1.25 mm Body smooth kernel) 5B (5 mm Body smooth kernel) 5 L (5 mm Lung sharp kernel). Siemens Emotion 6: 2.5 mm sharp (B70s) and 2.5 mm smooth (B41s). Dot shapes indicates ventilation mode: spontaneously breathing (squares) and mechanically ventilated subjects (circles) Effects of changes of slice thickness and kernel in distribution and amount of VHYPER% (red mask). Hyperaeration calculated with -950 HU as threshold is also shown (yellow mask) Bias between reconstructions in VHYPER% after correction, in the different lung pathology and ventilation mode groups

Discussion

The main findings of the present study are: 1) CT reconstruction mainly affected the estimation of VHYPER% and VNORMAL%; 2) the bias between reconstructions is mathematically determined and can be described by a power function; 3) such mathematical relationship can be used to correct the bias between two reconstructions, independent of lung pathology and ventilation mode. In the last decade, several techniques have been introduced and investigated to assess at the bedside lung aeration, including lung ultrasound (LUS) [14, 15] and electrical impedance tomography (EIT) [16]. The latter is able to provide an estimation of hyper-aeration [17]. Nonetheless, CT remains the gold standard because of its strong correlation with physical density [15], and in the last decades it provided precious pathophysiological information in acute respiratory distress syndrome (ARDS) [18], therefore studies validating emerging monitoring techniques such as LUS and EIT often use CT as reference [19]. Thus, a standardization of hyperaeration measurement by CT is mandatory allowing the analysis and comparison of previously acquired datasets. The present study shows that hyperaeration measured with a given slice thickness-kernel combination can be estimated using data from a different CT reconstruction. To our knowledge, this is the first study attempting to correct hyperaeration assessment bias in injured lung and during mechanical ventilation. A similar approach was proposed for spontaneously breathing chronic emphysema patients [11]. Using a comparable method, we found a similar performance on reconstruction bias correction in two different manufacturer scanners and in patients with different lung findings as well as type of ventilation. The discrepancy between quantitative CT analyses was primarily found in the hyperaerated and the normally aerated regions of the lung, as previously described [7, 8]. Thus, a single correction formula for VHYPER% can virtually completely eliminate bias between reconstructions, being applied directly to VHYPER%, and by subtraction to VNORMAL%. The hyperaeration assessment was initially studied by pulmonologists for staging chronic emphysema [20]. In this domain, though not unanimous, there is agreement that specific reconstructions, such as smooth B30s kernel wit 1 mm slice thickness or comparable, should be preferred to others due to their good correlation with quantitative histology [21, 22], using a threshold of -950 HU to define hyperaerated regions [11, 23]. In intensive care applications of lung CT analysis, quantification of hyperaeration is a functional and physiological rather than an anatomical assessment [6], thus a direct comparison with a morphological characteristic of the lung is not viable, and a threshold of -900 HU was usually employed [1, 4, 9, 13]. Due to the lack of a gold standard for comparison, safe amounts of hyperaeration at CT analysis are difficult to be determined [3, 4]. The results of the present work show that VHYPER % calculated with different reconstructions can be compared once corrected by a mathematical formula. Other studies proposed to extrapolate quantitative CT analysis from a limited number of slices, potentially reducing dose exposure [24, 25]. We propose a mathematical allowing switching to thicker or thinner slices as needed without affecting VHYPER% measurement. Applications of our method has relevance mainly in the research field, with potential indirect relevant clinical implications. Efforts should be made in order to obtain a standard CT assessment of hyper-aeration, as in recent experimental studies it was correlated with an increased neutrophilic inflammation in a model of ARDS [26]. CT scans performed for clinical reasons in critically ill patients often require the administration of contrast medium: this caused the exclusion of half of the screened patients in the present cohort. Due to the minimal perfusion of hyper-aerated lung regions, one could hypothesise that limited changes will occur in this compartment due to the presence of iodine; however, the validity of our method in CT scans acquired with contrast medium remains to be tested. Our study has some limitations to be addressed. First, the retrospective design of the study led to limited possibility of controlling the correct execution of the chest scan, which was evaluated a posteriori by visual inspection of the images. Second, mechanically ventilated patients were not examined during a respiratory hold leading to artefacts and possibly inaccurate absolute estimation of VHYPER%. However, since in this study no comparison was performed between subjects but only on different CT scan reconstructions of the same patient, the mentioned limits should not affect the validity of the results. Third, reported correction formulas cannot be directly applied on other scanners, for which further calibration would be necessary to apply externally the findings of our study. Moreover, this technique, while able to reduce notably the discrepancy between VHYPER% estimation with two different reconstructions, is unable to clarify which specific reconstruction settings should be chosen: further studies are warranted to identify those that better correlate with histological findings, inflammation markers and, potentially, clinical outcomes.

