Michelle O'Brien1, Ludwig Bueermann2.
Abstract
A comparison has been made of the air-kerma standards for low-energy x rays at the National Institute of Standards and Technology (NIST) and the Physikalisch-Technische Bundesanstalt (PTB). The comparison involved a series of measurements at the PTB and the NIST using the air-kerma standards and two NIST reference-class transfer ionization chamber standards. Results are presented for the reference radiation beam qualities in the range from 25 kV to 50 kV for low energy x rays, including the techniques used for mammography dose traceability. The tungsten generated reference radiation qualities, between 25 kV and 50 kV used for this comparison, are new to NIST; therefore this comparison will serve as the preliminary comparison for NIST and a verification of the primary standard correction factors. The mammography comparison will repeat two previously unpublished comparisons between PTB and NIST. The results show the standards to be in reasonable agreement within the standard uncertainty of the comparison of about 0.4 %.Entities:
Keywords: air kerma; free-air ionization chamber; mammography qualities; primary standard; reference radiation qualities; x-ray calibration; x-rays
Year: 2009 PMID: 27504231 PMCID: PMC4646528 DOI: 10.6028/jres.114.023
Source DB: PubMed Journal: J Res Natl Inst Stand Technol ISSN: 1044-677X
Physical constants used in the determination of the air-kerma rate
| Physical constant | Value | Relative standard uncertainty |
|---|---|---|
| ρair | 1.293 kg m−3 | 0.01 |
| | 33.97 J C−1 | 0.15 |
| 1 – | 1.0000 | 0.01 |
Density of dry air at 273.15 K and 101 325 Pa.
Main characteristics of the primary standards used in the comparison
| Characteristic | NIST Ritz | NIST Attix | PTB PK100 |
|---|---|---|---|
| Air-path length / cm | 12.74 | 21.27 | 9.72 |
| Plate separation / cm | 9.0 | Variable | 23.4 |
| Collecting plate length / cm | 7.003 | Variable | 2.0021 |
| Aperture diameter / cm | 1.0 | 1.0 or 0.5 | 2.0008 |
| Measuring volume / cm3 | 5.502 | Variable | 6.2947 |
| Polarizing voltage / V | −5000 | −2500 | 6000 |
Main characteristics of ionization chambers used as transfer standards
| Model | Serial number | Type | Sensitive volume | Collector diameter |
|---|---|---|---|---|
| A11 | 114 | PLANAR | 0.62 cm3 | 20 mm |
| A15 | 103 | PLANAR | 2.46 cm3 | 20 mm |
Characteristics of the reference radiation qualities used for the comparison at a measurement position of 50 cm
| Reference radiation | Gernerating potential | Additional filtration | Half-value layer (HVL) | Air-kerma rate | |
|---|---|---|---|---|---|
| kV | mm A1 | mm Cu | mm A1 | mGy s−1 | |
| PTB | |||||
|
| |||||
| BIPM10 | 10 | 0 | 0.033 | 0.28 | |
| BIPM25 | 25 | 0.374 | 0.239 | 0.28 | |
| BIPM30 | 30 | 0.208 | 0.163 | 0.42 | |
| BIPM50a | 50 | 4.00 | 2.291 | 0.19 | |
| BIPM50b | 50 | 1.00 | 1.065 | 0.28 | |
|
| |||||
| NIST | |||||
|
| |||||
| BIPM10 | 10 | 0 | 0.037 | 2.60 | |
| BIPM25 | 25 | 0.373 | 0.240 | 1.40 | |
| BIPM30 | 30 | 0.203 | 0.167 | 4.20 | |
| BIPM40 | 40 | 3.97 | 0.208 | 2.649 | 0.043 |
| BIPM50a | 50 | 3.97 | 2.291 | 0.730 | |
| BIPM50b | 50 | 1.07 | 1.038 | 3.20 | |
Characteristics of the mammography reference radiation qualities used for the comparison at a measurement position of 100 cm
| Reference radiation | Tube voltage | Additional filtration | Half-value layer (HVL) | Air-kerma rate |
