Lígia Brancalion Catapani1, Adriana da Costa Gonçalves2, Nathalia Morano Candeloro1, Lídia Aparecida Rossi3, Elaine Caldeira de Oliveira Guirro4.
Abstract
BACKGROUND: Skin function is dependent on its biomechanical characteristics, resistance, malleability, and elasticity. Therapeutic ultrasound may increase cutaneous malleability thus and optimize the rehabilitation process on specific diseases. The aim of this study is to evaluate possible alterations of biomechanical characteristics of the normal skin after therapeutic ultrasound application.Entities:
Keywords: Elasticity; Skin; Therapeutic ultrasound
Year: 2016 PMID: 27536356 PMCID: PMC4988011 DOI: 10.1186/s40349-016-0065-8
Source DB: PubMed Journal: J Ther Ultrasound ISSN: 2050-5736
Fig. 1Cutometer probe application with specific adhesive rings fixed in the anterior forearm and delimited by a neoprene template
Fig. 2Application of therapeutic ultrasound in delimited area with neoprene template
Values of the R0 variable before and after therapeutic ultrasound application at normal skin
| Variable R0 | Number | Average (mm) | Standard deviation (mm) | Standard error (mm) |
|---|---|---|---|---|
| Pre-intervention | 31 | 0.3273 | 0.1038 | 0.0186 |
| Post-intervention | 31 | 0.3795* | 0.1042 | 0.0187 |
| Difference | −0.05226 | 0.05396 | 0.00969 |
*Differs from pre- (p = 0.001)
Values of the R2 variable before and after therapeutic ultrasound application at normal skin
| Variable R2 | Number | Average | Standard deviation | Standard error |
|---|---|---|---|---|
| Pre-intervention | 31 | 0.8419 | 0.0946 | 0.0170 |
| Post-intervention | 31 | 0.8884* | 0.0624 | 0.0112 |
| Difference | −0.0464 | 0.0675 | 0.0121 |
*Differs from pre- (p = 0.001)
Values of the R6 variable before and after therapeutic ultrasound application at normal skin
| Variable R6 | Number | Average | Standard deviation | Standard error |
|---|---|---|---|---|
| Pre-intervention | 31 | 0.4605 | 0.1079 | 0.0194 |
| Post-intervention | 31 | 0.3629* | 0.0934 | 0.0168 |
| Difference | 0.0976 | 0.0827 | 0.0149 |
*Differs from pre- (p = 0.001)
Fig. 3Values of the R0 (mm) variable before and after therapeutic ultrasound application on normal skin
Fig. 4Values of the R2 variable before and after therapeutic ultrasound application on normal skin
Fig. 5Values of the R6 variable before and after therapeutic ultrasound application on normal skin