Transcutaneous PCO2 Monitoring in Newborn Infants During General Anesthesia Is Technically Feasible.

Victoria Karlsson1, Bengt Sporre, Johan Ågren.   

Abstract

BACKGROUND: Transcutaneous (TC) measurement of PCO2 (TC PCO2) is a well-established method to monitor assisted ventilation in neonatal intensive care, but its use in the operating room is limited, and the data regarding its performance during general anesthesia of the newborn are lacking. The aim of this study is to evaluate the performance of continuous TC PCO2 monitoring during general anesthesia in newborn infants.
METHODS: Infants (n = 25) with a gestational age of 23 to 41 weeks and a birth weight of 548 to 4114 g were prospectively enrolled. During general anesthesia and surgery, TC PCO2 was measured continuously and recorded at 1-minute intervals. Five-minute mean values were compared with simultaneously obtained blood gas (BG) analyses of PCO2. Only the first paired TC and BG samples were used in this analysis. We defined precision as 2.1 times the standard deviation of the difference of the 2 samples. P < .01 was considered statistically significant.
RESULTS: We obtained samples from 25 infants. The difference between TC and BG was 0.3 ± 0.7 kPa (mean ± standard deviation) giving a precision of 1.47 kPa. Nineteen of twenty-five (76%) sample pairs displayed a difference of <1 kPa (99% confidence interval, 48%-92%, P = .016). The difference in paired samples was similar for different gestational and postnatal ages and did not appear to be affected by electrocautery.
CONCLUSIONS: In this small study, we did not demonstrate that TC CO2 monitoring was accurate at P < .01. This partly reflects the small size of the study, resulting in wide 99% confidence bounds.

Entities:  

Mesh:

Year:  2016        PMID: 27464976      PMCID: PMC5017270          DOI: 10.1213/ANE.0000000000001462

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


Monitoring of assisted ventilation is of vital importance in the operating room (OR). In the newborn, several factors contribute to the unique challenge of providing appropriate gas exchange during general anesthesia. These include (but are not limited to) the process of postnatal cardiorespiratory adaptation with persistence of fetal shunting, the neonate’s propensity for atelectasis formation, and the use of anesthesia delivery systems primarily designed for use in larger patients. Also, in abdominal surgery, either changes in abdominal pressure or surgical manipulation might cause rapid changes in minute ventilation and ventilation-perfusion mismatch and impact on gas exchange. Carbon dioxide is a potent vasoactive substance, and there is a clear relation between the partial pressure of carbon dioxide and cerebral perfusion.[1] Both hypo- and hypercapnia are related to adverse neurologic outcome.[2] Most importantly, induced hypocapnia in the neonate is associated with cerebral artery vasospasm, reduced cerebral perfusion, and an increased risk of ischemic periventricular white matter injury.[3,4] Because the magnitude of CO2 changes are not easily anticipated or estimated during assisted ventilation, Pco2 needs to be continuously monitored to ensure proper management. During anesthesia and surgery, this is most commonly performed by continuous measurement of end-tidal (ET) Pco2 and complemented by intermittent analysis of Pco2 in arterial or capillary blood. The ET method is well established for use in adults and older children,[5,6] but many of the factors that contribute to the complex respiratory physiology of the newborn and the presence of pulmonary immaturity and/or disease also impact on the accuracy of ET Pco2 monitoring.[7] Consequently, in the neonatal intensive care unit (NICU), continuous Pco2 is almost exclusively obtained by transcutaneous (TC) measurements. In the NICU, TC monitoring of Pco2 is considered to be the standard of care both for absolute measurement and trend analysis, and TC readings have been found to correlate well with blood Pco2.[8-10] However, the use of TC Pco2 monitoring in intraoperative care of neonates is limited, and to our knowledge, its accuracy has not been established in this population. We hypothesized that TC Pco2 monitoring would be a relevant method in the operating room as well. To assess the accuracy of TC measurements, we compared TC Pco2 and blood gas (BG) analyses obtained during general anesthesia in newborn infants.

METHODS

The investigation was approved of by the regional ethical review board, and in all cases, written parental consent was obtained before study inclusion. The study was not registered because patients were not assigned to treatment groups. Infant Data

Subjects

We prospectively and consecutively enrolled neonates with a postconceptional age (PCA) of <44 weeks who were admitted to the NICU at Uppsala University Children’s Hospital and scheduled for surgery. Enrollment was consecutive but limited by the availability of the primary investigator (V.K.). Data on each infant’s gestational age, birth weight, PCA, and weight at day of study are given in Table 1.
Table 1.