|---|---|---|---|---|
| (kV) | (mm) | (mm A1) | mGy s−1 | |
| PTB | ||||
|
| ||||
| MMV 20 | 20 | 0.03 Mo | 0.22 | 0.555 |
| MMV 25 | 25 | 0.03 Mo | 0.29 | 0.108 |
| MMV 28 | 28 | 0.03 Mo | 0.32 | 1.070 |
| MMV 30 | 30 | 0.03 Mo | 0.33 | 1.030 |
| MMV 35 | 35 | 0.03 Mo | 0.37 | 1.070 |
| MMH 20 | 20 | 0.030 Mo + 2.0 Al | 0.46 | 0.013 |
| MMH 25 | 25 | 0.030 Mo + 2.0 Al | 0.56 | 0.062 |
| MMH 28 | 28 | 0.030 Mo + 2.0 Al | 0.61 | 0.069 |
| MMH 30 | 30 | 0.030 Mo + 2.0 Al | 0.64 | 0.071 |
| MMH 35 | 35 | 0.030 Mo + 2.0 Al | 0.73 | 0.070 |
|
| ||||
| NIST | ||||
|
| ||||
| Mo/Mo23 | 23 | 0.032 Mo | 0.271 | 0.330 |
| Mo/Mo25 | 25 | 0.032 Mo | 0.296 | 0.460 |
| Mo/Mo28 | 28 | 0.032 Mo | 0.332 | 0.680 |
| Mo/Mo30 | 30 | 0.032 Mo | 0.351 | 0.400 |
| Mo/Mo35 | 35 | 0.032 Mo | 0.392 | 0.840 |
| Mo/Mo25x | 25 | 0.030 Mo + 2.0 Al | 0.566 | 0.018 |
| Mo/Mo28x | 28 | 0.030 Mo + 2.0 Al | 0.626 | 0.038 |
| Mo/Mo30x | 30 | 0.030 Mo + 2.0 Al | 0.660 | 0.055 |
| Mo/Mo35x | 35 | 0.030 Mo + 2.0 Al | 0.748 | 0.120 |
Correction factors used in the comparison for the NIST Ritz standard
| Correction factor | Generating potential (kV) | Relative standard uncertainty | |||||
|---|---|---|---|---|---|---|---|
| 25 | 30 | 40 | 50a | 50b | Type A | Type B | |
| Air attenuation | 1.039 | 1.054 | 1.005 | 1.006 | 1.012 | 0.01 | 0.01 |
| Scattered radiation | 0.9966 | 0.9967 | 0.9976 | 0.9974 | 0.9952 | 0.07 | |
| Electron loss | 1.000 | 1.000 | 1.000 | 1.005 | 1.001 | 0.05 | |
| Ion recombination | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | 0.03 | |
| Fluorescence | 0.9961 | 0.9962 | 0.99828 | 0.9981 | 0.9976 | 0.03 | |
| Aperture transmission | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | 0.04 | |
| Field distortion | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | 0.1 | |
| Wall transmission | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | 0.01 | |
| Humidity | 0.998 | 0.998 | 0.998 | 0.998 | 0.998 | 0.03 | |
These are nominal values for T = 293.15 K and p = 101 325 Pa. Each measurement is corrected using the air temperature and pressure at time of measurement.
Correction factors used in the comparison for the PTB PK100 standard: Mammography radiation qualities, attenuated (PTB MMH-series)
| Correction factor | Generating potential (kV) | Relative standard uncertainty (%) | |||||
|---|---|---|---|---|---|---|---|
| 20 | 25 | 28 | 30 | 35 | Type A | Type B | |
| Air attenuation | 1.0138 | 1.0116 | 1.0110 | 1.0106 | 1.0098 | 0.05 | 0.05 |
| Scattered radiation | 0.9916 | 0.9919 | 0.9920 | 0.9921 | 0.9923 | 0.05 | |
| Electron loss | 1.0000 | 1.0000 | 1.0000 | 1.0000 | 1.0000 | 0.05 | |
| Volume recombination | 1.0001 | 1.0004 | 1.0004 | 1.0005 | 1.0005 | 0.05 | 0.05 |
| Initial recombination | 1.0005 | 1.0005 | 1.0005 | 1.0005 | 1.0005 | 0.05 | |
| Guard srip attenuation | 1.0029 | 1.0024 | 1.0023 | 1.0023 | 1.0021 | 0.05 | 0.05 |
| Aperture edge transmission | 0.9996 | 0.9995 | 0.9995 | 0.9994 | 0.9994 | 0.05 | |
| Field distortion | 0.9920 | 0.9920 | 0.9920 | 0.9920 | 0.9920 | 0.15 | |
| Wall transmission | 1.0000 | 1.0000 | 1.0000 | 1.0000 | 1.0000 | 0.05 | |
| Humidity | 0.9980 | 0.9980 | 0.9980 | 0.9980 | 0.9980 | 0.05 | |
These are nominal values for T = 293.15 K and p = 101 325 Pa. Each measurement is corrected using the air temperature and pressure measured at the time.
