Literature DB >> 27504129

Robustness of IPSA optimized high-dose-rate prostate brachytherapy treatment plans to catheter displacements.

Joel Poder1, May Whitaker2.   

Abstract

PURPOSE: Inverse planning simulated annealing (IPSA) optimized brachytherapy treatment plans are characterized with large isolated dwell times at the first or last dwell position of each catheter. The potential of catheter shifts relative to the target and organs at risk in these plans may lead to a more significant change in delivered dose to the volumes of interest relative to plans with more uniform dwell times.
MATERIAL AND METHODS: This study aims to determine if the Nucletron Oncentra dwell time deviation constraint (DTDC) parameter can be optimized to improve the robustness of high-dose-rate (HDR) prostate brachytherapy plans to catheter displacements. A set of 10 clinically acceptable prostate plans were re-optimized with a DTDC parameter of 0 and 0.4. For each plan, catheter displacements of 3, 7, and 14 mm were retrospectively applied and the change in dose volume histogram (DVH) indices and conformity indices analyzed.
RESULTS: The robustness of clinically acceptable prostate plans to catheter displacements in the caudal direction was found to be dependent on the DTDC parameter. A DTDC value of 0 improves the robustness of planning target volume (PTV) coverage to catheter displacements, whereas a DTDC value of 0.4 improves the robustness of the plans to changes in hotspots.
CONCLUSIONS: The results indicate that if used in conjunction with a pre-treatment catheter displacement correction protocol and a tolerance of 3 mm, a DTDC value of 0.4 may produce clinically superior plans. However, the effect of the DTDC parameter in plan robustness was not observed to be as strong as initially suspected.

Entities:  

Keywords:  DTDC; IPSA; brachytherapy; catheter displacement; prostate cancer

Year:  2016        PMID: 27504129      PMCID: PMC4965496          DOI: 10.5114/jcb.2016.60499

Source DB:  PubMed          Journal:  J Contemp Brachytherapy        ISSN: 2081-2841


Purpose

Prostate cancer is the most common male malignancy in the Western world, and as life expectancy increase, the prevalence is also expected to increase in an aging population [1]. Radiotherapy is an important therapeutic modality for the treatment of patients with localized or locally advanced prostate cancer [2] utilizing both external beam radiotherapy (EBRT) and brachytherapy. Over the last few decades, significant advances in technology related to high-dose-rate (HDR) brachytherapy have seen an increase in its use as a localized boost to EBRT of the prostate. An important technological advancement in HDR brachytherapy is the evolution from forward planning to inverse planning techniques [3, 4, 5]. The inverse planning optimization algorithm currently implemented in the Nucletron Oncentra (Nucletron B.V., Veenendaal, The Netherlands) brachytherapy treatment planning system (TPS) is the inverse planning simulated annealing (IPSA) optimization algorithm. Inverse planning simulated annealing is based on contoured anatomy and optimizes dwell times using a simulated annealing algorithm [6]. The algorithm is constrained by user specific surface and volumetric dose constraints for both the target volume and organs at risk to calculate clinically acceptable treatment plans [7]. Inverse planning simulated annealing optimized brachytherapy treatment plans are characterized with large isolated dwell times at the first or last dwell position of each catheter. The central dwell positions however consist of extremely short, or zero, dwell times [8]. There is concern amongst users that these large isolated dwell times may lead to hot spots, either inside or outside the target. Also, the potential of catheter shifts relative to the target and organs at risk may lead to a more significant change in delivered dose to the volumes of interest relative to plans with more uniform dwell times. Recently, the Nucletron Oncentra TPS has added the dwell time deviation constraint (DTDC) parameter to the IPSA optimization process. This parameter constrains the allowable dwell times in the optimization process and can be set to a value between 0 and 1 in increments of 0.1. A value of 0 corresponds to completely unrestricted dwell times and a value of 1 results in homogeneous dwell times [9]. The displacement of catheters relative to the target and organs at risk during the time between imaging and patient treatment has been reported by a number of groups [10, 11, 12, 13]. The displacements have predominantly been reported along the patient longitudinal axis and in the caudal direction [10] primarily due to acute edema between the prostate and perineal skin [14]. Previous work from our group [11] demonstrated a median catheter displacement of 7.5 mm in caudal direction (range 2.9-23.9 mm) in the time from planning CT to treatment (approximately 1-3 hours). Tiong et al. [10] have reported significant adverse effects on the tumor control probability for catheter displacements larger than 3 mm, including underdosage of the target and overdosage to critical structures. Due to these findings, our department has implemented a clinical protocol, in which internal catheter positions are verified and corrected immediately prior to treatment delivery with a tolerance of 3 mm. This study aims to determine if the DTDC parameter can be optimized to improve the robustness of HDR prostate brachytherapy plans to catheter displacements relative to patient anatomy. A set of 10 clinically acceptable prostate plans were re-optimized with a DTDC parameter of 0 and 0.4. The values of 0 and 0.4 were chosen to reflect the change that is currently occurring in our clinical protocol. For each plan, catheter displacements of 3, 7, and 14 mm were retrospectively applied, and the change in DVH indices and conformity indices analyzed.