Conclusions

In quantitative lung CT, the relationship between hyperaerated percent volume at different slice thickness-convolution kernel settings can be described by a single function, independent of lung pathology, ventilation mode. This approach was feasible in the two different CT scanners used in the present study.
  26 in total

Review 1.  What has computed tomography taught us about the acute respiratory distress syndrome?

Authors:  L Gattinoni; P Caironi; P Pelosi; L R Goodman
Journal:  Am J Respir Crit Care Med       Date:  2001-11-01       Impact factor: 21.405

2.  Equating quantitative emphysema measurements on different CT image reconstructions.

Authors:  Seth T Bartel; Andrew J Bierhals; Thomas K Pilgram; Cheng Hong; Kenneth B Schechtman; Susan H Conradi; David S Gierada
Journal:  Med Phys       Date:  2011-08       Impact factor: 4.071

Review 3.  Monitoring respiration: what the clinician needs to know.

Authors:  Lorenzo Ball; Yuda Sutherasan; Paolo Pelosi
Journal:  Best Pract Res Clin Anaesthesiol       Date:  2013-06

4.  Tidal hyperinflation during low tidal volume ventilation in acute respiratory distress syndrome.

Authors:  Pier Paolo Terragni; Giulio Rosboch; Andrea Tealdi; Eleonora Corno; Eleonora Menaldo; Ottavio Davini; Giovanni Gandini; Peter Herrmann; Luciana Mascia; Michel Quintel; Arthur S Slutsky; Luciano Gattinoni; V Marco Ranieri
Journal:  Am J Respir Crit Care Med       Date:  2006-10-12       Impact factor: 21.405

5.  Bedside ultrasound assessment of positive end-expiratory pressure-induced lung recruitment.

Authors:  Belaïd Bouhemad; Hélène Brisson; Morgan Le-Guen; Charlotte Arbelot; Qin Lu; Jean-Jacques Rouby
Journal:  Am J Respir Crit Care Med       Date:  2010-09-17       Impact factor: 21.405

6.  Effects of CT section thickness and reconstruction kernel on emphysema quantification relationship to the magnitude of the CT emphysema index.

Authors:  David S Gierada; Andrew J Bierhals; Cliff K Choong; Seth T Bartel; Jon H Ritter; Nitin A Das; Cheng Hong; Thomas K Pilgram; Kyongtae T Bae; Bruce R Whiting; Jason C Woods; James C Hogg; Barbara A Lutey; Richard J Battafarano; Joel D Cooper; Bryan F Meyers; G Alexander Patterson
Journal:  Acad Radiol       Date:  2010-02       Impact factor: 3.173

7.  Image reconstruction affects computer tomographic assessment of lung hyperinflation.

Authors:  Andreas W Reske; Harald Busse; Marcelo B P Amato; Matthias Jaekel; Thomas Kahn; Peter Schwarzkopf; Dierk Schreiter; Udo Gottschaldt; Matthias Seiwerts
Journal:  Intensive Care Med       Date:  2008-06-08       Impact factor: 17.440

8.  Comparative Effects of Volutrauma and Atelectrauma on Lung Inflammation in Experimental Acute Respiratory Distress Syndrome.

