An experimental extrapolation technique using the Gafchromic EBT3 film for relative output factor measurements in small x-ray fields.

Johnny E Morales1, Martin Butson2, Scott B Crowe3, Robin Hill2, J V Trapp4.   

Abstract

PURPOSE: An experimental extrapolation technique is presented, which can be used to determine the relative output factors for very small x-ray fields using the Gafchromic EBT3 film.
METHODS: Relative output factors were measured for the Brainlab SRS cones ranging in diameters from 4 to 30 mm(2) on a Novalis Trilogy linear accelerator with 6 MV SRS x-rays. The relative output factor was determined from an experimental reducing circular region of interest (ROI) extrapolation technique developed to remove the effects of volume averaging. This was achieved by scanning the EBT3 film measurements with a high scanning resolution of 1200 dpi. From the high resolution scans, the size of the circular regions of interest was varied to produce a plot of relative output factors versus area of analysis. The plot was then extrapolated to zero to determine the relative output factor corresponding to zero volume.
RESULTS: Results have shown that for a 4 mm field size, the extrapolated relative output factor was measured as a value of 0.651 ± 0.018 as compared to 0.639 ± 0.019 and 0.633 ± 0.021 for 0.5 and 1.0 mm diameter of analysis values, respectively. This showed a change in the relative output factors of 1.8% and 2.8% at these comparative regions of interest sizes. In comparison, the 25 mm cone had negligible differences in the measured output factor between zero extrapolation, 0.5 and 1.0 mm diameter ROIs, respectively.
CONCLUSIONS: This work shows that for very small fields such as 4.0 mm cone sizes, a measureable difference can be seen in the relative output factor based on the circular ROI and the size of the area of analysis using radiochromic film dosimetry. The authors recommend to scan the Gafchromic EBT3 film at a resolution of 1200 dpi for cone sizes less than 7.5 mm and to utilize an extrapolation technique for the output factor measurements of very small field dosimetry.

Entities:  

Mesh:

Year:  2016        PMID: 27487885     DOI: 10.1118/1.4958679

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Alkali Metal Salts of 10,12-Pentacosadiynoic Acid and Their Dosimetry Applications.

Authors:  Amy V Hall; Osama M Musa; David K Hood; David C Apperley; Dmitry S Yufit; Jonathan W Steed
Journal:  Cryst Growth Des       Date:  2021-03-25       Impact factor: 4.076

2.  Survey of 5 mm small-field output factor measurements in Australia.

Authors:  Christopher P Oliver; Duncan J Butler; Viliami Takau; Ivan Williams
Journal:  J Appl Clin Med Phys       Date:  2018-01-25       Impact factor: 2.102

3.  A novel method for the determination of field output factors and output correction factors for small static fields for six diodes and a microdiamond detector in megavoltage photon beams.

Authors:  Božidar Casar; Eduard Gershkevitsh; Ignasi Mendez; Slaven Jurković; M Saiful Huq
Journal:  Med Phys       Date:  2018-12-24       Impact factor: 4.071

4.  An Experimental Slope Method for a More Accurate Measurement of Relative Radiation Doses using Radiographic and Radiochromic Films and Its Application to Megavoltage Small-Field Dosimetry.

Authors:  Raghavendra Holla; D Khanna; Bhaskaran K Pillai; K V Jafar Ali; P S Renil Mon; C O Clinto; Tharmarnadar Ganesh
Journal:  J Med Phys       Date:  2019 Jul-Sep

5.  Expired EBT3 Films' Sensitivity for the Measurement of X-ray and UV Radiation: An Optical Analysis.

Authors:  Hamad Yahia Abu Mhanna; Ahmad Fairuz Omar; Yasmin Md Radzi; Hanan Fawaz Akhdar; Haytham Al Ewaidat
Journal:  Materials (Basel)       Date:  2022-04-15       Impact factor: 3.748

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.