This correction includes the re-absorption of scattered radiation and of fluorescent photons.
Relative standard uncertainties (in %) associated with the standards
| Source of uncertainty | NIST Attix | NIST Ritz | PTB | |||
|---|---|---|---|---|---|---|
| Type A | Type B | Type A | Type B | Type A | Type B | |
| Ionization current | 0.13 | 0.06 | 0.05 | 0.06 | 0.10 | 0.06 |
| Volume | 0.01 | 0.07 | 0.04 | 0.01 | 0.06 | |
| Positioning | 0.01 | 0.01 | 0.01 | |||
| Correction factors (excl. | 0.04 | 0.09 | 0.03 | 0.14 | 0.10 | 0.20 |
| Physical constants | 0.15 | 0.15 | 0.15 | |||
|
| ||||||
|
| 0.14 | 0.20 | 0.07 | 0.22 | 0.14 | 0.26 |
|
| ||||||
| 0.24 | 0.23 | 0.30 | ||||
Relative standard uncertainties (in %) associated with the calibration of the transfer ionization chambers
| Source of uncertainty | NIST Attix | NIST Ritz | PTB | |||
|---|---|---|---|---|---|---|
| Type A | Type B | Type A | Type B | Type A | Type B | |
| Air-kerma rate
| 0.14 | 0.20 | 0.07 | 0.22 | 0.14 | 0.26 |
| Ionization current | 0.09 | 0.06 | 0.05 | 0.06 | 0.10 | 0.06 |
| Positioning | 0.01 | 0.01 | 0.04 | |||
| Monitor normalization | 0.05 | |||||
| Air density correction | 0.04 | |||||
|
| ||||||
| 0.17 | 0.21 | 0.09 | 0.23 | 0.17 | 0.28 | |
|
| ||||||
| 0.27 | 0.24 | 0.33 | ||||
Relative standard uncertainties (in %) associated with the comparison results R
| Source of uncertainty | Type A | Type B |
|---|---|---|
| Comparison result | 0.24 | 0.28 |
| 0.37 | ||
Measured PTB results for the calibration of the NIST transfer chambers at both laboratories
| Reference radiation | NIST transfer chambers calibration coefficients | |||
|---|---|---|---|---|
| ( | A11 SN114 | A11 SN114 | A11 SN114 | A11 SN114 |
| BIPM10 | 3.326 | 3.315 | 1.341 | 1.347 |
| BIPM25 | 3.268 | 3.245 | 1.259 | 1.265 |
| BIPM30 | 3.262 | 3.243 | 1.269 | 1.277 |
| BIPM50a | 3.203 | 3.162 | 1.207 | 1.213 |
| BIPM50b | 3.247 | 3.210 | 1.221 | 1.226 |
| MMV 20 | 3.259 | 3.239 | 1.254 | 1.263 |
| MMV 25 | 3.267 | 3.245 | 1.245 | 1.254 |
| MMV 28 | 3.271 | 3.250 | 1.243 | 1.251 |
| MMV 30 | 3.273 | 3.252 | 1.242 | 1.251 |
| MMV 35 | 3.276 | 3.254 | 1.239 | 1.248 |
| MMH 20 | 3.293 | 3.250 | 1.235 | 1.240 |
| MMH 25 | 3.281 | 3.250 | 1.225 | 1.233 |
| MMH 28 | 3.278 | 3.249 | 1.222 | 1.230 |
| MMH 30 | 3.274 | 3.245 | 1.221 | 1.228 |
| MMH 35 | 3.268 | 3.240 | 1.218 | 1.226 |
Measured PTB results for the calibration of the NIST transfer chambers at both laboratories
| Reference radiation | NIST transfer chambers calibration coefficients | |||
|---|---|---|---|---|
| ( | A11 SN114 | A11 SN114 | A11 SN114 | A11 SN114 |
| BIPM25 | 3.260 | 3.245 | 1.258 | 1.262 |
| BIPM30 | 3.247 | 3.231 | 1.260 | 1.269 |
| BIPM40 | 3.202 | 3.186 | 1.209 | 1.218 |