Infant Data

Transcutaneous Pco2 Monitoring

TC partial pressure of carbon dioxide (TC Pco2; kilopascal) was measured continuously by use of the TC technique using the E5280 probe (TCM 4/40, Radiometer, Denmark) with a probe temperature of 43°C according to manufacturer recommendations. Heating of the probe increases capillary blood flow as well as the diffusion of CO2 and O2 through the skin. Before probe placement and at 4-hour intervals, the electrode was calibrated against air and test gas. If calibration failed, the electrode membrane was replaced, and the electrode was recalibrated. The probe was placed on the upper chest, and the measurement site was changed at least every hour to minimize skin redness/thermal injury. After each probe placement, the electrode was allowed to stabilize for 15 minutes before recording TC Pco2 values at 1-minute intervals.

Blood Gas CO2 Analysis

BG analysis was performed using capillary blood samples (100 μL/sample) obtained from each infant’s warmed foot and was immediately analyzed for Pco2 (BG Pco2; kilopascal) using an automated BG analyzer (ABL 800, Radiometer). By warming the foot, the capillary blood is considered arterialized and has been shown to correlate very well with arterial blood CO2 in newborn infants.[11]

Measurement Procedure

In preparation for anesthesia, the TC probe was placed simultaneously with the application of electrocardiogram leads and pulse oximetry. Anesthesia was induced with thiopental (3 mg/kg), atropine (0.02 mg/kg), fentanyl (1–10 μg/kg), and neuromuscular block obtained with atracurium (0.5 mg/kg). Anesthesia was maintained with sevoflurane supplemented with intermittent dosing of fentanyl. All patients were intubated using uncuffed endotracheal tubes (2.5–3.0 mm) with pharyngeal pack and were managed using an anesthesia delivery system and ventilator (FLOW-i, Maquet, Sweden) in pressure control (infants weighing <2.8 kg) or pressure-regulated volume control mode (infants weighing >2.8 kg). Tidal volumes were set at 7 mL/kg body weight with a respiratory rate of 40–60/min and adjusted according to standard monitoring (ET Pco2). Infants already on the ventilator before anesthesia were initially kept on their NICU ventilator settings. The first blood sampling was performed after the anesthesia had been induced, the infant was intubated (including placement of a pharyngeal pack), maintained on the ventilator, and before the start of surgery. Additional blood sampling (1–2 per infant) was undertaken during the surgicy and recovery, but it was not included further in this analysis. All study data monitoring and collection were performed by the same person who had no role in taking care of the patient. The OR team was blinded to the TC Pco2 data but it had access to the BG data.

Treatment of Data

Statistical analysis was performed using Statistical Package for Social Science, SPSS version 20 (IBM Corp, Armonk, NY). The TC Pco2 value was calculated as the mean of the 5 TC Pco2 values immediately before BG sampling. Limits of agreement were evaluated by the use of Bland–Altman analysis.[12] The analysis was based on the difference between paired measurements of TC and BG Pco2 (ie, TC Pco2BG Pco2). Bias was calculated as the average difference. Precision was calculated as the standard deviation of the difference × 2.1 (calculated as the 2-sided t-inverse statistical for P = .05 and 24 degrees of freedom × square root (1 + 1/25). Differences were assessed by Student paired t test. Data are expressed as mean ± standard deviation or median (range). A P value of <.01 was considered statistically significant. A priori, we considered an accuracy ±1 kPa to be clinically acceptable based on previously reported results from older infants during anesthesia[13,14] and from the NICU environment.[8] The number of subjects was based on convenience. There was no power analysis before the study.

RESULTS

The analyzed data include BG and TC Pco2 pairs from 25 infants (Figure 1). Bland–Altman analysis revealed a bias of 0.3 ± 0.7 kPa and a precision of ±1.47 kPa (Figure 2). The TC Pco2 to BG Pco2 difference was ≤1 kPa in 19 of 25 (76%) data pairs (99% confidence interval, 48%–92%, P = .016 versus random chance (ie, 50%).
Figure 1.

Correlation between transcutaneous (TC) and blood gas (BG) Pco2 data pairs (n = 25). The diagonal line is the line of identity (X = Y).

