The analysis of in-air reverberation patterns from medical ultrasound transducers.

T Quinn1, P K Verma1.   

Abstract

National guidelines for routine ultrasound quality assurance include the measurement of transducer sensitivity using the in-air reverberation pattern generated from a transducer operating in air. The in-air reverberation method has been compared with other well-known measures of transducer sensitivity, such as the Sonora FirstCall probe tester (Sonora Medical Systems Inc, Longmont, CO, USA) and phantom-based images. There is good agreement between the in-air reverberation images and the Sonora FirstCall probe data for individual crystal sensitivity. However, the in-air reverberation approach is cheaper and easier to implement for linear and curvilinear transducers. The methods used for measuring the in-air reverberation pattern and the influence of scanner parameters such as gain and output power have been investigated. In general, reliance on a visual assessment of reverberation depth can lead to inconsistencies in the measurement of reverberation depth, when taken over a period of several months. The value of scanner parameters, in particular scanner gain, is also important when trying to measure changes in reverberation depth due to changes in transducer performance. A more accurate assessment of changes in transducer sensitivity, using the in-air reverberation method, is found by measuring the peak pixel grey scale values within a chosen reverberation band in the image. This quantitative approach can be taken a step further by assessing changes within the whole in-air reverberation pattern, by applying a two-dimensional cross correlation between two reverberation images to assess changes in transducer performance.

Entities:  

Keywords:  Quality assurance; reverberation; transducer; ultrasound

Year:  2014        PMID: 27433190      PMCID: PMC4760524          DOI: 10.1177/1742271X13517693

Source DB:  PubMed          Journal:  Ultrasound        ISSN: 1742-271X


  1 in total

1.  A review of two alternative ultrasound quality assurance programmes.

Authors:  N J Dudley; K Griffith; G Houldsworth; M Holloway; M A Dunn
Journal:  Eur J Ultrasound       Date:  2001-03
  1 in total
  4 in total

1.  Quality assurance testing of transoesophageal echocardiography probes.

Authors:  Christopher McLeod; Kirsty McNeill; Karne McBride; Scott Inglis; Stephen D Pye
Journal:  Ultrasound       Date:  2016-08-23

2.  Modeling and Experimental Characterization of Bonding Delaminations in Single-Element Ultrasonic Transducer.

Authors:  Wenxiang Ding; Maxime Bavencoffe; Marc Lethiecq
Journal:  Materials (Basel)       Date:  2021-04-27       Impact factor: 3.623

3.  Method for automatic detection of defective ultrasound linear array transducers based on uniformity assessment of clinical images - A case study.

Authors:  Robert Lorentsson; Nasser Hosseini; Jan-Olof Johansson; Wiebke Rosenberg; Benny Stenborg; Lars Gunnar Månsson; Magnus Båth
Journal:  J Appl Clin Med Phys       Date:  2018-01-11       Impact factor: 2.102

4.  Automated quality control of ultrasound based on in-air reverberation patterns.

Authors:  Pepijn van Horssen; Arnold Schilham; Dennis Dickerscheid; Niels van der Werf; Han Keijzers; Ronald van Almere; Joost Kuijer; Rob Peters; Mark Hofman
Journal:  Ultrasound       Date:  2017-09-29
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.