Material and methods

Initial plans

A set of 10 clinical prostate HDR brachytherapy plans were chosen for analysis. These CT plans were created between 2012 and 2015 on the Nucletron Oncentra Brachytherapy TPS (v4.3, Nucletron B.V., Veenendaal, The Netherlands). The prostate planning target volume (PTV), urethra, and rectum were all contoured by the same radiation oncologist at the time of treatment. Prostate volumes varied between 25.1 and 59.4 cm3 and the number of catheters used was between 14 and 24. All patients received 2 fractions of 9.5 Gy with 2 weeks between fractions. Catheter insertion (using Oncosmart, ProGuide Sharp Needle, 6F, Nucletron B.V., Veenendaal, The Netherlands), CT scan, planning, and treatment are all performed on the same day, and the mean time between the planning CT and treatment was 182 minutes.

IPSA optimization and DTDC

Each plan was optimized using the IPSA algorithm using the parameters outlined in Table 1. As per clinical protocol, plans were initially optimized with the DTDC parameter set to 0. The plans were then re-optimized with the DTDC parameter set to 0.4 and all other parameters kept constant. The dwell time characteristics of each plan were then compared using the plan modulation index (M), as defined by Smith et al. [9]. The plan modulation index is defined as the maximum deviation of dwell time from the average dwell time for each catheter, normalized to the maximum dwell time for the treatment plan, averaged over all catheters in the plan.
Table 1

Inverse planning simulated annealing (IPSA) optimization parameters used for patient plan optimization

ROIUsageSurfaceVolume
WeightMin (cGy)Max (cGy)WeightWeightMin (cGy)Max (cGy)Weight
ProstatePTV100950142510010010095030
RectumOrgan665505047550
UrethraOrgan120950998505095099850

PTV – planning target volume

Inverse planning simulated annealing (IPSA) optimization parameters used for patient plan optimization PTV – planning target volume

Catheter displacements

Catheter displacements in the caudal direction were then simulated for each plan. Offsets of 3, 7, and 14 mm were performed. Displacements of this magnitude were chosen as they corresponded to clinically relevant catheter displacements, as found in a previous study by our group [11]. Our center has implemented a clinical protocol, in which catheter displacements ≥ 3 mm are corrected for, immediately prior to treatment by altering the indexer length at the treatment console. Implanting the catheters past the prostate base into the bladder allowed for extra dwell positions beyond the prostate in the event of a caudal shift. Physical re-insertion was not performed.