| BIPM50a | 3.207 | 3.183 | 1.209 | 1.216 |
| BIPM50b | 3.262 | 3.237 | 1.225 | 1.232 |
| Mo/Mo23 | 3.264 | 3.252 | 1.247 | 1.262 |
| Mo/Mo25 | 3.267 | 3.246 | 1.244 | 1.254 |
| Mo/Mo28 | 3.260 | 3.248 | 1.250 | 1.252 |
| Mo/Mo30 | 3.271 | 3.246 | 1.240 | 1.245 |
| Mo/Mo35 | 3.270 | 3.242 | 1.243 | 1.242 |
| Mo/Mo25x | 3.291 | 3.269 | 1.234 | 1.240 |
| Mo/Mo28x | 3.293 | 3.268 | 1.227 | 1.236 |
| Mo/Mo30x | 3.291 | 3.261 | 1.227 | 1.229 |
| Mo/Mo35x | 3.262 | 3.249 | 1.219 | 1.230 |
Results of the comparison for each of the transfer chambers shown as the ratio of the average calibration coefficient NK, NIST / NK, PTB for both chambers, as well as the mean
| Reference radiation | A11 SN114 | A15 SN103 | Mean |
|---|---|---|---|
| BIPM25 | 0.999 | 0.998 | 0.998 |
| BIPM30 | 0.996 | 0.993 | 0.995 |
| BIPM50a | 1.004 | 1.002 | 1.003 |
| BIPM50b | 1.007 | 1.004 | 1.005 |
| Mo/Mo25 | 1.000 | 0.999 | 1.000 |
| Mo/Mo28 | 0.998 | 1.003 | 1.001 |
| Mo/Mo30 | 0.999 | 0.997 | 0.998 |
| Mo/Mo35 | 0.997 | 0.999 | 0.998 |
| Mo/Mo25x | 1.004 | 1.006 | 1.005 |
| Mo/Mo28x | 1.005 | 1.005 | 1.005 |
| Mo/Mo30x | 1.005 | 1.003 | 1.004 |
| Mo/Mo35x | 1.000 | 1.002 | 1.001 |
Correction factors used in the comparison for the NIST Attix mammography standard
| Correction factor | Reference radiation qualities | Relative standard uncertainty (%) | |||||
|---|---|---|---|---|---|---|---|
| Mo / Mo23 | Mo / Mo25 | Mo / Mo28 | Mo / Mo30 | Mo / Mo35 | Type A | Type B | |
| Air attenuation | 1.055 | 1.051 | 1.047 | 1.043 | 1.040 | 0.01 | |
| Scattered radiation | 0.9949 | 0.9950 | 0.9950 | 0.9951 | 0.9952 | 0.07 | |
| Electron loss | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | 0.05 | |
| Ion recombination | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | 0.04 | |
| Humidity | 0.998 | 0.998 | 0.998 | 0.998 | 0.998 | 0.03 | |
These are nominal values for T = 293.15 K and p = 101 325 Pa. Each measurement is corrected using the air temperature and pressure measured at the time.
Correction factors used in the comparison for the NIST Attix mammography standard
| Correction factor | Reference radiation qualities | Relative standard uncertainty (%) | ||||
|---|---|---|---|---|---|---|
| Mo / Mo25x | Mo / Mo28x | Mo / Mo30x | Mo / Mo35x | Type A | Type B | |
| Air attenuation | 1.021 | 1.024 | 1.021 | 1.020 | 0.01 | |
| Scattered radiation | 0.9955 | 0.9956 | 0.9956 | 0.9957 | 0.07 | |
| Electron loss | 1.000 | 1.000 | 1.000 | 1.000 | 0.05 | |
| Ion recombination | 1.000 | 1.000 | 1.000 | 1.000 | 0.04 | |
| Humidity | 0.998 | 0.998 | 0.998 | 0.998 | 0.03 | |
These are nominal values for T = 293.15 K and p = 101 325 Pa. Each measurement is corrected using the air temperature and pressure measured at the time.
