A R Kalukin1, B Winn2, Y Wang2, C Jacobsen2, Z H Levine3, J Fu3.
Abstract
For two-dimensional x-ray imaging of thin films, the technique of scanning transmission x-ray microscopy (STXM) has achieved images with feature sizes as small as 40 nm in recent years. However, calibration of three-dimensional tomographic images that are produced with STXM data at this scale has not yet been described in the scientific literature, and the calibration procedure has novel problems that have not been encountered by x-ray tomography carried out at a larger scale. In x-ray microtomography, for example, one always has the option of using optical imaging on a section of the object to verify the x-ray projection measurements; with STXM, on the other hand, the sample features are too small to be resolved by light at optical wavelengths. This fact implies that one must rely on procedures with higher resolution, such as atomic force microscopy (AFM), for the calibration. Such procedures, however, generally depend on a highly destructive sectioning of the sample, and are difficult to interpret because they give surface information rather than depth information. In this article, a procedure for calibration is described that overcomes these limitations and achieves a calibration of an STXM tomography image with an AFM image and a scanning electron microscopy image of the same object. A Ge star-shaped pattern was imaged at a synchrotron with a scanning transmission x-ray microscope. Nineteen high-resolution projection images of 200 × 200 pixels were tomographically reconstructed into a three-dimensional image. Features in two-dimensional images as small as 40 nm and features as small as 80 nm in the three-dimensional reconstruction were resolved. Transverse length scales based on atomic force microscopy, scanning electron microscopy, x-ray transmission and tomographic reconstruction agreed to within 10 nm. Toward the center of the sample, the pattern thickness calculated from projection images was (51 ± 15) nm vs (80 ± 52) nm for tomographic reconstruction, where the uncertainties are evaluated at the level of two standard deviations.Entities:
Keywords: atomic force microscopy; scanning electron microscope; x-ray microscopy
Year: 2000 PMID: 27551641 PMCID: PMC4877146 DOI: 10.6028/jres.105.067
Source DB: PubMed Journal: J Res Natl Inst Stand Technol ISSN: 1044-677X
Fig. 1:Scanning transmission x-ray microscopy image of Ge test pattern; color gray scale display. For all images in this article, brightness increases with measured thickness.
Fig. 2:SEM image of Ge test pattern; color gray scale display.
Fig. 3:Scanning transmission x-ray microscope image of Ge test pattern; zero tilt angle, 200 × 200, 26 nm pixels; color gray scale display.
Fig. 5:AFM image of Ge test pattern, showing contaminating particles in the inner rings; 512 × 512, 14.7 nm pixels; color gray scale display.
Fig. 4:Scanning transmission x-ray microscope image of Ge test pattern; 50° tilt angle, 200 × 200 26-nm pixels; color gray scale display.
Fig. 6:View of three-dimensional reconstruction image of test pattern, 52 nm voxels; volume isosurface display.
Fig. 7:Three-dimensional reconstructed x-ray image of contamination at center of Ge test pattern, 52 nm voxels; volume isosurface display.
Comparison of feature width and thickness measurements of Ge pattern for several imaging modalities at two regions within the pattern. All values are in nm. The assigned uncertainties are expanded uncertainties with a coverage factor of k = 2 (i.e., two standard deviation estimates)
| Method | Bar width | Thickness | ||
|---|---|---|---|---|
| Star edge (nm) | 3 μm ring (nm) | Star edge (nm) | 3 μm ring (nm) | |
| SEM | 85 ± 10 | |||
| AFM | 390 ± 60 | 95 ± 15 | 210 ± 10 | > 20 |
| 2-D x-ray | 390 ± 40 | 85 ± 40 | 140 ± 20 | 34 ± 10 |
| 2-D x-ray scaled to AFM | 390 ± 40 | 85 ± 40 | 210 ± 30 | 51 ± 15 |
| 3-D x-ray | 85 ± 52 | 80 ± 52 | ||