Plan analysis

All patient plans were assessed by evaluating dose volume histogram (DVH) indices and dose quality indices. Dose volume histogram indices used for plan evaluation are outlined in Table 2. Furthermore, a normal tissue (NT) contour was created by adding a 2 mm margin around the PTV and subtracting this expanded contour from the external contour, e.g. NT = Body – (PTV + 2 mm). These parameters were automatically calculated by the TPS and are highly dependent on the size of the histogram bin used for calculation [8]. Because of this, a conformity index (CI), a dose inhomogeneity index (DHI), and an overdose volume index (ODI) were also calculated for each plan.
Table 2

Clinically acceptable dose volume histogram (DVH) indices

Volume typeDose (%)Dose (cGy)Volume (%)
PTV100950≥ 90
PTV1501425< 30
PTV2001900< 15
Rectum706650
Urethra12011400

PTV – planning target volume

Clinically acceptable dose volume histogram (DVH) indices PTV – planning target volume The CI used in this study is the one introduced by van't Riet et al. [15] and is shown in Equation 1: where V is the volume of the PTV receiving a dose greater than or equal to the 100% isodose, V is the volume of the PTV, and V is the volume of the 100% isodose. The DHI parameter gives an indication of the homogeneity of the dose within the PTV, which was first introduced by Wu et al. [16] and is shown in Equation 2: where V is the volume of the PTV receiving a dose greater than or equal to the 150% isodose, and V is as described above. Finally, the ODI parameter [17] indicates the amount of high dose (greater than 200%) within the PTV: where V is the volume of the PTV receiving a dose greater than or equal to the 200% isodose. The change in DVH and dose quality indices was then calculated as a function of catheter displacement for both DTDC values of 0 and 0.4. The change in these indices with increasing catheter displacement gives an indication of the robustness of the plans to changes in catheter position relative to the targets and organs at risk between planning CT and treatment. Statistical significance between the DTDC values was verified using a paired t-test with α = 0.05 (corresponding to a 5% significance level).

Results

Initial clinical plans

Initial clinical prostate plans IPSA optimized with a DTDC value of 0 produced a large spread of dwell times, relative to those plans optimized with a DTDC value of 0.4. The plan modulation index (M) for each plan variant is given in Table 3 along with the total dwell time, normalized to the air kerma strength of the source. The average M for the 0 DTDC case (± 1 SD) was equal to 0.44 ± 0.07, whereas for the 0.4 DTDC case M = 0.20 ± 0.06. The effect of increasing the DTDC parameter is to limit the maximum dwell time in any catheter; this is reflected by the decreasing value of M, as more homogeneous dwell time distribution is created within each catheter. The total dwell time, normalized to the air kerma strength of the source was also seen to decrease for plans optimized with a DTDC value of 0.4, relative to a DTDC of 0. This is due to the reduction in large isolated dwell times at the first or last dwell positions of each catheter.
Table 3

Plan modulation index (M) and normalized total dwell time (cGy-1cm-2)

PatientPlan modulation index (M)Normalized total dwell time (cGy-1cm-2)
DTDC 0DTDC 0.4DTDC 0DTDC 0.4
10.480.098.32 × 10-6 7.73 × 10-6
20.340.232.75 × 10-6 2.55 × 10-6
30.430.252.57 × 10-6 2.33 × 10-6
40.410.2310.44 × 10-6 9.82 × 10-6
50.570.1715.27 × 10-6 14.32 × 10-6
60.430.1812.45 × 10-6 12.34 × 10-6
70.420.246.32 × 10-6 6.12 × 10-6
80.480.198.72 × 10-6 8.29 × 10-6
90.400.218.37 × 10-6 7.80 × 10-6
100.440.207.27 × 10-6 6.80 × 10-6

DTDC – dwell time deviation constraint

Plan modulation index (M) and normalized total dwell time (cGy-1cm-2) DTDC – dwell time deviation constraint Table 4 shows the change in DVH and dose quality indices when re-optimizing the plans with a DTDC parameter of 0.4. By changing the DTDC value to 0.4, the coverage of the PTV is improved, as reflected by the increase in PTV V100% and CI, however, only the difference in CI was found to be statistically significant. Statistically significant reductions in NT V100%, NT V150%, and NT V200% were also found when changing DTDC from 0 to 0.4. This, along with the improvement in CI, is due to the reduction in the large isolated dwell times just outside of the PTV, which are delivering higher doses to the adjacent healthy tissue. On the other hand, there is a statistically significant increase in PTV V150%, PTV V200%, DHI, and ODI for DTDC 0.4 versus DTDC 0. The rectum V150% was largely unaffected by the DTDC change, and a small reduction in the urethra V120% was observed.
Table 4

