Superiority of biphasic shocks in the defibrillation of dogs by epicardial patches and catheter electrodes.

G C Flaker1, J C Schuder, W C McDaniel, H Stoeckle, M Dbeis.   

Abstract

Currently available internal cardiac defibrillators use a uniphasic, truncated exponential waveform morphology of about 6 msec in duration at an energy level of 23 to 33 joules. To determine if improved defibrillation could be achieved with a different waveform morphology, we implanted 4.5 cm2 titanium patches to the left and right ventricle of 28 dogs. After ventricular fibrillation was induced, defibrillation was attempted using 7, 12, 13, or 17 joules. A 5 msec rectangular uniphasic waveform morphology was compared with a 10 msec rectangular biphasic waveform with the lagging 5 msec pulse of half the amplitude of the leading 5 msec. In an additional seven dogs, a transvenous bipolar catheter was placed with the distal electrode in the right ventricular apex and the proximal electrode in the superior vena cava. Biphasic and uniphasic shocks were compared at 14 joules. In the patch-patch system, the biphasic waveform was superior to the uniphasic waveform at 7 joules (67% versus 35%, p less than 0.001) and at 12 joules (93% versus 78%, p less than 0.001). No statistically significant differences were achieved at 13 joules or 17 joules. In the catheter electrode system with a delivered energy of 14 joules, the biphasic waveform was more effective than the uniphasic waveform (87% versus 27%, p less than 0.001). Manufacturers of automatic implantable defibrillators should consider this information in the design of future automatic implantable defibrillators.

Entities:  

Mesh:

Year:  1989        PMID: 2750650     DOI: 10.1016/0002-8703(89)90187-7

Source DB:  PubMed          Journal:  Am Heart J        ISSN: 0002-8703            Impact factor:   4.749


  6 in total

1.  Analysis of electric field stimulation of single cardiac muscle cells.

Authors:  L Tung; J R Borderies
Journal:  Biophys J       Date:  1992-08       Impact factor: 4.033

2.  Biphasic electrical field stimulation aids in tissue engineering of multicell-type cardiac organoids.

Authors:  Loraine L Y Chiu; Rohin K Iyer; John-Paul King; Milica Radisic
Journal:  Tissue Eng Part A       Date:  2008-09-10       Impact factor: 3.845

3.  Mechanisms of cardiac cell excitation with premature monophasic and biphasic field stimuli: a model study.

Authors:  M G Fishler; E A Sobie; N V Thakor; L Tung
Journal:  Biophys J       Date:  1996-03       Impact factor: 4.033

4.  The development and application of the implantable cardioverter defibrillator.

Authors:  Walter Coats; Lenard Politte; Greg C Flaker
Journal:  Mo Med       Date:  2010 Jan-Feb

5.  [Influence of waveform and configuration of electrodes on the defibrillation threshold of implantable cardioverter-defibrillators].

Authors:  M Block; D Hammel; G Breithardt
Journal:  Herzschrittmacherther Elektrophysiol       Date:  1997-03

6.  Simultaneous comparison of many triphasic defibrillation waveforms.

Authors:  Ron Davis; Robert Malkin
Journal:  Open Biomed Eng J       Date:  2012-02-14
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.