Literature DB >> 27460679

Deviations of different systems for guided implant surgery.

V Laederach1, K Mukaddam1, M Payer2, A Filippi1, S Kühl1.   

Abstract

OBJECTIVES: To compare the deviation of different systems for Guided Implant Surgery (GIS) related to the specific tolerance between drills and sleeves.
MATERIAL AND METHODS: Four different systems for GIS and their appropriate sleeves were used: Camlog Guide (CG), Straumann Guided Surgery (SG), SIC Guide (SIG), and NobelGuide (NG). System-appropriate metal sleeves were inserted into plexiglass boxes, and guided drilling procedure was performed (i) holding the drills in the most centric position of the sleeves and (ii) applying forces eccentrically. Digital microscope images of the plexiglass boxes were taken and axial deviations were calculated based on the Pythagorean Theorem, whereas coronal and apical deviations were measured with a corresponding software-device and calculated by subtracting the measured deviations from the original diameter of the drills. Statistically significant differences between centric and eccentric drilling were determined applying the t-test for independent data.
RESULTS: The axial deviation ranged from 0° (SG) to 5.64° (CG). The apical deviations varied between 0.01 mm (SIG) and 3.2 mm (NG) and the coronal deviations ranged from 0.01 mm (SIG) to 1.60 mm (NG). In terms of angular deviation, there were statistically significant differences between centric and eccentric drilling for all four systems. Coronal and apical deviations, showed no statistical significance between centric and eccentric drilling for SIG and NG, in contrast to CG and SG.
CONCLUSIONS: The clinician may have considerable impact on the accuracy of GIS when applying eccentric forces.
© 2016 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  accuracy; drill guide; guided surgery; sleeve; tolerance

Mesh:

Year:  2016        PMID: 27460679     DOI: 10.1111/clr.12930

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  3 in total

1.  Digital implant planning and guided implant surgery - workflow and reliability.

Authors:  O Schubert; J Schweiger; M Stimmelmayr; E Nold; J-F Güth
Journal:  Br Dent J       Date:  2019-01-25       Impact factor: 1.626

2.  In Vitro Comparison between Metal Sleeve-Free and Metal Sleeve-Incorporated 3D-Printed Computer-Assisted Implant Surgical Guides.

Authors:  Kyung Chul Oh; June-Sung Shim; Ji-Man Park
Journal:  Materials (Basel)       Date:  2021-01-29       Impact factor: 3.623

3.  Stereolithographic Surgical Guide with a Combination of Tooth and Bone Support: Accuracy of Guided Implant Surgery in Distal Extension Situation.

Authors:  Chia-Cheng Lin; Masahiro Ishikawa; Takeo Maida; Hsin-Chung Cheng; Keng-Liang Ou; Takashi Nezu; Kazuhiko Endo
Journal:  J Clin Med       Date:  2020-03-05       Impact factor: 4.241

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.