Initial dose volume histogram (DVH) and dose indices before catheter displacement

ParameterDTDC 0DTDC 0.4 p
PTV V100% 92.9 ± 1.9%93.4 ± 1.7%≤ 0.2620
PTV V150% 18.9 ± 3.6%23.0 ± 3.7%≤ 0.0001
PTV V200% 7.5 ± 1.6%9.0 ± 1.4%≤ 0.0002
Rectum V70% 0.1 ± 0.1%0.1 ± 0.1%≤ 0.3498
Urethra V120% 10.4 ± 5.6%5.9 ± 5.9%≤ 0.0421
NT V100% 11.0 ± 3.0 cc5.7 ± 1.9 cc≤ 0.0001
NT V150% 3.3 ± 1.4 cc1.2 ± 0.7 cc≤ 0.0001
NT V200% 1.5 ± 0.2 cc0.5 ± 0.3 cc≤ 0.0018
CI0.691 ± 0.0460.748 ± 0.042≤ 0.0002
DHI0.204 ± 0.0390.246 ± 0.038≤ 0.0001
ODI0.081 ± 0.0180.097 ± 0.014≤ 0.0004

PTV – planning target volume, NT – normal tissue, CI – conformity index, DHI – dose inhomogeneity index, ODI – overdose index, V100%, V150%, V200%, V70%, V120% – volume of relevant structure receiving 100%, 150%, 200%, 70%, and 120% of the prescribed isodose, respectively, DTDC – dwell time deviation constraint

Initial dose volume histogram (DVH) and dose indices before catheter displacement PTV – planning target volume, NT – normal tissue, CI – conformity index, DHI – dose inhomogeneity index, ODI – overdose index, V100%, V150%, V200%, V70%, V120% – volume of relevant structure receiving 100%, 150%, 200%, 70%, and 120% of the prescribed isodose, respectively, DTDC – dwell time deviation constraint

Effect of catheter displacements

The effect of catheter displacements on DVH and dose quality indices are shown in Tables 5, 6, and 7 for catheter shifts of 3, 7, and 14 mm, respectively. Overall, a DTDC value of 0 improves the robustness of PTV coverage to catheter displacements relative to a DTDC value of 0.4. This is reflected in the smaller changes in PTV V100% (Figure 1) and CI (Figure 2) for all three catheter displacement values. The dwell positions moving out of the PTV in the DTDC 0 plans have smaller weights relative to those in the DTDC 0.4 plans, resulting in smaller changes in PTV V100% and CI with catheter displacement.
Table 5

Change in dose volume histogram (DVH) and dose indices for a 3 mm catheter displacement

ParameterDTDC 0DTDC 0.4 p
PTV V100% 0.788 ± 0.751%–1.299 ± 0.916%≤ 0.0001
PTV V150% 1.786 ± 1.540%0.146 ± 0.537%≤ 0.0009
PTV V200% 1.051 ± 0.829%–0.034 ± 0.183%≤ 0.0021
Rectum V70% 0.078 ± 0.081%0.080 ± 0.090%≤ 0.8905
Urethra V120% –5.628 ± 4.878%–4.019 ± 5.047%≤ 0.0088
NT V100% –1.042 ± 1.314 cc1.314 ± 1.661 cc≤ 0.0454
NT V150% –1.033 ± 0.893 cc0.468 ± 0.504 cc≤ 0.0001
NT V200% –0.568 ± 0.674 cc0.171 ± 0.318 cc≤ 0.0052
CI0.002 ± 0.017–0.024 ± 0.022≤ 0.0011
DHI0.017 ± 0.0050.005 ± 0.006≤ 0.0033
ODI0.011 ± 0.0080.001 ± 0.002≤ 0.0035