Authors:  Andreas Güldner; Anja Braune; Lorenzo Ball; Pedro L Silva; Cynthia Samary; Angelo Insorsi; Robert Huhle; Ines Rentzsch; Claudia Becker; Liane Oehme; Michael Andreeff; Marcos F Vidal Melo; Tilo Winkler; Paolo Pelosi; Patricia R M Rocco; Jörg Kotzerke; Marcelo Gama de Abreu
Journal:  Crit Care Med       Date:  2016-09       Impact factor: 7.598

9.  Extrapolation in the analysis of lung aeration by computed tomography: a validation study.

Authors:  Andreas W Reske; Anna Rau; Alexander P Reske; Manja Koziol; Beate Gottwald; Michaele Alef; Jean-Claude Ionita; Peter M Spieth; Pierre Hepp; Matthias Seiwerts; Alessandro Beda; Silvia Born; Gerik Scheuermann; Marcelo B P Amato; Hermann Wrigge
Journal:  Crit Care       Date:  2011-11-23       Impact factor: 9.097

10.  Preliminary study of ventilation with 4 ml/kg tidal volume in acute respiratory distress syndrome: feasibility and effects on cyclic recruitment - derecruitment and hyperinflation.

Authors:  Jaime Retamal; Javiera Libuy; Magdalena Jiménez; Matías Delgado; Cecilia Besa; Guillermo Bugedo; Alejandro Bruhn
Journal:  Crit Care       Date:  2013-01-28       Impact factor: 9.097

View more
  5 in total

Review 1.  Lung imaging: how to get better look inside the lung.

Authors:  Lorenzo Ball; Veronica Vercesi; Federico Costantino; Karthikka Chandrapatham; Paolo Pelosi
Journal:  Ann Transl Med       Date:  2017-07

2.  Ultra-low-dose sequential computed tomography for quantitative lung aeration assessment-a translational study.

Authors:  Lorenzo Ball; Anja Braune; Francesco Corradi; Claudia Brusasco; Alessandro Garlaschi; Thomas Kiss; Thomas Bluth; Francesca Simonassi; Alice Bergamaschi; Jörg Kotzerke; Marcus J Schultz; Marcelo Gama de Abreu; Paolo Pelosi
Journal:  Intensive Care Med Exp       Date:  2017-04-04

3.  Magnetic Resonance Imaging for Quantitative Assessment of Lung Aeration: A Pilot Translational Study.

Authors:  Lorenzo Ball; Anja Braune; Peter Spieth; Moritz Herzog; Karthikka Chandrapatham; Volker Hietschold; Marcus J Schultz; Nicolò Patroniti; Paolo Pelosi; Marcelo Gama de Abreu
Journal:  Front Physiol       Date:  2018-08-13       Impact factor: 4.566

4.  Automatic Lung Segmentation and Quantification of Aeration in Computed Tomography of the Chest Using 3D Transfer Learning.

Authors:  Lorenzo Maiello; Lorenzo Ball; Marco Micali; Francesca Iannuzzi; Nico Scherf; Ralf-Thorsten Hoffmann; Marcelo Gama de Abreu; Paolo Pelosi; Robert Huhle
Journal:  Front Physiol       Date:  2022-02-04       Impact factor: 4.566

5.  An Experimental Pre-Post Study on the Efficacy of Respiratory Physiotherapy in Severe Critically III COVID-19 Patients.

Authors:  Denise Battaglini; Salvatore Caiffa; Giovanni Gasti; Elena Ciaravolo; Chiara Robba; Jacob Herrmann; Sarah E Gerard; Matteo Bassetti; Paolo Pelosi; Lorenzo Ball
Journal:  J Clin Med       Date:  2021-05-15       Impact factor: 4.241

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.