Correction factors used in the comparison for the PTB PK100 standard: BIPM reference radiation qualities (PTB F-series)
| Correction factor | Generating potential (kV) | Relative standard uncertainty (%) | |||||
|---|---|---|---|---|---|---|---|
| 10 | 25 | 30 | 50a | 50b | Type A | Type B | |
| Air attenuation | 1.2274 | 1.0340 | 1.0485 | 1.0046 | 1.0095 | 0.05 | 0.05 |
| Scattered radiation | 0.9854 | 0.9901 | 0.9893 | 0.9939 | 0.9927 | 0.05 | |
| Electron loss | 1.0000 | 1.0000 | 1.0000 | 1.0000 | 1.0000 | 0.05 | |
| Volume recombination | 1.0014 | 1.0017 | 1.0026 | 1.0012 | 1.0018 | 0.05 | 0.05 |
| Initial recombination | 1.0005 | 1.0005 | 1.0005 | 1.0005 | 1.0005 | 0.05 | |
| Guard srip attenuation | 1.0393 | 1.0075 | 1.0110 | 1.0015 | 1.0027 | 0.05 | 0.05 |
| Aperture edge transmission | 0.9998 | 0.9991 | 0.9992 | 0.9971 | 0.9979 | 0.05 | |
| Field distortion | 0.9920 | 0.9920 | 0.9920 | 0.9920 | 0.9920 | 0.15 | |
| Wall transmission | 1.0000 | 1.0000 | 1.0000 | 1.0000 | 1.0000 | 0.05 | |
| Humidity | 0.9980 | 0.9980 | 0.9980 | 0.9980 | 0.9980 | 0.05 | |
These are nominal values for T = 293.15 K and p = 101 325 Pa. Each measurement is corrected using the air temperature and pressure measured at the time.
This correction includes the re-absorption of scattered radiation and of fluorescent photons.
Correction factors used in the comparison for the PTB PK100 standard: Mammography radiation qualities, non-attenuated (PTB MMV-series)
| Correction factor | Generating potential (kV) | Relative standard uncertainty (%) | |||||
|---|---|---|---|---|---|---|---|
| 20 | 25 | 28 | 30 | 35 | Type A | Type B | |
| Air attenuation | 1.0333 | 1.0275 | 1.0256 | 1.0250 | 1.0232 | 0.05 | 0.05 |
| Scattered radiation | 0.9899 | 0.9904 | 0.9906 | 0.9907 | 0.9909 | 0.05 | |
| Electron loss | 1.0000 | 1.0000 | 1.0000 | 1.0000 | 1.0000 | 0.05 | |
| Volume recombination | 1.0001 | 1.0004 | 1.0004 | 1.0005 | 1.0005 | 0.05 | 0.05 |
| Initial recombination | 1.0005 | 1.0005 | 1.0005 | 1.0005 | 1.0005 | 0.05 | |
| Guard srip attenuation | 1.0068 | 1.0056 | 1.0052 | 1.0051 | 1.0047 | 0.05 | 0.05 |
| Aperture edge transmission | 0.9997 | 0.9997 | 0.9996 | 0.9996 | 0.9996 | 0.05 | |
| Field distortion | 0.9920 | 0.9920 | 0.9920 | 0.9920 | 0.9920 | 0.15 | |
| Wall transmission | 1.0000 | 1.0000 | 1.0000 | 1.0000 | 1.0000 | 0.05 | |
| Humidity | 0.9980 | 0.9980 | 0.9980 | 0.9980 | 0.9980 | 0.05 | |
These are nominal values for T = 293.15 K and p = 101 325 Pa. Each measurement is corrected using the air temperature and pressure measured at the time.
This correction includes the re-absorption of scattered radiation and of fluorescent photons.