PTV – planning target volume, NT – normal tissue, CI – conformity index, DHI – dose inhomogeneity index, ODI – overdose index, V100%, V150%, V200%, V70%, V120% – volume of relevant structure receiving 100%, 150%, 200%, 70%, and 120% of the prescribed isodose, respectively, DTDC – dwell time deviation constraint

Table 6

Change in dose volume histogram (DVH) and dose indices for a 7 mm catheter displacement

ParameterDTDC 0DTDC 0.4 p
PTV V100% –2.803 ± 2.516%–8.99 ± 2.283%≤ 0.0001
PTV V150% 2.199 ± 2.670%–1.262 ± 1.158%≤ 0.0002
PTV V200% 1.255 ± 1.260%–0.689 ± 0.427%≤ 0.0008
Rectum V70% 0.271 ± 0.324%0.265 ± 0.324%≤ 0.7793
Urethra V120% –6.461 ± 5.879%–4.333 ± 6.804%≤ 0.0205
NT V100% 1.072 ± 3.121 cc4.381 ± 1.996 cc≤ 0.0189
NT V150% –0.399 ± 1.639 cc1.507 ± 0.753 cc≤ 0.0041
NT V200% –0.471 ± 0.940 cc0.641 ± 0.517 cc≤ 0.0014
CI–0.046 ± 0.044–0.136 ± 0.046≤ 0.0005
DHI0.029 ± 0.0240.011 ± 0.016≤ 0.0007
ODI0.016 ± 0.0120.002 ± 0.004≤ 0.0020

PTV – planning target volume, NT – normal tissue, CI – conformity index, DHI – dose inhomogeneity index, ODI – overdose index, V100%, V150%, V200%, V70%, V120% – volume of relevant structure receiving 100%, 150%, 200%, 70%, and 120% of the prescribed isodose, respectively, DTDC – dwell time deviation constraint

Table 7

Change in dose volume histogram (DVH) and dose indices for a 14 mm catheter displacement

ParameterDTDC 0DTDC 0.4 p
PTV V100% –18.69 ± 5.197%–25.74 ± 5.313%≤ 0.0003
PTV V150% 0.522 ± 3.343%–3.589 ± 2.063%≤ 0.0001
PTV V200% 0.566 ± 1.455%–1.628 ± 0.786%≤ 0.0005
Rectum V70% 0.690 ± 0.917%0.665 ± 0.956%≤ 0.4049
Urethra V120% –4.657 ± 7.543%–0.592 ± 10.624%≤ 0.0523
NT V100% 8.928 ± 4.198 cc13.495 ± 2.333 cc≤ 0.0002
NT V150% 1.582 ± 1.210 cc3.981 ± 1.197 cc≤ 0.0008
NT V200% 0.453 ± 0.786 cc1.756 ± 0.747 cc≤ 0.0036
CI–0.251 ± 0.072–0.352 ± 0.064≤ 0.0003
DHI0.056 ± 0.0400.041 ± 0.039≤ 0.0013
ODI0.027 ± 0.0180.013 ± 0.010≤ 0.0043

PTV – planning target volume, NT – normal tissue, CI – conformity index, DHI – dose inhomogeneity index, ODI – overdose index, V100%, V150%, V200%, V70%, V120% – volume of relevant structure receiving 100%, 150%, 200%, 70%, and 120% of the prescribed isodose, respectively, DTDC – dwell time deviation constraint

Fig. 1

The change in PTV V100% as a function of catheter displacement for plans optimized with DTDC set to 0 and 0.4 (error bars showing 95% confidence interval)

Fig. 2

The change in PTV V200% as a function of catheter displacement for plans optimized with DTDC set to 0 and 0.4 (error bars showing 95% confidence interval)

