Kirill V Kovtunov1, Alexey S Romanov1, Oleg G Salnikov1, Danila A Barskiy2, Eduard Y Chekmenev2, Igor V Koptyug1.
Abstract
1H MRI of gases can potentially enable functional lung imaging to probe gas ventilation and other functions. In this work, 1H MR images of hyperpolarized and thermally polarized propane gas were obtained using UTE (ultrashort echo time) pulse sequence. A 2D image of thermally polarized propane gas with ~0.9×0.9 mm2 spatial resolution was obtained in less than 2 seconds, demonstrating that even non-hyperpolarized hydrocarbon gases can be successfully utilized for conventional proton MRI. The experiments were also performed with hyperpolarized propane gas and demonstrated acquisition of high-resolution multi-slice FLASH 2D images in ca. 510 s and non slice-selective 2D UTE MRI images in ca. 2 s. The UTE approach adopted in this study can be potentially used for medical lung imaging. Furthermore, the possibility to combine UTE with selective suppression of 1H signals from one of the two gases in a mixture is demonstrated in this MRI study. The latter can be useful for visualizing industrially important processes where several gases may be present, e.g., gas-solid catalytic reactions.Entities:
Year: 2016 PMID: 27478870 PMCID: PMC4966642 DOI: 10.18383/j.tom.2016.00112
Source DB: PubMed Journal: Tomography ISSN: 2379-1381
Figure 1.The schematic representation of the experimental setup for producing parahydrogen-induced polarization (PHIP) hyperpolarized (HP) propane via heterogeneous pairwise hydrogenation of propene with parahydrogen (A). An overlay of hydrogen 1 (1H) magnetic resonance imaging (1H MRI) FLASH image of hyperpolarized (HP) propane flowing into the 10-mm nuclear magnetic resonance (NMR) tube via 1/16″ OD Teflon capillary (B). The field of view (FOV) was 3.1 × 3.1 cm with 256 × 32 matrix size, and the total acquisition time was ∼510 seconds. Note that the NMR tube is schematically shown, and its length does not correlate with the actual scale of the 2-dimensional (2D) magnetic resonance (MR) image.
Figure 2.Heterogeneous hydrogenation of propene to propane with pH2 over Rh/TiO2 catalyst with partial preservation of spin order of parahydrogen in the final HP product (A). 1H NMR spectra of HP propane (blue) and thermally polarized propane (black) obtained in heterogeneous hydrogenation of propene with parahydrogen; the signal enhancement was ∼30-fold (B). 1H MRI of HP propane flowing through a 1/16″ OD Teflon capillary inside a 10-mm NMR tube obtained with a FLASH pulse sequence (C). The FOV was 3.1 × 2.5 cm with 256 × 32 matrix size, and the total acquisition time was ∼510 seconds with slice thickness of ∼0.7 cm. The signal intensity color scale ranges from black (zero) to blue (maximum).
Figure 3.MR images of 15-mm NMR tube filled with thermally polarized propane (A) and PHIP HP propane (the same experimental setup as for Figure 1) (B). The FOV was 3 × 3 cm with 32 × 32 matrix size, the total acquisition time was ∼2 seconds, and the slice thickness in the z direction was 10 mm.
Figure 4.1H MR image (A) and corresponding 1H NMR spectrum of a sample comprising a 10-mm NMR tube with propylene placed inside a 15-mm tube with propane (B). 1H MR image (C) and corresponding 1H NMR spectrum (D) of the same phantom with application of selective suppression pulse for propane NMR signals. The FOV for both images was 5 × 5 cm with 128 × 128 matrix size and 10-mm slice thickness along the z axis; the total acquisition time was ∼8 seconds. The 1H NMR spectrum was obtained using the same ultrashort echo time (UTE) pulse sequence as that for 1H MRI but with the gradients switched off and the spectral bandwidth appropriately reduced.