Figure 2.

Bland–Altman plot of the difference between transcutaneous (TC) and blood gas (BG) Pco2 (n = 25) versus the average of the 2. The lines represent the TC to BG bias (solid line) and the precision (±2.10 × SD; broken lines).

Correlation between transcutaneous (TC) and blood gas (BG) Pco2 data pairs (n = 25). The diagonal line is the line of identity (X = Y). Bland–Altman plot of the difference between transcutaneous (TC) and blood gas (BG) Pco2 (n = 25) versus the average of the 2. The lines represent the TC to BG bias (solid line) and the precision (±2.10 × SD; broken lines). Transcutaneous to Blood Gas Pco2 Difference Transcutaneous (TC) and blood gas (BG) Pco2 in infants with a postnatal age of <1 week or >1 week. The diagonal line is the line of identity (X = Y). Transcutaneous (TC) and blood gas (BG) Pco2 in preterm or term infants. The diagonal line is the line of identity (X = Y). The TC to BG Pco2 bias (Table 2) and data pairs are displayed further in subgroups depending on postnatal age (<1 week or >1 week; Figure 3) and gestational age (preterm or term; Figure 4).
Table 2.

Transcutaneous to Blood Gas Pco2 Difference

Figure 3.

Transcutaneous (TC) and blood gas (BG) Pco2 in infants with a postnatal age of <1 week or >1 week. The diagonal line is the line of identity (X = Y).

Figure 4.

Transcutaneous (TC) and blood gas (BG) Pco2 in preterm or term infants. The diagonal line is the line of identity (X = Y).

DISCUSSION

The present relatively small investigation does not demonstrate that TC measurements accurately reflect Pco2 in newborn infants during general anesthesia. The precision was 1.47 kPa, less than our accepted accuracy of 1 kPa. Although the TC Pco2 determination in 76% of the subjects was within our acceptable range of ±1 kPa of the concurrent BG Pco2, this did not reach statistical significance at P < .01, as demonstrated by the lower confidence bound of 48% (less than random sample). The recognition that extremes of Pco2, even for brief periods of time, are associated with long-term neonatal morbidity emphasizes the need for reliable and continuous monitoring of Pco2 during general anesthesia. Arterial BG analysis from intermittent sampling from an indwelling catheter is considered the gold standard in measuring Pco2. However, the dynamics of neonatal anesthetic management mandate the use of continuous measurements. Also, placing arterial lines in the tiniest infants is often painful, cumbersome, resource consuming, and it carries a risk of vascular thrombosis.[15] The most commonly applied method in anesthesia, ET CO2 monitoring, is known to be less accurate in infants with lung disease, high respiratory rate, and small tidal volumes,[16] and it is therefore of variable value at times. TC monitoring of Pco2 is well established in the NICU, and several studies have demonstrated a good correlation between TC and BG Pco2 in this setting.[8] The method is independent of pulmonary disease, tidal volume, high respiratory rates, and ventilator mode (eg, small tidal volumes or high-frequency oscillatory ventilation), but it may be influenced by skin edema and hypoperfusion.[17] Two previous studies have evaluated TC for intraoperative use in pediatric patients.[13,18] Both studies report a good agreement between TC and BG Pco2, but those studies did not include any newborn infants. The authors of these studies recommend using TC as a complement to ET monitoring of Pco2. In a review from the study by Molloy and Deakins[19] in 2006 conclude that TC is superior to ET and promote the use of TC for noninvasive trend monitoring of Pco2, particularly in infants with pulmonary disease. Arguably, management of a newborn in the NICU is distinct from that in the OR, where access to the small patient is limited by sterile draping and for the purpose of maintaining thermal stability. To measure TC Pco2, the electrode needs to be calibrated, heated, and repositioned at regular intervals. These procedures did not create a barrier to the performance of the measurements in our study, suggesting that TC Pco2 monitoring requires minimal staff training. To conclude, because of the small size of our study, we were unable to demonstrate that TC monitoring of Pco2 in the OR was sufficiently accurate for clinical use. A larger study is required to address this question.

DISCLOSURES

Name: Victoria Karlsson, RN, MMSc. Contribution: This author helped design the study, collect the data, analyze the data, and prepare the manuscript. Name: Bengt Sporre, MD. Contribution: This author helped design the study, collect the data, and prepare the manuscript. Name: Johan Ågren, MD, PhD. Contribution: This author helped design the study, review the data, analyze the data, and prepare the manuscript. This manuscript was handled by: James A. DiNardo, MD.
  19 in total

Review 1.  Noninvasive monitoring of oxygen and carbon dioxide.