The change in PTV V100% as a function of catheter displacement for plans optimized with DTDC set to 0 and 0.4 (error bars showing 95% confidence interval) The change in PTV V200% as a function of catheter displacement for plans optimized with DTDC set to 0 and 0.4 (error bars showing 95% confidence interval) Change in dose volume histogram (DVH) and dose indices for a 3 mm catheter displacement PTV – planning target volume, NT – normal tissue, CI – conformity index, DHI – dose inhomogeneity index, ODI – overdose index, V100%, V150%, V200%, V70%, V120% – volume of relevant structure receiving 100%, 150%, 200%, 70%, and 120% of the prescribed isodose, respectively, DTDC – dwell time deviation constraint Change in dose volume histogram (DVH) and dose indices for a 7 mm catheter displacement PTV – planning target volume, NT – normal tissue, CI – conformity index, DHI – dose inhomogeneity index, ODI – overdose index, V100%, V150%, V200%, V70%, V120% – volume of relevant structure receiving 100%, 150%, 200%, 70%, and 120% of the prescribed isodose, respectively, DTDC – dwell time deviation constraint Change in dose volume histogram (DVH) and dose indices for a 14 mm catheter displacement PTV – planning target volume, NT – normal tissue, CI – conformity index, DHI – dose inhomogeneity index, ODI – overdose index, V100%, V150%, V200%, V70%, V120% – volume of relevant structure receiving 100%, 150%, 200%, 70%, and 120% of the prescribed isodose, respectively, DTDC – dwell time deviation constraint Conversely, a DTDC value of 0.4 improves the robustness of the plans to changes in hotspots, reflected by statistically significant differences in changes to PTV V150%, PTV V200% (Figure 3), DHI, and ODI (Figure 4) compared to plans optimized with a DTDC value of 0 for catheter displacements up to 14 mm. This behavior can be explained by considering that the isolated dwell times at the end of the catheters often exist just outside the PTV before a catheter shift is implemented. Therefore, the V150% and V200% volumes are surrounding these dwell positions and, as a catheter shift is implemented, they move further away from the PTV and into healthy tissue.
Fig. 3

The change in CI as a function of catheter displacement for plans optimized with DTDC set to 0 and 0.4 (error bars showing 95% confidence interval)

Fig. 4

The change in ODI as a function of catheter displacement for plans optimized with DTDC set to 0 and 0.4 (error bars showing 95% confidence interval)

The change in CI as a function of catheter displacement for plans optimized with DTDC set to 0 and 0.4 (error bars showing 95% confidence interval) The change in ODI as a function of catheter displacement for plans optimized with DTDC set to 0 and 0.4 (error bars showing 95% confidence interval) For a catheter displacement of 3 mm, plans optimized with a DTDC value of 0.4 were found to be more robust in terms of NT V100%, NT V150%, and NT V200%. Conversely, for larger catheter shifts of 7 and 14 mm, the plans optimized with DTDC 0 were more robust. This is due to the fact that for a catheter shift of 3 mm, one of the large isolated dwell positions in the DTDC 0 plans moves into the normal tissue. However, for larger shifts, subsequent dwell positions moving into the normal tissue have significantly smaller dwell times compared to those in the DTDC 0.4 plans, resulting in smaller changes in NT V100%, NT V150%, and NT V200%. The urethra V120% was more sensitive to catheter displacements than the rectum V70% for both values of DTDC. A DTDC value of 0.4 improved the robustness of the plans to changes in urethra V120% compared to plans optimized with a DTDC value of 0, with statistically significant differences for catheter displacements of 3 and 7 mm. There was also no statistically significant difference found between DTDC values for the rectum V70%.

