INTRODUCTION: Laser Vision Correction is not always possible for the treatment of myopia. The aim of this paper is to present results after MyoRing implantation in myopic eyes who in their majority were not eligible for LVC. Safety, effectivity and patient selection is discussed. The aim of the treatment was to be spectacle independent in everyday-life.Entities:
Keywords: CISIS; Corrección de la visión con láser; LASIK; LIO fáquicas; Miopía; MyoRing; Myopia; Phakic IOL
Mesh:
Year: 2016 PMID: 27461520 PMCID: PMC5484779 DOI: 10.1016/j.optom.2016.06.003
Source DB: PubMed Journal: J Optom ISSN: 1989-1342
The nomogram shows the dimensions of the MyoRing used for a certain refractive correction and the number of eyes treated with each MyoRing dimension.
| MyoRing diameter (mm) | MyoRing thickness (microns) | SE range (dioptres) | Number of treatments |
|---|---|---|---|
| 5 | 280 | −9.0 to −13.0 | 14 |
| 6 | 280 | −5.5 to −6.5 | 3 |
| 7 | 280 | −2.5 to −3.5 | 2 |
Preoperative and postoperative spherical equivalent (SE) and corrective power in dioptres for each MyoRing dimension used in the study.
| MyoRing diameter (mm) | MyoRing thickness (microns) | SE preoperatively (dioptres) | SE postoperatively (dioptres) | Corrected power (dioptres) |
|---|---|---|---|---|
| 5 | 280 | −11.72 ± 2.6 | −0.88 ± 1.2 | 10.8 ± 1.4 |
| 6 | 280 | −6.83 ± 0.8 | −0.67 ± 0.8 | 6.17 ± 0.2 |
| 7 | 280 | −2.81 ± 0.3 | +0.5 ± 0.0 | 3.31 ± 0.3 |
The values show mean and standard deviation.
Preoperative and postoperative sphere (sph) and cylinder (cyl) in dioptres for each MyoRing dimension used in the study.
| MyoRing diameter (mm) | MyoRing thickness (microns) | sph preoperatively (dioptres) | cyl preoperatively (dioptres) | sph postoperatively (dioptres) | cyl postoperatively (dioptres) |
|---|---|---|---|---|---|
| 5 | 280 | −11.00 ± 2.7 | −1.45 ± 1.2 | −0.45 ± 0.4 | −0.88 ± 0.9 |
| 6 | 280 | −7.58 ± 2.5 | −1.75 ± 1.3 | −1.42 ± 2 | −1.33 ± 0.6 |
| 7 | 280 | −2.5 ± 0.4 | −0.63 ± 0.2 | 1.0 ± 0.4 | −1.0 ± 0.7 |
The values show mean and standard deviation.
Figure 1Uncorrected distance visual acuity (UDVA) before and after MyoRing treatment.
Figure 2Corrected distance visual acuity (CDVA) before and after MyoRing treatment.
Preoperative and postoperative uncorrected (UDVA) and corrected (CDVA) distance visual acuity in logMAR.
| UDVA (logMAR) | CDVA (logMAR) | |
|---|---|---|
| Preoperatively | 1.77 ± 0.5 | 0.05 ± 0.1 |
| Postoperatively | 0.23 ± 0.3 | 0.05 ± 0.1 |
The values show mean and standard deviation.