Authors:  A O Soubani
Journal:  Am J Emerg Med       Date:  2001-03       Impact factor: 2.469

Review 2.  Transcutaneous carbon dioxide monitoring.

Authors:  J M Rennie
Journal:  Arch Dis Child       Date:  1990-04       Impact factor: 3.791

Review 3.  Intravascular catheter complications in the neonatal intensive care unit.

Authors:  Marcus C Hermansen; Mary Goetz Hermansen
Journal:  Clin Perinatol       Date:  2005-03       Impact factor: 3.430

4.  Transcutaneous PCO2 and PO2: a multicenter study of accuracy.

Authors:  B W Palmisano; J W Severinghaus
Journal:  J Clin Monit       Date:  1990-07

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Noninvasive monitoring of carbon dioxide in infants and children with congenital heart disease: end-tidal versus transcutaneous techniques.

Authors:  John Wilson; Pierantonio Russo; JoAnne Russo; Joseph D Tobias
Journal:  J Intensive Care Med       Date:  2005 Sep-Oct       Impact factor: 3.510

7.  Noninvasive monitoring of carbon dioxide during mechanical ventilation in older children: end-tidal versus transcutaneous techniques.

Authors:  J W Berkenbosch; J Lam; R S Burd; J D Tobias
Journal:  Anesth Analg       Date:  2001-06       Impact factor: 5.108

Review 8.  Transcutaneous carbon dioxide monitoring in infants and children.

Authors:  Joseph D Tobias
Journal:  Paediatr Anaesth       Date:  2009-02-19       Impact factor: 2.556

9.  PaCO2 and neurodevelopment in extremely low birth weight infants.

Authors:  Lara A McKee; Jorge Fabres; George Howard; Myriam Peralta-Carcelen; Waldemar A Carlo; Namasivayam Ambalavanan
Journal:  J Pediatr       Date:  2009-05-17       Impact factor: 4.406

10.  Significance of hypocarbia in the development of periventricular leukomalacia in preterm infants.

Authors:  Christina Giannakopoulou; Eftichia Korakaki; Antonia Manoura; Stellios Bikouvarakis; Maria Papageorgiou; Dimitrios Gourgiotis; Eleftheria Hatzidaki
Journal:  Pediatr Int       Date:  2004-06       Impact factor: 1.524

View more
  5 in total

1.  Beyond Anesthesia Toxicity: Anesthetic Considerations to Lessen the Risk of Neonatal Neurological Injury.

Authors:  Mary Ellen McCann; Jennifer K Lee; Terrie Inder
Journal:  Anesth Analg       Date:  2019-11       Impact factor: 5.108

2.  Transcutaneous CO2 versus end-tidal CO2 in neonates and infants undergoing surgery: a prospective study.

Authors:  Arvind Chandrakantan; Ronald Jasiewicz; Ruth A Reinsel; Kseniya Khmara; Jonathan Mintzer; Joseph D DeCristofaro; Zvi Jacob; Peggy Seidman
Journal:  Med Devices (Auckl)       Date:  2019-05-06

3.  Validation of a new combined transcutaneous tcPCO2 and tcPO2 sensor in children in the operating theater.

Authors:  Jan J van Wijk; Willem van Weteringen; Sanne E Hoeks; Lonneke M Staals
Journal:  Paediatr Anaesth       Date:  2021-12-15       Impact factor: 2.129

4.  An Update on Patent Ductus Arteriosus and What is Coming Next.

Authors:  Ömer Erdeve; Emel Okulu; Yogen Singh; Richard Sindelar; Mehmet Yekta Oncel; Gianluca Terrin; Giovanni Boscarino; Ali Bülbül; Hannes Sallmon; Begüm Atasay; Fahri Ovalı; Ronald I Clyman
Journal:  Turk Arch Pediatr       Date:  2022-03

Review 5.  Current state of noninvasive, continuous monitoring modalities in pediatric anesthesiology.

Authors:  Jan J van Wijk; Frank Weber; Robert J Stolker; Lonneke M Staals
Journal:  Curr Opin Anaesthesiol       Date:  2020-12       Impact factor: 2.733

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.