Discussion

As expected, the plan modulation index (M) was observed to decrease with an increased value of DTDC. The calculated values of M = 0.44 ± 0.07 (DTDC = 0) and M = 0.20 ± 0.06 (DTDC = 0.4) are in close agreement with those found in a previous study by Smith et al. [9]. The increase value of DTDC was also observed to increase the PTV V100%, V150%, and V200%. The increase in PTV V150% of 4% and PTV V200% of 1.5% when changing the DTDC value from 0 to 0.4 was also observed in the same study [9]. This increase resulted in DVH indices that were still clinically acceptable for treatment according to our local protocol as outlined in Table 2. Contrary to expectation, the DTDC value of 0 produced plans that were more robust to catheter displacements in terms of target coverage. However, for a catheter displacement of 3 mm, the average CI was 0.693 ± 0.049 and 0.723 ± 0.045 for DTDC = 0 and DTDC = 0.4, respectively. Our center has implemented a clinical protocol, in which catheter displacements ≥ 3 mm are corrected for, immediately prior to treatment by adjusting the indexer length on the treatment control system. Therefore, if used in combination with this catheter correction protocol, plans optimized with a DTDC value of 0.4 may be clinically superior to those optimized with DTDC = 0, especially when considering that using a DTDC value of 0.4 improves robustness to changes in hotspots and dose to OARs and healthy tissue. One recent advance in the field of HDR prostate brachytherapy has been the use of 3D trans-rectal ultrasound (TRUS) based treatment planning [18]. The use of this technique has been shown to significantly reduce the time between imaging and treatment compared to CT based treatment planning. Milickovic et al. [18] have shown that for an average time between imaging and treatment of 51.2 minutes, the average needle displacement was found to be 1 mm. This displacement is small relative to those noted in other studies [11, 14], and is likely due to the reduction in time between imaging and treatment. Therefore, one current initiative of our group is to reduce the time between imaging and treatment, and the introduction of 3D TRUS based planning is being investigated. A previous study by our group [11] has shown that catheter displacements in the cranial direction occurred for only 3 of 48 cases, with the remainder occurring in the caudal direction. One limitation of this study is that catheter displacements were only considered along the longitudinal axis in the caudal direction. Catheter shifts in the lateral and anterior-posterior directions due to edema were also not considered. A further limitation is that only two values of DTDC were considered. Preliminary calculations showed that small changes in DTDC, e.g. from 0.2-0.4 do not significantly affect the robustness of the plans to catheter displacements. Furthermore, values of 0 and 0.4 were chosen to reflect the change that is currently occurring in our clinical protocol. Historically, plans have been optimized using a DTDC value of 0; however previous studies [9] have shown that a DTDC value of 0.4 gives plan modulation equivalent to graphical optimization without significantly compromising plan quality.

Conclusions

The results of this study indicate that the robustness of clinically acceptable prostate plans to catheter displacements in the caudal direction are dependent on the DTDC parameter. A DTDC value of 0 improves the robustness of PTV coverage to catheter displacements relative to a DTDC value of 0.4. Whereas a DTDC value of 0.4 improves the robustness of the plans to changes in hotspots compared to a DTDC value of 0. For a catheter displacement of 3 mm, plans optimized with a DTDC value of 0.4 were found to be more robust in terms of the dose to normal tissue. However, for larger catheter shifts, the plans optimized with DTDC 0 were more robust, due to larger shifts moving relatively small weight dwell positions into the normal tissue compared to the DTDC 0.4 plans. When used in combination with a pre-treatment catheter displacement correction protocol and a tolerance of 3 mm, a DTDC value of 0.4 may produce clinically superior plans. In future work, attempts will be made to measure the actual dwell times delivered by the afterloader as compared to those calculated by the TPS for a range of DTDC values and the effect on the dose examined.
  16 in total

1.  Use of cone-beam imaging to correct for catheter displacement in high dose-rate prostate brachytherapy.

Authors:  Rick Holly; Gerard C Morton; Raxa Sankreacha; Niki Law; Thomas Cisecki; D Andrew Loblaw; Hans T Chung
Journal:  Brachytherapy       Date:  2010-12-28       Impact factor: 2.362

2.  High dose rate afterloading brachytherapy for prostate cancer: catheter and gland movement between fractions.

Authors:  Peter J Hoskin; Peter J Bownes; Peter Ostler; Ken Walker; Linda Bryant
Journal:  Radiother Oncol       Date:  2003-09       Impact factor: 6.280

3.  A small tolerance for catheter displacement in high-dose rate prostate brachytherapy is necessary and feasible.

Authors:  Albert Tiong; Sean Bydder; Martin Ebert; Nikki Caswell; David Waterhouse; Nigel Spry; Peter Camille; David Joseph
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-07-18       Impact factor: 7.038

4.  Prostate HDR brachytherapy catheter displacement between planning and treatment delivery.

Authors:  May Whitaker; George Hruby; Aimee Lovett; Nitya Patanjali
Journal:  Radiother Oncol       Date:  2011-08-31       Impact factor: 6.280

5.  A conformation number to quantify the degree of conformality in brachytherapy and external beam irradiation: application to the prostate.

Authors:  A van't Riet; A C Mak; M A Moerland; L H Elders; W van der Zee
Journal:  Int J Radiat Oncol Biol Phys       Date:  1997-02-01       Impact factor: 7.038

6.  A comparison of inverse optimization algorithms for HDR/PDR prostate brachytherapy treatment planning.

Authors:  Anna M Dinkla; Rob van der Laarse; Emmie Kaljouw; Bradley R Pieters; Kees Koedooder; Niek van Wieringen; Arjan Bel
Journal:  Brachytherapy       Date:  2014-10-22       Impact factor: 2.362

7.  A dose homogeneity index for evaluating 192Ir interstitial breast implants.

Authors:  A Wu; K Ulin; E S Sternick
Journal:  Med Phys       Date:  1988 Jan-Feb       Impact factor: 4.071

8.  Dramatic increase in prostate cancer cases by 2021.

Authors:  Harvey Quon; Andrew Loblaw; Robert Nam
Journal:  BJU Int       Date:  2011-04-20       Impact factor: 5.588

9.  The influence of the dwell time deviation constraint (DTDC) parameter on dosimetry with IPSA optimisation for HDR prostate brachytherapy.

Authors:  Ryan L Smith; Vanessa Panettieri; Craig Lancaster; Natasha Mason; Rick D Franich; Jeremy L Millar
Journal:  Australas Phys Eng Sci Med       Date:  2014-12-07       Impact factor: 1.430

10.  Measurement of craniocaudal catheter displacement between fractions in computed tomography-based high dose rate brachytherapy of prostate cancer.

Authors:  Yongbok Kim; I-Chow J Hsu; Jean Pouliot
Journal:  J Appl Clin Med Phys       Date:  2007-09-17       Impact factor: 2.102

View more
  4 in total

1.  An audit of high dose-rate prostate brachytherapy treatment planning at six Swedish clinics.

Authors:  Frida Dohlmar; Sakarias Johansson; Torbjörn Larsson; Michael Sandborg; Åsa Carlsson Tedgren
Journal:  J Contemp Brachytherapy       Date:  2021-02-18

2.  Influence of dwell time homogeneity error weight parameter on treatment plan quality in inverse optimized high-dose-rate cervix brachytherapy using SagiPlan.

Authors:  Mohammad Amin Mosleh-Shirazi; Elham Shahcheraghi-Motlagh; Mohammad Hadi Gholami; Alireza Shakibafard; Sareh Karbasi; Reza Fardid
Journal:  J Contemp Brachytherapy       Date:  2019-06-28

3.  HDR prostate brachytherapy plan robustness and its effect on in-vivo source tracking error thresholds: A multi-institutional study.

Authors:  Joel Poder; Dylan Koprivec; Yashiv Dookie; Andrew Howie; Dean Cutajar; Antonio L Damato; Nicolas Côté; Marco Petasecca; Joseph Bucci; Anatoly Rosenfeld
Journal:  Med Phys       Date:  2022-04-19       Impact factor: 4.506

4.  Comparison of the IPSA and HIPO algorithms for interstitial tongue high-dose-rate brachytherapy.

Authors:  Chang Heon Choi; So-Yeon Park; Jong Min Park; Hong-Gyun Wu; Jin-Ho Kim; Jung-In Kim
Journal:  PLoS One       Date:  2018-10-04       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.